Capillary electrophoretic determination of ammonia using headspace single-drop microextraction

被引:43
|
作者
Pranaityte, Birute
Jermak, Svetlana
Naujalis, Evaldas
Padarauskas, Audrius
机构
[1] Vilnius State Univ, Dept Analyt & Environm Chem, LT-03225 Vilnius, Lithuania
[2] Inst Semicond Phys, Lab Metrol Chem, LT-01108 Vilnius, Lithuania
关键词
single-drop microextraction; capillary electrophoresis; ammonia;
D O I
10.1016/j.microc.2006.10.001
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new method involving headspace single-drop microextraction (SDME) and capillary electrophoresis (CE) is developed for the preconcentration and determination of ammonia (as dissolved NH3 and ammonium ion). An aqueous microdrop (5 mu L) containing 1 mmol/L H3PO4 and 0.5 mmol/L KH2PO4 (as internal standard) was used as the acceptor phase. Common experimental parameters (sample and acceptor phase pH, extraction temperature, extraction time) affecting the extraction efficiency were investigated. Proposed SDME-CE method provided about 14-fold enrichment in about 20 rnin. The calibration curve was linear for concentrations of NH4 in the range from 5 to 100 mu mol/L (R-2 = 0.996). The LOD (S/N=3) was estimated to be 1.5 mu mol/L of NH4+. Such detection sensitivity is high enough for ammonia determination in common environmental and biological samples. Finally, headspace SDME was applied to determine ammonia in human blood, seawater and milk samples with spiked recoveries in the range of 96-107%. (C) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:48 / 52
页数:5
相关论文
共 50 条