Purpose: To understand the propagation of electrons through the acceleration gradient of the Xoft micro‐miniature x‐ray source via electrostatic modeling of the fields and x‐ray pinhole camera imaging. Materials and Methods: 1) A 3D model of the source was built using OmniTrak3, and electron trajectories were traced from the hot cathode to the x‐ray producing anode. A finite element model of the filament temperature profile was input to a Richardson‐Dushman thermionic emission/extraction model to determine the electron emission density distribution. Assumptions made were that the field was graded linearly along the length, and that electrical connection between conductive and resistive components was ideal. 2) A ShadoCam4 x‐ray sensitive camera was used to acquire images of sources through a 30 μm pinhole mounted 15 mm from the source. The camera was 105 mm from the pinhole, providing a magnification of 7×. The camera was read out through a USB computer interface. Images were typically acquired over ten second integration times. Results: The actual images showed distinct patterns that were identifiable with model predictions. Structures associated with particular emission locations and launch angles on the filament were clearly observed. Conversely, patterns in the images were explainable in terms of parameters such as the absolute location of the filament. Image patterns which were relatively similar to the nominal had no correlation with measured spatial distribution patterns of the sources, but where there were notable variations in the image patterns it was possible to infer correlations with polar and/or azimuthal measurements. Conclusions: Electron trajectories within the accelerating structure were explored through pinhole imaging and computer modeling, and sensitivities to mechanical tolerances were established that were valuable in setting manufacturing tolerances. This study was funded by Xoft, Inc. 3 Field Precision LLC, Albuquerque, NM. 4 Rad‐icon Imaging Corp, Santa Clara, CA. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2652 / 2652
页数:1
相关论文
共 50 条
  • [41] Modeling of Inhomogeneous Intensity Distribution of X-Ray Source in Radiographic Images
    Yuksel, Ayhan
    Dokur, Zuemray
    Korurek, Mehmet
    Olmez, Tamer
    23RD INTERNATIONAL SYMPOSIUM ON COMPUTER AND INFORMATION SCIENCES, 2008, : 515 - 519
  • [42] A TECHNIQUE FOR CALIBRATING AN ELECTRON-BEAM EVAPORATOR X-RAY SOURCE
    KINZIG, RJ
    JOURNAL OF ELECTRONIC MATERIALS, 1991, 20 (09) : 681 - 686
  • [43] The quasimonochromatic X-ray radiation source on the basis of the electron beam of the microtron
    Adischev, YN
    Zabaev, VN
    Potylitsyn, AP
    Salikhbaev, US
    Safarov, AN
    Skripnik, BA
    Stepnov, YY
    Khakberdiev, I
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2003, 201 (01): : 300 - 306
  • [44] Laser electron-beam X-ray source for medical applications
    Bessonov, EG
    Vinogradov, AV
    Gorbunkov, MV
    Tur'yanskii, AG
    Feshchenko, RM
    Shabalin, YV
    PHYSICS-USPEKHI, 2003, 46 (08) : 872 - 876
  • [45] Electron beam production and characterization for the PLEIADES Thomson x-ray source
    Brown, WJ
    Hartemann, FV
    Tremaine, AM
    Springer, PT
    Le Sage, GP
    Barty, CPJ
    Rosenzweig, JB
    Crane, JK
    Cross, RR
    Fittinghoff, DN
    Gibson, DJ
    Slaughter, DR
    Anderson, S
    ADVANCED ACCELERATOR CONCEPTS, 2002, 647 : 830 - 838
  • [46] Space-resolved electron density and temperature evaluation by x-ray pinhole camera method in an ECR plasma
    Finocchiaro, G.
    Naselli, E.
    Mishra, B.
    Biri, S.
    Mazzaglia, M.
    Pidatella, A.
    Racz, R.
    Torrisi, G.
    Mascali, D.
    PHYSICS OF PLASMAS, 2024, 31 (06)
  • [47] A Hard X-Ray Pinhole Camera System for Fast Electron Bremsstrahlung Measurements in the HL-2A Tokamak
    Zhang, Y. P.
    Mazon, D.
    Zhang, J.
    Zhang, P. F.
    Malard, P.
    Xu, H. B.
    Zhou, J.
    Peysson, Y.
    Zou, X. L.
    Yang, J. W.
    Yuan, G. L.
    Isobe, M.
    Song, X. Y.
    Li, X.
    Liu, Yi
    Shi, Z. B.
    Xu, M.
    Duan, X. R.
    FUSION SCIENCE AND TECHNOLOGY, 2021, 77 (01) : 1 - 8
  • [48] Electron beam and X-ray resists
    Jones, Richard G.
    Tate, Philip C.Miller
    Advanced Materials for Optics and Electronics, 1994, 4 (02): : 139 - 153
  • [49] The structure of X-ray emissions from triggered lightning leaders measured by a pinhole-type X-ray camera
    Schaal, M. M.
    Dwyer, J. R.
    Arabshahi, S.
    Cramer, E. S.
    Lucia, R. J.
    Liu, N. Y.
    Rassoul, H. K.
    Smith, D. M.
    Matten, J. W.
    Reid, A. G.
    Hill, J. D.
    Jordan, D. M.
    Uman, M. A.
    JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 2014, 119 (02) : 982 - 1002
  • [50] PHASE-SPACE PROFILE MEASUREMENT USING AN X-RAY PINHOLE CAMERA
    OGATA, A
    MITSUHASHI, T
    KATSURA, T
    YAMAMOTO, N
    KAWAMOTO, T
    PROCEEDINGS OF THE 1989 IEEE PARTICLE ACCELERATOR CONFERENCE, VOLS 1-3: ACCELERATOR SCIENCE AND TECHNOLOGY, 1989, : 1498 - 1500