Although IMRT has been shown clinically to increase skin doses for some patients, it has also been shown that intensity modulated delivery does not, of itself, increase skin doses. The reason for this apparent difference is that inverse planning can result in solutions that give high fluence to tangential beam segments near the skin surface, in an attempt to counter the build-up region. In cases where the clinical target volume (CTV) stops short of the skin surface, but the planning target volume (PTV) does not, there is no clinical reason to treat the skin. The CTV-PTV margin exists purely to ensure that fields are large enough to allow for geometrical uncertainties. With an objective function based on the doses to the PTV, it is possible for a plan that gives excess fluence to the skin to have a lower objective function, and hence to be preferred in an optimization. We describe I technique of plan evaluation, based on analysis of a plan by recalculating several plans in which the isocentre has been offset by a distance equal to the CTV-PTV margin. We demonstrate that changes to a plan that reduce a PTV-based objective can give a worse dose distribution to the CTV when systematic and random set-up errors are accounted for, and increase skin dose. Several possible strategies for avoiding this problem are discussed, including the use of the skin as an organ at risk, modification of the PTV to avoid the skin, and the use of 'pretend bolus' applied in planning but not in treatment. The latter gave the best results. The possibility of using the evaluation method itself, as the basis of an objective function for optimization, is discussed.
引用
收藏
页码:4919 / 4928
页数:10
相关论文
共 50 条
  • [41] Conformal intensity-modulated radiotherapy (IMRT) delivered by robotic linac - testing IMRT to the limit?
    Webb, S
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07): : 1639 - 1654
  • [42] Optimization criteria in intensity-modulated radiotherapy
    Sauer, Otto A.
    Zeitschrift fur Medizinische Physik, 2003, 13 (02): : 99 - 107
  • [43] An optimization algorithm for intensity modulated radiotherapy - The simulated dynamics with dose-volume constraints
    Hou, Q
    Wang, J
    Chen, Y
    Galvin, JM
    MEDICAL PHYSICS, 2003, 30 (01) : 61 - 68
  • [44] Intensity-modulated radiotherapy (IMRT) for carcinoma of the maxillary sinus: A comparison of IMRT planning systems
    Ahmed, Raef S.
    Ove, Roger
    Duan, Jun
    Popple, Richard
    Cobb, Glenn B.
    MEDICAL DOSIMETRY, 2006, 31 (03) : 224 - 232
  • [45] The Impact of Normal Tissue Constraints On PTV Dose Homogeneity for Intensity Modulated Radiotherapy (IMRT), Volume Modulated Arc Therapy (VMAT) and Tomotherapy
    Peng, J.
    McDonald, D.
    Ashenafi, M.
    Ellis, A.
    Vanek, K.
    MEDICAL PHYSICS, 2014, 41 (06) : 322 - 322
  • [46] Evaluation of the effect of random setup errors on dose delivery in Intensity Modulated Radiotherapy
    Bharati, Avinav
    Mandal, Susama Rani
    Srivastava, Anoop Kumar
    Rastogi, Madhup
    Khurana, Rohini
    Hadi, Rahat
    Sapru, Shantanu
    Gandhi, Ajeet Kumar
    Mishra, S. P.
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2020, 26 (01): : 55 - 60
  • [47] Significant reduction of normal structure dose by proton radiotherapy (PT) compared with intensity modulated radiotherapy (IMRT) for unilateral treatment of tonsil carcinoma
    Parvathaneni, U.
    Yu, T. K.
    Riley, B.
    Mason, B. E.
    Woo, S. Y.
    Smith, A.
    Garden, A. S.
    Rosenthal, D. I.
    RADIOTHERAPY AND ONCOLOGY, 2007, 82 : S11 - S11
  • [48] Effect of Intensity Modulated Radiotherapy (IMRT) on the immunity, physical status and clinical effect of locally advanced NSCLC patients
    Xu, Jun-kai
    PAKISTAN JOURNAL OF MEDICAL SCIENCES, 2021, 37 (05) : 1480 - 1485
  • [49] Pulsed Reduced Dose-Rate Intensity Modulated Radiotherapy (IMRT) Delivery for Use in the High Dose Re-Irradiation Setting
    Price, R.
    Lin, T.
    Kuritzky, N.
    Ma, C.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [50] Intensity modulated radiation therapy (IMRT) surface dose measurements using a PTW advanced Markus chamber
    Chen, F. Q.
    Gupta, R.
    Metcalfe, P.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2010, 33 (01) : 23 - 34