High failure rate for an epicardial implantable cardioverter-defibrillator lead: Implications for long-term follow-up of patients with an implantable cardioverter-defibrillator

被引:43
|
作者
Brady, PA [1 ]
Friedman, PA [1 ]
Trusty, JM [1 ]
Grice, S [1 ]
Hammill, SC [1 ]
Stanton, MS [1 ]
机构
[1] Mayo Clin & Mayo Fdn, Div Cardiovasc Dis, Rochester, MN 55905 USA
关键词
D O I
10.1016/S0735-1097(97)00529-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. The purpose of this study was to determine the risk of epicardial lead failure during long-term follow-up and its mode of presentation, Background. Despite the high prevalence of epicardial lead-based implantable cardioverter-defibrillators, their long-term performance is unknown, and appropriate follow-up has not been established, Methods. The study group comprised all patients in whom an epicardial lead system was implanted at the Mayo Clinic between October 31, 1984 and November 3, 1994, The number of lead fractures and leads with fluid within the insulation and the made of presentation were determined retrospectively by review of patient visits, radiographs of lead systems and data derived from formal lead testing. Results. At 4 years, the survival rate free of lead malfunction, using formal lead testing, for 160 Medtronic epicardial patches (models 6897 and 6921) was 72% compared with 92.5% for the 179 Cardiac Pacemaker, Inc. (CPI) patches (models 0040 and 0041) (p = = 0.01). In addition, five Medtronic patches in three patients had fluid within the lead insulation but no obvious fracture, No CPI patches had fluid identified within the leads. Of 330 Medtronic epicardial pace/sense leads (model 6917), the 4-year survival rate Free of lead malfunction as assessed by lead testing was 94%. In all, 18 presentations of lead malfunction were found in 17 patients (2 patients had more than one lead fracture at different times). In 11 (58%) of these presentations, the patients were asymptomatic despite the presence of obvious lead fracture. Conclusions. Epicardial lead malfunction is common on long-term follow-up, and some leads have a failure rate of 28% at 4 years. Many patients with fractured leads remain asymptomatic, despite involvement of multiple leads in some casts, Therefore, consideration should be given to regular periodic lead testing in addition to routine X-ray examination, as as symptomatic lead malfunction can present with normal chest X-ray findings. (C) 1998 by the American College of Cardiology.
引用
收藏
页码:616 / 622
页数:7
相关论文
共 50 条
  • [21] An algorithm to predict implantable cardioverter-defibrillator lead failure
    Gunderson, BD
    Patel, AS
    Bounds, CA
    Shepard, RK
    Wood, MA
    Ellenbogen, KA
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (09) : 1898 - 1902
  • [22] Long-Term Performance of the Riata/ST Implantable Cardioverter-Defibrillator Lead
    Stroeker, Erwin
    de Asmundis, Carlo
    Vanduynhoven, Philippe
    De Vadder, Katrien
    De Vusser, Philip
    Mullens, Wilfried
    Chierchia, Gian-Battista
    Brugada, Pedro
    Czapla, Jens
    La Meir, Mark
    Wellens, Francis
    Van Herendael, Hugo
    Rivero-Ayerza, Maximo
    AMERICAN JOURNAL OF CARDIOLOGY, 2016, 117 (05): : 807 - 812
  • [23] Over 10 years with an implantable cardioverter-defibrillator - a long term follow-up of 60 patients
    Sterlinski, Maciej
    Przybylski, Andrzej
    Gepner, Katarzyna
    Syska, Pawel
    Maciag, Aleksander
    Lewandowski, Michal
    Kowalik, Ilona
    Pytkowski, Manusz
    Sitkowska-Rysiak, Ewa
    Lichomski, Jerzy
    Szwed, Hanna
    Sadowski, Zygmunt
    KARDIOLOGIA POLSKA, 2010, 68 (09) : 1023 - 1031
  • [24] Selecting patients for an Implantable cardioverter-defibrillator
    Mittal, Suneet
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 51 (03) : 297 - 299
  • [25] Benefits of the implantable cardioverter-defibrillator
    Epstein, Andrew E.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 52 (14) : 1122 - 1127
  • [26] THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    DELUCA, SA
    GIRARD, M
    AMERICAN FAMILY PHYSICIAN, 1989, 39 (01) : 141 - 143
  • [27] THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    MIROWSKI, M
    REID, PR
    MOWER, MM
    WATKINS, L
    PLATIA, EV
    GRIFFITH, LSC
    JUANTEGUY, JM
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (03): : 534 - 540
  • [28] Implantable cardioverter-defibrillator shocks
    Raitt, Merritt H.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 51 (14) : 1366 - 1368
  • [29] Development of the Implantable Cardioverter-Defibrillator
    Maron, Barry J.
    Estes, N. A. Mark
    Rowin, Ethan J.
    Maron, Martin S.
    Reynolds, Matthew R.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2023, 82 (04) : 353 - 373
  • [30] THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - AN UPDATE
    MIROWSKI, M
    JOURNAL OF CARDIOVASCULAR MEDICINE, 1984, 9 (03): : 191 - &