EVALUATION OF THE I-STAT PORTABLE CLINICAL ANALYZER FOR MEASUREMENT OF IONIZED CALCIUM AND SELECTED BLOOD CHEMISTRY VALUES IN ASIAN ELEPHANTS (ELEPHAS MAXIMUS)

被引:0
|
作者
Tarbert, Danielle K. [1 ]
Behling-Kelly, Erica [2 ]
Priest, Heather [2 ,3 ]
Childs-Sanford, Sara [1 ]
机构
[1] Cornell Univ, Coll Vet Med, Sect Zool Med, Dept Clin Sci, 930 Campus Rd, Ithaca, NY 14853 USA
[2] Cornell Univ, Coll Vet Med, Dept Populat Med & Diagnost Sci, 930 Campus Rd, Ithaca, NY 14853 USA
[3] Charles River Labs Inc, 6995 Longley Ln, Reno, NV 89511 USA
关键词
Asian elephant; i-STAT; ionized calcium; venous blood gas; biochemical analytes; method comparison; SAMPLES; PARAMETERS; ANALYTES; URINE; DOGS; PH;
D O I
10.1638/2016-0150R1.1
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The i-STAT (R) portable clinical analyzer (PCA) provides patient-side results for hematologic, biochemical, and blood gas values when immediate results are desired. This analyzer is commonly used in nondomestic animals; however, validation of this method in comparison with traditional benchtop methods should be performed for each species. In this study, the i-STAT PCA was compared with the Radiometer ABL 800 Flex benchtop analyzer using 24 heparinized whole blood samples obtained from healthy E. maximus. In addition, the effect of sample storage was evaluated on the i-STAT PCA. Analytes evaluated were hydrogen ion concentration (pH), glucose, potassium (K+), sodium (Na+), bicarbonate (HCO3), total carbon dioxide (TCO2), partial pressure of carbon dioxide (PCO2), and ionized calcium (iCa(2+)). Statistical analysis using correlation coefficients, Passing-Bablok regression analysis, and Bland-Altman plots found good agreement between results from samples run immediately after phlebotomy and 4 hr postsampling on the i-STAT PCA with the exception of K+, which is known to change with sample storage. Comparison of the results from the two analyzers at 4 hr postsampling found very strong or strong correlation in all values except K+, with statistically significant bias in all values except glucose and PCO2. Despite bias, mean differences assessed via Bland-Altman plots were clinically acceptable for all analytes excluding K+. Within the reference range for iCa(2+), the iCa(2+) values obtained by the i-STAT PCA and Radiometer ABL 800 Flex were close in value, however in light of the constant and proportionate biases detected, overestimation at higher values and underestimation at lower values of iCa(2+) by the i-STAT PCA would be of potential concern. This study supports the use of the i-STAT PCA for the evaluation of these analytes, with the exception of K+, in the Asian elephant.
引用
收藏
页码:319 / 327
页数:9
相关论文
共 25 条
  • [1] AN EVALUATION OF BLOOD-GAS AND IONIZED CALCIUM MEASUREMENT OF THE I-STAT ANALYZER
    MOCK, T
    YATSCOFF, RW
    CLINICAL CHEMISTRY, 1995, 41 (06) : S182 - S182
  • [2] Evaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood
    Yilmaz, Oznur
    Karapinar, Tolga
    VETERINARY CLINICAL PATHOLOGY, 2019, 48 (01) : 31 - 35
  • [3] Evaluation of the i-STAT Portable Clinical Analyzer in dogs
    Matburger, C
    Henke, J
    Hirschberger, J
    Matburger, S
    Erhardt, W
    TIERARZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2000, 28 (02): : 132 - 137
  • [4] Evaluation of the i-STAT® Portable Clinical Analyzer for the determination of blood creatinine.
    Marver, CL
    Santrach, PJ
    Burritt, MF
    CLINICAL CHEMISTRY, 1999, 45 (06) : A33 - A33
  • [5] Evaluation of the i-STAT portable clinical analyzer for use in a helicopter
    Burritt, MF
    Santrach, PJ
    Hankins, DG
    Herr, D
    Newton, NC
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1996, 56 : 121 - 128
  • [6] Evaluation of i-STAT portable clinical analyzer in a hemodialysis unit
    Gault, MH
    Harding, CE
    CLINICAL BIOCHEMISTRY, 1996, 29 (02) : 117 - 124
  • [7] Evaluation of the i-STAT portable clinical analyzer in chickens (Gallus gallus)
    Steinmetz, Hanspeter W.
    Vogt, Rainer
    Kastner, Sabine
    Riond, Barbara
    Hatt, Jean-Michel
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2007, 19 (04) : 382 - 388
  • [8] EVALUATION OF BLOOD-GAS AND PH METHODS ON THE I-STAT(TM) PORTABLE CLINICAL ANALYZER
    BOTTJEN, PL
    HAVEN, MC
    CLINICAL CHEMISTRY, 1995, 41 (06) : S196 - S196
  • [9] Selected Blood Chemistry Values in Mobility-Impaired Broiler Breeder Hens with Suspected Calcium Tetany Using the i-STAT® Handheld Clinical Analyzer
    Martin, Michael P.
    Wineland, Michael
    Fletcher, Oscar J.
    Barnes, H. John
    AVIAN DISEASES, 2011, 55 (03) : 340 - 345
  • [10] EVALUATION OF A NOVEL POINT-OF-CARE SYSTEM, THE I-STAT PORTABLE CLINICAL ANALYZER
    ERICKSON, KA
    WILDING, P
    CLINICAL CHEMISTRY, 1993, 39 (02) : 283 - 287