EVALUATION OF THE I-STAT PORTABLE CLINICAL ANALYZER FOR MEASUREMENT OF IONIZED CALCIUM AND SELECTED BLOOD CHEMISTRY VALUES IN ASIAN ELEPHANTS (ELEPHAS MAXIMUS)

被引:0
|
作者
Tarbert, Danielle K. [1 ]
Behling-Kelly, Erica [2 ]
Priest, Heather [2 ,3 ]
Childs-Sanford, Sara [1 ]
机构
[1] Cornell Univ, Coll Vet Med, Sect Zool Med, Dept Clin Sci, 930 Campus Rd, Ithaca, NY 14853 USA
[2] Cornell Univ, Coll Vet Med, Dept Populat Med & Diagnost Sci, 930 Campus Rd, Ithaca, NY 14853 USA
[3] Charles River Labs Inc, 6995 Longley Ln, Reno, NV 89511 USA
关键词
Asian elephant; i-STAT; ionized calcium; venous blood gas; biochemical analytes; method comparison; SAMPLES; PARAMETERS; ANALYTES; URINE; DOGS; PH;
D O I
10.1638/2016-0150R1.1
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The i-STAT (R) portable clinical analyzer (PCA) provides patient-side results for hematologic, biochemical, and blood gas values when immediate results are desired. This analyzer is commonly used in nondomestic animals; however, validation of this method in comparison with traditional benchtop methods should be performed for each species. In this study, the i-STAT PCA was compared with the Radiometer ABL 800 Flex benchtop analyzer using 24 heparinized whole blood samples obtained from healthy E. maximus. In addition, the effect of sample storage was evaluated on the i-STAT PCA. Analytes evaluated were hydrogen ion concentration (pH), glucose, potassium (K+), sodium (Na+), bicarbonate (HCO3), total carbon dioxide (TCO2), partial pressure of carbon dioxide (PCO2), and ionized calcium (iCa(2+)). Statistical analysis using correlation coefficients, Passing-Bablok regression analysis, and Bland-Altman plots found good agreement between results from samples run immediately after phlebotomy and 4 hr postsampling on the i-STAT PCA with the exception of K+, which is known to change with sample storage. Comparison of the results from the two analyzers at 4 hr postsampling found very strong or strong correlation in all values except K+, with statistically significant bias in all values except glucose and PCO2. Despite bias, mean differences assessed via Bland-Altman plots were clinically acceptable for all analytes excluding K+. Within the reference range for iCa(2+), the iCa(2+) values obtained by the i-STAT PCA and Radiometer ABL 800 Flex were close in value, however in light of the constant and proportionate biases detected, overestimation at higher values and underestimation at lower values of iCa(2+) by the i-STAT PCA would be of potential concern. This study supports the use of the i-STAT PCA for the evaluation of these analytes, with the exception of K+, in the Asian elephant.
引用
收藏
页码:319 / 327
页数:9
相关论文
共 25 条
  • [21] Measurement of ionized calcium in canine blood samples collected in prefilled and self-filled heparinized syringes using the i-STAT point-of-care analyzer
    Tappin, Simon
    Rizzo, Francesca
    Dodkin, Steve
    Papasouliotis, Kostas
    Tasker, Severine
    Murphy, Kate
    VETERINARY CLINICAL PATHOLOGY, 2008, 37 (01) : 66 - 72
  • [22] Evaluation of the i-STAT portable point-of-care analyzer for determination of blood gases and acid-base status in newborn calves
    Bleul, Ulrich
    Goetz, Eva
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2014, 24 (05) : 519 - 528
  • [23] Haematology, blood chemistry and urine parameters of free-ranging plains viscachas (Lagostomus maximus) in Argentina determined by use of a portable blood analyser (i-STAT®) and conventional laboratory methods
    Wenker, C. J.
    Hunziker, D.
    Lopez, J.
    Oppliger, H.
    Forrer, R.
    Lutz, H.
    JOURNAL OF VETERINARY MEDICINE SERIES A-PHYSIOLOGY PATHOLOGY CLINICAL MEDICINE, 2007, 54 (05): : 260 - 264
  • [24] Commercial Hy-Line W-36 pullet and laying hen venous blood gas and chemistry profiles utilizing the portable i-STAT®1 analyzer
    Schaal, T. P.
    Arango, J.
    Wolc, A.
    Brady, J. V.
    Fulton, J. E.
    Rubinoff, I.
    Ehr, I. J.
    Persia, M. E.
    O'Sullivan, N. P.
    POULTRY SCIENCE, 2016, 95 (02) : 466 - 471
  • [25] Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens
    Sauer, Z. C.
    Taylor, K.
    Wolc, A.
    Viall, A.
    Fulton, J. E.
    Settar, P.
    Rubinoff, I
    Schaal, T.
    Sato, Y.
    POULTRY SCIENCE, 2020, 99 (07) : 3487 - 3490