Simulator testing of total knee replacements

被引:0
|
作者
Blunn, GW [1 ]
Bell, CJ [1 ]
Walker, PS [1 ]
Chatterjee, S [1 ]
Perry, J [1 ]
Campbell, P [1 ]
Haider, H [1 ]
Paul, JP [1 ]
机构
[1] UCL, Inst Orthopaed, Ctr Biomed Engn, London, England
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In order to determine the effect of improvements in design and materials on wear of total knee replacement, a valid test method is required. This test method should reproduce the wear patterns seen on retrieved knee replacements. We have tested various designs of implants using a force-control knee simulator and analysed the wear on these implants. The tests were conducted in 50 per cent bovine serum for up to ten million cycles. On condylar replacements, wear was associated with scratching and burnishing. Wear rates between different designs could be discriminated. Delamination wear occurred only in those samples which had been artificially aged. Tests showed that wear on a meniscal knee design equalled that, or was greater than that, observed on a condylar design. On meniscal designs wear was associated mainly with the meniscal bearing surface.
引用
收藏
页码:113 / 125
页数:13
相关论文
共 50 条
  • [41] Cost, safety, and rehabilitation of same-stage, bilateral total knee replacements compared to two-stage total knee replacements
    Raymond C. W. Wan
    Jason C. H. Fan
    Yuk-Wah Hung
    Ka-Bon Kwok
    Carmen K. M. Lo
    Kwong-Yin Chung
    Knee Surgery & Related Research, 33
  • [42] Cost, safety, and rehabilitation of same-stage, bilateral total knee replacements compared to two-stage total knee replacements
    Wan, Raymond C. W.
    Fan, Jason C. H.
    Hung, Yuk-Wah
    Kwok, Ka-Bon
    Lo, Carmen K. M.
    Chung, Kwong-Yin
    KNEE SURGERY & RELATED RESEARCH, 2021, 33 (01)
  • [43] FRICTION TORQUE STUDIES OF TOTAL JOINT REPLACEMENTS USE OF A SIMULATOR
    SCALES, JT
    KELLY, P
    GODDARD, D
    ANNALS OF THE RHEUMATIC DISEASES, 1969, 28 (5S) : 30 - &
  • [44] Premature failure of Kinemax Plus total knee replacements
    Reay, E.
    Wu, J.
    Holland, J.
    Deehan, D.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (05): : 604 - 611
  • [45] The pros and cons of modularity in total knee replacements - Reply
    Morrey, BF
    Babis, GC
    Trousdale, RT
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (08): : 1481 - 1481
  • [46] An analysis of rotating-platform total knee replacements
    Bourne, RB
    Masonis, J
    Anthony, M
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) : 173 - 180
  • [47] TOTAL CONDYLAR AND OXFORD TOTAL KNEE REPLACEMENTS - A 3.3 YEAR COMPARISON
    BOURNE, RB
    RORABECK, CH
    DAVEY, R
    MACNAB, J
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1985, 67 (02): : 331 - 331
  • [48] Anticipating Urinary Retention Following Total Hip and Total Knee Replacements
    Hamed, Y.
    Ramesh, A.
    Taylor, R.
    Michaud, R.
    MALAYSIAN ORTHOPAEDIC JOURNAL, 2020, 14 (03) : 110 - 113
  • [49] STRESSES IN POLYETHYLENE COMPONENTS OF CONTEMPORARY TOTAL KNEE REPLACEMENTS
    BARTEL, DL
    RAWLINSON, JJ
    BURSTEIN, AH
    RANAWAT, CS
    FLYNN, WF
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1995, (317) : 76 - 82
  • [50] EXCISION ARTHROPLASTY FOR INFECTED CONSTRAINED TOTAL KNEE REPLACEMENTS
    LETTIN, AWF
    NEIL, MJ
    CITRON, ND
    AUGUST, A
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (02): : 220 - 224