Bacterial Contamination of Ultrasound Probes at a Tertiary Referral University Medical Center

被引:11
|
作者
Chu, Kelly [1 ]
Obaid, Haron [1 ]
Babyn, Paul [1 ]
Blondeau, Joseph [2 ,3 ,4 ,5 ]
机构
[1] Univ Saskatchewan, Royal Univ Hosp, Dept Med Imaging, Saskatoon, SK S7N 0W8, Canada
[2] Royal Univ Hosp, Dept Clin Microbiol, Saskatoon, SK S7N 0W8, Canada
[3] Univ Saskatchewan, Dept Microbiol & Immunol, Saskatoon, SK S7N 0W8, Canada
[4] Univ Saskatchewan, Dept Pathol, Saskatoon, SK S7N 0W8, Canada
[5] Univ Saskatchewan, Dept Ophthalmol, Saskatoon, SK S7N 0W8, Canada
关键词
bacteria; contamination; infection; probe; ultrasound; GEL; OUTBREAK; REDUCE; RISK;
D O I
10.2214/AJR.13.12407
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to assess the adequacy of our institution's ultrasound probe-disinfecting protocols, determine compliance with the guidelines, and then implement changes if needed. MATERIALS AND METHODS. We first assessed the prevalence of bacterial contamination (and thus the potential for bacterial transmission) by swabbing all ultrasound probes (n = 31) in the radiology department and culturing the samples. Next, in conditions simulating the typical work environment, we determined the efficacy of our probe-disinfecting protocols by seeding probes with 10(4)-10(9) CFU/mL of methicillin-resistant Staphylococcus aureus (MRSA), disinfecting the seeded probes with 0.5% accelerated hydrogen peroxide, and then swabbing the disinfected probes to assess for bacterial growth. RESULTS. Seven of 31 (22.6%) probes were positive for bacterial growth-none of which were endocavity probes (0/4). Four of 14 visibly soiled probes (28.6%) showed bacterial growth, and four of seven probes positive for bacteria (57.1%) were visibly soiled. No MRSA grew after seeding probes with MRSA and then disinfecting with 0.5% accelerated hydrogen peroxide. Sonography guidelines and general disinfecting guidelines were reviewed. CONCLUSION. Our protocols for disinfecting nonendocavity and endocavity probes are compliant with sonography guidelines and general disinfecting guidelines. Although limited by a small sample size, our study showed that our protocol for disinfecting endocavity probes seems adequate. With a 25.9% bacterial contamination rate for nonendocavity probes, the adequacy of our protocol for disinfecting nonendocavity probes is more debatable; however, this bacterial contamination rate is at the lower end of the values reported in the literature. With the use of an effective disinfectant, education of sonographers, and implemented changes, we hope to decrease bacterial contamination rates and thus decrease the potential for bacterial transmission.
引用
收藏
页码:928 / 932
页数:5
相关论文
共 50 条
  • [1] Bacterial contamination of ultrasound probes
    Fries, U
    Ohrloff, C
    ULTRASONOGRAPHY IN OPHTHALMOLOGY XV, 1997, 61 : 497 - 501
  • [2] Bacterial contamination and cleanliness of emergency department ultrasound probes
    Sanz, Geoffrey E.
    Theoret, Jonathan
    Liao, Michael M.
    Erickson, Catherine
    Kendall, John L.
    CANADIAN JOURNAL OF EMERGENCY MEDICINE, 2011, 13 (06) : 384 - 389
  • [3] Incidence of residual bacterial contamination of transvaginal ultrasound probes
    Oide, Shiho
    Kuwata, Tomoyuki
    Wang, Liangcheng
    Imai, Ken
    Chikazawa, Kenro
    Horiuchi, Isao
    Takagi, Kenjiro
    Konno, Ryo
    JOURNAL OF MEDICAL ULTRASONICS, 2019, 46 (04) : 475 - 479
  • [4] Incidence of residual bacterial contamination of transvaginal ultrasound probes
    Shiho Oide
    Tomoyuki Kuwata
    Liangcheng Wang
    Ken Imai
    Kenro Chikazawa
    Isao Horiuchi
    Kenjiro Takagi
    Ryo Konno
    Journal of Medical Ultrasonics, 2019, 46 : 475 - 479
  • [5] Comparison of methods for evaluating bacterial contamination of ultrasound probes
    Harumi Koibuchi
    Shunji Hayashi
    Kazuhiko Kotani
    Yasutomo Fujii
    Kei Konno
    Yoshikazu Hirai
    Nobuyuki Taniguchi
    Journal of Medical Ultrasonics, 2009, 36 : 187 - 192
  • [6] Comparison of methods for evaluating bacterial contamination of ultrasound probes
    Koibuchi, Harumi
    Hayashi, Shunji
    Kotani, Kazuhiko
    Fujii, Yasutomo
    Konno, Kei
    Hirai, Yoshikazu
    Taniguchi, Nobuyuki
    JOURNAL OF MEDICAL ULTRASONICS, 2009, 36 (04) : 187 - 192
  • [7] Utility of a repeat endoscopic ultrasound at a tertiary referral center
    Dewitt, John M.
    Mcgreevy, Kathleen
    Sherman, Stuart
    Mchenry, Lee
    Leblanc, Julia K.
    GASTROINTESTINAL ENDOSCOPY, 2007, 65 (05) : AB200 - AB200
  • [8] OBSTETRIC ULTRASONOGRAPHY IN A TERTIARY MILITARY MEDICAL REFERRAL CENTER
    MCCOOL, RA
    BOMBARD, AT
    SOUTHERN MEDICAL JOURNAL, 1990, 83 (07) : 756 - 758
  • [9] Analysis of Glaucoma Referral Patterns at a Tertiary Medical Center
    Szulborski, Kira
    Alwreikat, Amer
    Roh, Shiyoung
    Ramsey, David J.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [10] THE CRRT EXPERIENCE IN A SINGLE TERTIARY REFERRAL MEDICAL CENTER
    Chen, Chih Chia
    Chiou, YuanYow
    PEDIATRIC NEPHROLOGY, 2021, 36 (09) : 2938 - 2939