Simplified Stokes polarimeter based on division-of-amplitude

被引:8
|
作者
Negara, Christian [1 ]
Li, Zheng [1 ,2 ]
Laengle, Thomas [1 ]
Beyerer, Juergen [1 ,2 ]
机构
[1] Fraunhofer Inst Optron Syst Technol & Image Expl, Fraunhoferstr 1, D-76131 Karlsruhe, Germany
[2] Karlsruhe Inst Technol, Kaiserstr 12, D-76131 Karlsruhe, Germany
关键词
polarimetry; ellipsometry; Stokes vector; polarization state detector; division-of-amplitude; beam splitter; retarder; quarter-wave plate; BEAM-SPLITTERS; OPTIMIZATION; ELLIPSOMETER; RETARDATION; CALIBRATION; FILMS;
D O I
10.1117/12.2532399
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A polarization state detector (PSD) measures the state of polarization of the detected light. The state of polarization is fully described by the Stokes vector containing four Stokes parameters. A division-of-amplitude photopolarimeter (DOAP) measures the four Stokes parameters by simultaneously acquiring four intensities using photodetectors. A key component of the DOAP is the first beam splitter, which splits up the incoming beam into two beams. The effect of the beam splitter on the state of polarization of the reflected (r) and transmitted (t) beam is determined by six parameters: R; T;Psi(r);Psi(t);Psi(r), and Delta(t). R and T are the reflectance and transmittance, and (Psi(r);Delta(r)) and (Psi(t);Delta(t)) are the ellipsometric parameters of the beam splitter in reflection and transmission, respectively. To measure the Stokes vector with high accuracy, the six optical parameters must be chosen appropriately. In previous work, the optimal parameters of the beam splitter have been determined as R = T = 1/2; cos(2) 2 Psi(r) = 1/3;Psi(t) = Pi=2 and (Delta(r) modulo Pi = Pi/2 by calculating the maximum of the absolute value of the determinant of the instrument matrix. Using additional quarter-wave plates eliminates the constraint on the retardance and hence simplifies the manufacturing process of the beam splitter, especially when broadband application is intended. To compensate a suboptimal value of Delta(r) - Delta(t), the azimuthal angles of the principal axes of the retarders must be adjusted, for which we provide analytic formulas. Hence, a DOAP with retarders is also optimal in the sense that the same values for the determinant and condition number of the instrument matrix are obtained. When using two additional retarders, it is necessary to mount both at the same light path in order to obtain an optimal DOAP. We will show that is also possible to get an optimal DOAP with only one additional quarter-wave plate instead of two, if one of the Wollaston prisms is rotated.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Calibration and testing of a sixteen-beam grating-based division-of-amplitude photopolarimeter
    Cui, Y
    Azzam, RMA
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (12): : 5552 - 5558
  • [22] Spectral polarization measurements by use of the grating division-of-amplitude photopolarimeter
    Krishnan, S
    Hampton, S
    Rix, J
    Taylor, B
    Azzam, RMA
    APPLIED OPTICS, 2003, 42 (07) : 1216 - 1227
  • [23] Single-layer-coated beam splitters for the division-of-amplitude photopolarimeter
    Azzam, RMA
    Sudradjat, FF
    APPLIED OPTICS, 2005, 44 (02) : 190 - 196
  • [24] A new method of data processing of division-of-amplitude polarization navigation angle sensor
    Dai, Yansheng
    Li, Leilei
    Chen, Jiabin
    Zhang, Xusheng
    Zhang, Zhongyu
    2017 29TH CHINESE CONTROL AND DECISION CONFERENCE (CCDC), 2017, : 6578 - 6583
  • [25] Fast ellipsometry and Mueller-matrix ellipsometry using the division-of-amplitude photopolarimeter
    Krishnan, S
    Nordine, PC
    INTERNATIONAL SYMPOSIUM ON POLARIZATION ANALYSIS AND APPLICATIONS TO DEVICE TECHNOLOGY, 1996, 2873 : 152 - 156
  • [26] Multi-point calibration method for division-of-amplitude photopolarimetric detection system
    Wang, Fangbin
    Liu, Tao
    Hong, Jin
    Zhu, Darong
    Xu, Lu
    Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering, 2017, 46 (05):
  • [27] Spectroscopic ellipsometry using the grating division-of-amplitude photopolarimeter (G-DOAP)
    Krishnan, S
    Hampton, S
    Azzam, RMA
    THIN SOLID FILMS, 2004, 455 : 24 - 32
  • [28] General method for the calculation of Stokes parameters with a division of wavefront polarimeter
    Soto, GR
    Smith, PR
    REVISTA MEXICANA DE FISICA, 1998, 44 (04) : 392 - 396
  • [29] CALIBRATION, PROPERTIES, AND APPLICATIONS OF THE DIVISION-OF-AMPLITUDE PHOTOPOLARIMETER AT 632.8 AND 1523-NM
    KRISHNAN, S
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1992, 9 (09): : 1615 - 1622
  • [30] Bilayer pellicle and bilayer-coated prism beamsplitters for the division-of-amplitude photopolarimeter
    Azzam, RMA
    De, A
    Sudradjat, FF
    OPTICAL ENGINEERING, 2005, 44 (07) : 1 - 9