摘要
To obtain maximal resolution in scanning tunneling microscopy (STM) and atomic force microscopy, the size of the protruding tip orbital has to be minimized. Beryllium as tip material is a promising candidate for enhanced resolution because a beryllium atom has just four electrons, leading to a small covalent radius of only 96 pm. Besides that, beryllium is conductive and has a high elastic modulus, which is a necessity for a stable tip apex. However, beryllium tips that are prepared ex situ are covered with a robust oxide layer, which cannot be removed by just heating the tip. Here, the authors present a successful preparation method that combines the heating of the tip by field emission and a mild collision with a clean metal plate. That method yields a clean, oxide-free tip surface as proven by a work function of Phi(expt)=5.5 eV as deduced from a current-distance curve. Additionally, a STM image of the Si-(111)-(7x7) is presented to prove the single-atom termination of the beryllium tip. (C) 2010 American Vacuum Society. [DOI: 10.1116/1.3294706]
引用
收藏
页数:3
相关论文
共 50 条
  • [1] Force microscopy with light-atom probes
    Hembacher, S
    Giessibl, FJ
    Mannhart, J
    SCIENCE, 2004, 305 (5682) : 380 - 383
  • [2] Preparation of platinum iridium scanning probe microscopy tips
    Sorensen, AH
    Hvid, U
    Mortensen, MW
    Morch, KA
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (07): : 3059 - 3067
  • [3] Preparation of platinum/iridium scanning probe microscopy tips
    Department of Physics, Technical University of Denmark, Building 307, DK-2800 Lyngby, Denmark
    Rev. Sci. Instrum., 7 (3059-3067):
  • [4] Time gated ion microscopy of light-atom interactions
    Tzallas, P.
    Bergues, B.
    Rompotis, D.
    Tsatrafyllis, N.
    Chatziathanassiou, S.
    Muschet, A.
    Veisz, L.
    Schroeder, H.
    Charalambidis, D.
    JOURNAL OF OPTICS, 2018, 20 (02)
  • [5] Fullerene tips for scanning probe microscopy
    Kelly, KF
    Sarkar, D
    Oldenburg, SJ
    Hale, GD
    Halas, NJ
    FULLERENES AND PHOTONICS III, 1996, 2854 : 114 - 122
  • [6] Cold-atom scanning probe microscopy
    Gierling, M.
    Schneeweiss, P.
    Visanescu, G.
    Federsel, P.
    Haeffner, M.
    Kern, D. P.
    Judd, T. E.
    Guenther, A.
    Fortagh, J.
    NATURE NANOTECHNOLOGY, 2011, 6 (07) : 446 - 451
  • [7] An electrochemical etching procedure for fabricating scanning tunneling microscopy and atom-probe field-ion microscopy tips
    Kim, YC
    Seidman, DN
    METALS AND MATERIALS INTERNATIONAL, 2003, 9 (04) : 399 - 404
  • [8] An electrochemical etching procedure for fabricating scanning tunneling microscopy and atom-probe field-ion microscopy tips
    Yeong-Cheol Kim
    David N. Seidman
    Metals and Materials International, 2003, 9 : 399 - 404
  • [9] Implementation of recycling routes for scanning probe microscopy tips
    de Souza, CP
    Andrade, MS
    Neves, BRA
    MICROSCOPY AND MICROANALYSIS, 2002, 8 (06) : 509 - 517
  • [10] Diamond as Advanced Material for Scanning Probe Microscopy Tips
    Kranz, Christine
    ELECTROANALYSIS, 2016, 28 (01) : 35 - 45