A two-stage rotary blood pump design with potentially lower blood trauma: a computational study

被引:28
|
作者
Thamsen, Bente [1 ]
Mevert, Ricardo [1 ]
Lommel, Michael [1 ]
Preikschat, Philip [1 ]
Gaebler, Julia [1 ]
Krabatsch, Thomas [2 ]
Kertzscher, Ulrich [1 ]
Hennig, Ewald [2 ]
Affeld, Klaus [1 ]
机构
[1] Charite, Biofluid Mech Lab, Augustenburger Pl 1, D-13353 Berlin, Germany
[2] Deutsch Herzzentrum Berlin, Dept Cardiothorac & Vasc Surg, Berlin, Germany
来源
关键词
Blood trauma; Computational fluid dynamics; Rotary blood pump; Two-stage pump; VENTRICULAR ASSIST DEVICES; SHEAR-STRESS; HEARTMATE II; DAMAGE; THROMBOSIS;
D O I
10.5301/ijao.5000482
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aim: In current rotary blood pumps, complications related to blood trauma due to shear stresses are still frequently observed clinically. Reducing the rotor tip speed might decrease blood trauma. Therefore, the aim of this project was to design a two-stage rotary blood pump leading to lower shear stresses. Methods: Using the principles of centrifugal pumps, two diagonal rotor stages were designed with an outer diameter of 22 mm. The first stage begins with a flow straightener and terminates with a diffusor, while a volute casing behind the second stage is utilized to guide fluid to the outlet. Both stages are combined into one rotating part which is pivoted by cup-socket ruby bearings. Details of the flow field were analyzed employing computational fluid dynamics (CFD). A functional model of the pump was fabricated and the pressure-flow dependency was experimentally assessed. Results: Measured pressure-flow performance of the developed pump indicated its ability to generate adequate pressure heads and flows with characteristic curves similar to centrifugal pumps. According to the CFD results, a pressure of 70 mmHg was produced at a flow rate of 5 L/min and a rotational speed of 3200 rpm. Circumferential velocities could be reduced to 3.7 m/s as compared to 6.2 m/s in a clinically used axial rotary blood pump. Flow fields were smooth with well-distributed pressure fields and comparatively few recirculation or vortices. Substantially smaller volumes were exposed to high shear stresses > 150 Pa. Conclusions: Hence, blood trauma might be reduced with this design. Based on these encouraging results, future in vitro investigations to investigate actual blood damage are intended.
引用
收藏
页码:178 / 183
页数:6
相关论文
共 50 条
  • [21] Multivariate Assessment of the Effect of Pump Design and Pump Gap Design Parameters on Blood Trauma
    Graefe, Roland
    Henseler, Andreas
    Steinseifer, Ulrich
    ARTIFICIAL ORGANS, 2016, 40 (06) : 568 - 576
  • [22] Experimental study on a two-stage rotary desiccant cooling system
    Ge, T. S.
    Li, Y.
    Wang, R. Z.
    Dai, Y. J.
    INTERNATIONAL JOURNAL OF REFRIGERATION-REVUE INTERNATIONALE DU FROID, 2009, 32 (03): : 498 - 508
  • [23] DESIGN OF TOTALLY HYDRODYNAMICALLY SUSPENDED IMPELLERS FOR MINIATURIZED ROTARY BLOOD PUMP
    Amaral, F. K.
    Timms, D.
    Bohning, F.
    Graefe, R.
    Schmitz-Rode, T.
    Steinseifer, U.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2010, 33 (07): : 425 - 425
  • [24] Novel fault-diagnosis algorithms of two-stage rotary piston vacuum pump
    Lv, Dongmei
    Han, Jiang
    Yu, Zhenhua
    Zhenkong Kexue yu Jishu Xuebao/Journal of Vacuum Science and Technology, 2015, 35 (05): : 533 - 537
  • [25] Analysis of a new PM motor design for a rotary dynamic blood pump
    Xu, LY
    Wang, FX
    Fu, MH
    Medvedev, A
    Smith, WA
    Golding, LAR
    ASAIO JOURNAL, 1997, 43 (05) : M559 - M564
  • [26] Parameter study for the miniaturization of a rotary blood pump drive system
    Pohlmann, Andre
    Hameyer, Kay
    2012 15TH INTERNATIONAL CONFERENCE ON ELECTRICAL MACHINES AND SYSTEMS (ICEMS 2012), 2012,
  • [27] Computational study of heuristics for two-stage flexible flowshops
    LISPI and Universite Claude Bernard, Villeurbanne, France
    Int J Prod Res, 5 (1399-1415):
  • [28] A computational study of heuristics for two-stage flexible flowshops
    Guinet, A
    Solomon, MM
    Kedia, PK
    Dussauchoy, A
    INTERNATIONAL JOURNAL OF PRODUCTION RESEARCH, 1996, 34 (05) : 1399 - 1415
  • [29] Design optimization of an axial blood pump with computational fluid dynamics
    Zhang, Yan
    Zhan, Zhao
    Gui, Xing-Min
    Sun, Han-Song
    Zhang, Hao
    Zheng, Zhe
    Zhou, Jian-Ye
    Zhu, Xiao-Dong
    Li, Guo-Rong
    Hu, Sheng-Shou
    Jin, Dong-Hai
    ASAIO JOURNAL, 2008, 54 (02) : 150 - 155
  • [30] Computational fluid dynamics performance prediction for the hydrodynamic bearings of the VentrAssist rotary blood pump
    Bertram, CD
    Qian, Y
    Reizes, JA
    ARTIFICIAL ORGANS, 2001, 25 (05) : 348 - 357