Optimum tension for partial ossicular replacement prosthesis reconstruction in the human middle ear

被引:48
|
作者
Morris, DP [1 ]
Bance, M [1 ]
van Wijhe, RG [1 ]
Kiefte, M [1 ]
Smith, R [1 ]
机构
[1] Dalhousie Univ, Div Otolaryngol, Ear & Audit Res Lab, Halifax, NS B3H 2Y9, Canada
来源
LARYNGOSCOPE | 2004年 / 114卷 / 02期
关键词
Ossicular prosthesis; ossicular replacement; cadaver; ear middle; laser Doppler vibrometer;
D O I
10.1097/00005537-200402000-00024
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective: Hearing results from ossiculoplasty are unpredictable. There are many potentially modifiable parameters. One parameter that has not been adequately investigated in the past is the effect of tension on the mechanical functioning of the prosthesis. Our goal was to investigate this parameter further, with the hypothesis that the mechanical functioning of partial ossicular replacement prostheses (PORP) from the stapes head to the eardrum will be affected by the tension that they are placed under. Methods: Fresh temporal bones were used to reconstruct a missing incus defect with a PORP-type prosthesis. Three different lengths of PORP were used, and the stapes vibrations were measured with a laser Doppler vibrometer using a calibrated standard sound in the ear canal. Eight temporal bones were used. Results: Tension had a very significant effect on stapes vibration. In general, loose prostheses resulted in the best overall vibration transmission. The effects were most marked at the lower frequencies. There was a slight advantage to tight prostheses in the higher frequencies, but much less than the decrement in lower frequencies with tight prostheses. Conclusion: In ossicular reconstruction, best stapes vibration results in our model are achieved by shorter prostheses, which result in lower tension.
引用
收藏
页码:305 / +
页数:5
相关论文
共 50 条
  • [21] Effect of Ossicular Prosthesis Biofilms on Middle Ear Scarring and Hearing Outcomes
    Jaryszak, Eric M.
    Sampson, Edith M.
    Antonelli, Patrick J.
    OTOLOGY & NEUROTOLOGY, 2009, 30 (08) : 1191 - 1195
  • [22] A NEW TECHNIQUE FOR TOTAL OSSICULAR REPLACEMENT IN THE MIDDLE-EAR
    SCHOBEL, H
    ARCHIVES OF OTO-RHINO-LARYNGOLOGY-ARCHIV FUR OHREN-NASEN-UND KEHLKOPFHEILKUNDE, 1981, 231 (2-3): : 583 - 584
  • [23] Middle-ear dynamics before and after ossicular replacement
    Ferris, P
    Prendergast, PJ
    JOURNAL OF BIOMECHANICS, 2000, 33 (05) : 581 - 590
  • [24] Titanium Clip Ball Joint: A Partial Ossicular Reconstruction Prosthesis
    Beutner, Dirk
    Luers, Jan Christoffer
    Bornitz, Matthias
    Zahnert, Thomas
    Huttenbrink, Karl-Bernd
    OTOLOGY & NEUROTOLOGY, 2011, 32 (04) : 646 - 649
  • [25] Autologous incus versus titanium partial ossicular replacement prosthesis in reconstruction of Austin type A ossicular defects: a prospective randomised clinical trial
    Amith, Naragund
    Mudhol, R. S.
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2017, 131 (05): : 391 - 398
  • [26] Reconstruction of the ossicular chain in the middle ear with glass ionomer cement
    Brask, T
    LARYNGOSCOPE, 1999, 109 (04): : 573 - 576
  • [27] Hearing results with a hydroxylapatite/titanium bell partial ossicular replacement prosthesis
    Smith, J
    Gardner, E
    Dornhoffer, JL
    LARYNGOSCOPE, 2002, 112 (10): : 1796 - 1799
  • [28] THE DEVELOPMENT OF THE OSSICULAR LIGAMENTS IN THE HUMAN MIDDLE-EAR
    HARTWEIN, JHJ
    RAUCHFUSS, A
    ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 1987, 244 (01): : 23 - 25
  • [29] Functional results of plastipore prostheses for middle ear ossicular chain reconstruction
    Bayazit, Y
    Göksu, N
    Beder, L
    LARYNGOSCOPE, 1999, 109 (05): : 709 - 711
  • [30] Cartilage versus partial ossicular replacement prosthesis in ossiculoplasty during cholesteatoma surgery
    Mobashir M.K.
    Fouad Y.A.
    Alshawadfy M.A.
    Hassaan M.R.
    Anany A.M.
    The Egyptian Journal of Otolaryngology, 2018, 34 (1) : 42 - 47