Investigation of the influence of blade configuration on the hemodynamic performance and blood damage of the centrifugal blood pump

被引:26
|
作者
Li, Yuan [1 ]
Yu, Jiachen [2 ]
Wang, Hongyu [1 ]
Xi, Yifeng [1 ]
Deng, Xiaoyan [1 ]
Chen, Zengsheng [1 ]
Fan, Yubo [1 ]
机构
[1] Beihang Univ, Key Lab Biomech & Mechanobiol, Beijing Adv Innovat Ctr Biomed Engn, Sch Biol Sci & Med Engn,Minist Educ, Beijing 100083, Peoples R China
[2] Beihang Univ, Sch SinoFrench Engineer, Beijing, Peoples R China
基金
国家重点研发计划; 中国国家自然科学基金;
关键词
blade configuration; centrifugal blood pump; hemolysis; platelet activation state; pressure head; VENTRICULAR ASSIST DEVICE; FLUID-DYNAMICS ANALYSIS; PLATELET ACTIVATION; SHEAR-STRESS; HEMOLYSIS; GEOMETRY; DESIGN; MODELS;
D O I
10.1111/aor.14265
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose The design and optimization of centrifugal blood pumps are crucial for improved extracorporeal membrane oxygenation system performance. Secondary flow passages are common in centrifugal blood pumps, allowing for a high volume of unstable flow. Traditional design theory offers minimal guidance on the design and optimization of centrifugal blood pumps, so it's critical to understand how design parameter variables affect hydraulic performance and hemocompatibility. Methods Computational fluid dynamics (CFD) was employed to investigate the effects of blade number, blade wrap angle, blade thickness, and splitters on pressure head, hemolysis, and platelet activation state. Eulerian and Lagrangian features were used to analyze the flow fields and hemocompatibility metrics such as scalar shear stress, velocity distribution, and their correlation. Results The equalization of frictional and flow losses allow impellers with more blades and smaller wrap angles to have higher pressure heads, whereas the trade-off between the volume of high scalar shear stress and exposure time allows impellers with fewer blades and larger blade wrap angles to have a lower HI; there are configurations that increase the possibility of platelet activation for both number of blades and wrap angles. The hydraulic performance and hemocompatibility of centrifugal blood pumps are not affected by blade thickness. Compared to the main blades, splitters can improve the blood compatibility of a centrifugal blood pump with a small reduction in pressure head, but there is a trade-off between the length and location of the splitter that suppresses flow losses while reducing the velocity gradient. According to correlation analysis, pressure head, HI, and the volume of high shear stress were all substantially connected, and exposure time had a significant impact on HI. The platelet activation state was influenced by the average scalar shear stress and the volume of low velocity. Conclusion The findings reveal the impact of design variables on the performance of centrifugal blood pumps with secondary flow passages, as well as the relationship between hemocompatibility, hydraulic performance, and flow characteristics, and are useful for the design and optimization of this type of blood pump, as well as the prediction of clinical complications.
引用
收藏
页码:1817 / 1832
页数:16
相关论文
共 50 条
  • [21] Hemodynamic exercise response in calves with an implantable biventricular centrifugal blood pump
    Linneweber, J
    Nonaka, K
    Takano, T
    Kawahito, S
    Schulte-Eistrup, S
    Motomura, T
    Ichikawa, S
    Mikami, M
    Stevens, S
    Schima, H
    Wolner, E
    Nosé, Y
    ARTIFICIAL ORGANS, 2001, 25 (12) : 1018 - 1021
  • [22] Experimental investigation of the influence of the hydraulic performance of an axial blood pump on intraventricular blood flow
    Liu, Guang-Mao
    Jiang, Fu-Qing
    Yang, Xiao-Han
    Wei, Run-Jie
    Hu, Sheng-Shou
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2021, 44 (12): : 980 - 989
  • [23] Computational fluid dynamics prediction of blood damage in a centrifugal pump
    Song, XW
    Throckmorton, AL
    Wood, HG
    Antaki, JF
    Olsen, DB
    ARTIFICIAL ORGANS, 2003, 27 (10) : 938 - 941
  • [24] Centrifugal Pump Generates Superior Hemodynamic Performance Compared to a new Diagonal Blood Pump in Neonatal and Pediatric ECMO Circuits
    Patel, Krishna
    Undar, Akif
    WORLD JOURNAL FOR PEDIATRIC AND CONGENITAL HEART SURGERY, 2022, 13 (02) : 235 - 241
  • [25] The influence of the number of impeller's blade on the performance of multistage centrifugal pump
    Yi, Ling
    Li, Zhipeng
    Chen, Yanhui
    Hong, Shunjun
    ADVANCES IN CHEMICAL ENGINEERING III, PTS 1-4, 2013, 781-784 : 2851 - 2856
  • [26] Investigation on parametric design of centrifugal pump blade
    School of Energy and Power Engineering, Lanzhou University of Technology, Lanzhou, Gansu 730050, China
    Paiguan Jixie, 2009, 5 (310-313):
  • [27] Performance of a newly developed implantable centrifugal blood pump
    Tsukiya, T
    Taenaka, Y
    Tatsumi, E
    Takano, H
    ASAIO JOURNAL, 2001, 47 (05) : 559 - 562
  • [28] Centrifugal blood pump scaling
    Lorenz, M.
    Smith, W.
    ASAIO Journal, 2000, 46 (02)
  • [30] Influence of blade wrap angle on centrifugal pump performance by numerical and experimental study
    Lei Tan
    Baoshan Zhu
    Shuliang Cao
    Hao Bing
    Yuming Wang
    Chinese Journal of Mechanical Engineering, 2014, 27 : 171 - 177