Development and evaluation of reaction gas chromatograph as a calibration method using an index compound

被引:1
|
作者
Watanabe, T [1 ]
Kato, K [1 ]
Matsumoto, N [1 ]
Maeda, T [1 ]
机构
[1] AIST Tsukuba Cent 3, Organ Analyt Chem Div, NMIJ, Tsukuba, Ibaraki 3058563, Japan
关键词
micro reactor; reaction gas chromatography; GC/MS; hydrogenation; hydrodechlorination; calibration method; index compound;
D O I
10.2116/bunsekikagaku.52.265
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In order to realize a technique that serves as a calibration method using an index compound, reaction gas chromatography was examined. An index compound represents a group which includes many compounds; compounds in this group can be calibrated using the index compound. The gas chromatograph/micro reactor/mass spectrometer (GC/MR/MS) consists of a capillary gas chromatograph/mass spectrometer and a micro reactor. By using the MR, alkenes and alkynes were transformed to alkanes having the same carbon skeleton. Ethylene in a cylinder was determined using ethane as an index compound. The results showed good agreement with a reference value. The hydrodechlorination of chlorobenzenes was examined. Chlorobenzene in a solution was determined by using benzene as an index compound. The experimental results and reference concentration showed a good agreement. The efficacy of this technique was demonstrated with these results.
引用
收藏
页码:265 / 268
页数:4
相关论文
共 50 条
  • [1] Gas chromatograph calibration of gas mixtures using a versatile precision volumetric apparatus
    Raal, J. David
    Nelson, Wayne Michael
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2022, 93 (05):
  • [2] PREPARATION OF GAS-MIXTURES FOR CALIBRATION OF A CHROMATOGRAPH BY A VOLUMETRIC GAS-FLOW METHOD
    BRASSENS, A
    ANALUSIS, 1975, 3 (10) : 576 - 579
  • [3] DETERMINATION OF ABSOLUTE CALIBRATION COEFFICIENTS FOR A GAS CHROMATOGRAPH USING MIXTURES OF UNKOWN CONCENTRATIONS
    FARZANE, NG
    CHERNYSH.DA
    ILYASOV, LV
    INDUSTRIAL LABORATORY, 1966, 32 (02): : 168 - &
  • [4] Evaluation of compound-independent calibration using gas chromatography with atomic emission detection
    Li, Gang
    Wu, Da
    Xie, Wenyan
    Zhang, Xiangmin
    Liu, Baizhan
    TALANTA, 2012, 95 : 36 - 41
  • [5] A benchmarking management index system and an evaluation method for natural gas development
    Yong R.
    Yang H.
    Ma Y.
    Yang C.
    Fang Y.
    Natural Gas Industry, 2023, 43 (02) : 24 - 31
  • [6] DETERMINATION OF ELECTRON POLARIZING CAPACITY OF A MOLECULE USING GAS CHROMATOGRAPH METHOD
    DOLOVA, IA
    KISELEV, AV
    YASHIN, YI
    ZHURNAL FIZICHESKOI KHIMII, 1972, 46 (11): : 2935 - 2936
  • [7] ANALYTICAL ATTACHMENT FOR INVESTIGATION OF HETEROGENEOUS CATALYTIC REACTION BY MICROREACTOR METHOD WITH COUPLED GAS-CHROMATOGRAPH
    SLOMKIEWICZ, PM
    PRZEMYSL CHEMICZNY, 1983, 62 (02): : 110 - 111
  • [8] A Method for Realizing Phantom Calibration and Geometry Calibration Accurately Based On a Geometry Evaluation Index
    Yang, S.
    Wu, S.
    Qi, H.
    Xu, Y.
    Zhou, L.
    MEDICAL PHYSICS, 2016, 43 (06) : 3640 - 3640
  • [9] MULTIPOINT CALIBRATION OF A GAS-CHROMATOGRAPH USING CRYOGENIC PRECONCENTRATION OF A SINGLE GAS STANDARD CONTAINING VOLATILE ORGANIC-COMPOUNDS
    RHODERICK, GC
    MILLER, WR
    ANALYTICAL CHEMISTRY, 1990, 62 (08) : 810 - 815
  • [10] A simple method for the preparation and injection of gas mixtures into a gas chromatograph using a two-component device
    Donval, J. P.
    Guyader, V
    Boissy, E.
    JOURNAL OF CHROMATOGRAPHY A, 2020, 1631