Method Precision and Frequent Causes of Errors Observed in Point-of-Care Glucose Testing A Proficiency Testing Program Perspective

被引:10
|
作者
Aslan, Berna [1 ]
Stemp, Julia [1 ]
Yip, Paul [2 ,3 ]
Gun-Munro, Jane [1 ]
机构
[1] Inst Qual Management Healthcare, Ontario Med Assoc, Toronto, ON M5G 1E6, Canada
[2] Univ Toronto, Dept Lab Med & Pathobiol, Toronto, ON, Canada
[3] Univ Hlth Network, Toronto, ON, Canada
关键词
POC glucose; External quality assessment; Proficiency testing; Analytic quality goals; Performance standards; QUALITY; METERS;
D O I
10.1309/AJCPP5YS2MVSKBYY
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Objectives: Method imprecision, error rates, and explanatory causes that were identified in the Institute for Quality Management in Healthcare point-of-care (POC) glucose proficiency testing (PT) program were assessed in comparison with results obtained from laboratory glucose PT surveys. Methods: POC and laboratory glucose PT data were assessed from September 2009 to June 2011. Peer group means and coefficients of variation (CVs) were estimated using the robust algorithm recommended in the International Organization for Standardization/International Electrotechnical Commission 13528(E). Discordant finding investigations were also reviewed to determine the causes of significant and recurring errors. Results: POC glucose CVs were higher than laboratory method CVs (median CV, 4.5% and 1.6%, respectively). While all laboratory glucose results were within the performance limits, 305 (0.59%) of 51,379 POC glucose results exceeded limits. Investigations were required for 277 (0.53%) POC results. Pre- and postanalytical errors accounted for 76% of the discordant findings. Using wrong PT items, sample mix-up on the bench, and reporting results for the wrong sample were the most frequent reasons, while 21% of discordant findings identified manufacturer issues, and 3% were of unknown origin. Conclusions: Both method CVs and error rates were higher in POC than in laboratory glucose methods, even though larger performance limits were used for the assessment of POC glucose.
引用
收藏
页码:857 / 863
页数:7
相关论文
共 50 条
  • [11] Remote connectivity for point-of-care glucose testing.
    Weber, S
    CLINICAL CHEMISTRY, 1998, 44 : A67 - A67
  • [12] The Application of Wearable Glucose Sensors in Point-of-Care Testing
    Zhang, Sheng
    Zeng, Junyan
    Wang, Chunge
    Feng, Luying
    Song, Zening
    Zhao, Wenjie
    Wang, Qianqian
    Liu, Chen
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2021, 9
  • [13] Quality management in point-of-care glucose testing.
    Ross, G
    Bever, FN
    Uddin, Z
    Hockman, E
    CLINICAL CHEMISTRY, 2000, 46 (06) : A7 - A7
  • [14] DESIGNING A POINT-OF-CARE PROGRAM FOR COAGULATION-TESTING
    MACIK, BG
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 1995, 119 (10) : 929 - 938
  • [15] Improving a point-of-care testing program with connectivity.
    Snyder, A
    CLINICAL CHEMISTRY, 2001, 47 (06) : A178 - A178
  • [16] Outcomes and benefits of connectivity for point-of-care glucose testing.
    Bauwens, L
    CLINICAL CHEMISTRY, 2001, 47 (06) : A178 - A178
  • [17] Improving Glucose Management with Point-of-Care Testing in the Emergency Department
    Nahal Beik
    Kimberly Terry
    Current Emergency and Hospital Medicine Reports, 2014, 2 (3) : 166 - 171
  • [18] Assessment of a critical limit protocol for point-of-care glucose testing
    Lum, G
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1996, 106 (03) : 390 - 395
  • [19] Global standardization and improved technologies for point-of-care glucose testing
    Kost, Gerald J.
    Tran, Nam K.
    Abad, Victor J.
    Louie, Richard F.
    CLINICA CHIMICA ACTA, 2008, 391 (1-2) : 127 - 128
  • [20] Improving Glucose Management with Point-of-Care Testing in the Emergency Department
    Beik, Nahal
    Terry, Kimberly
    CURRENT EMERGENCY AND HOSPITAL MEDICINE REPORTS, 2014, 2 (03): : 166 - 171