Body-mounted robotic assistant for MRI-guided low back pain injection

被引:24
|
作者
Li, Gang [1 ]
Patel, Niravkumar A. [1 ]
Hagemeister, Jan [1 ]
Yan, Jiawen [1 ]
Wu, Di [1 ]
Sharma, Karun [2 ]
Cleary, Kevin [2 ]
Iordachita, Iulian [1 ]
机构
[1] Johns Hopkins Univ, LCSR, Baltimore, MD 21218 USA
[2] Childrens Natl Hlth Syst, Sheikh Zayed Inst Pediat Surg Innovat, Washington, DC USA
关键词
MRI-guided robot; Body-mounted robot; Pain injection; Image-guided therapy; FEASIBILITY; ACCURACY; DESIGN; BIOPSY; SYSTEM;
D O I
10.1007/s11548-019-02080-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose This paper presents the development of a body-mounted robotic assistant for magnetic resonance imaging (MRI)-guided low back pain injection. Our goal was to eliminate the radiation exposure of traditional X-ray guided procedures while enabling the exquisite image quality available under MRI. The robot is designed with a compact and lightweight profile that can be mounted directly on the patient's lower back via straps, thus minimizing the effect of patient motion by moving along with the patient. The robot was built with MR-conditional materials and actuated with piezoelectric motors so it can operate inside the MRI scanner bore during imaging and therefore streamline the clinical workflow by utilizing intraoperative MR images. Methods The robot is designed with a four degrees of freedom parallel mechanism, stacking two identical Cartesian stages, to align the needle under intraoperative MRI-guidance. The system targeting accuracy was first evaluated in free space with an optical tracking system, and further assessed with a phantom study under live MRI-guidance. Qualitative imaging quality evaluation was performed on a human volunteer to assess the image quality degradation caused by the robotic assistant. Results Free space positioning accuracy study demonstrated that the mean error of the tip position to be 0.51 +/- 0.27 mm and needle angle to be 0.70 degrees +/- 0.38 degrees. MRI-guided phantom study indicated the mean errors of the target to be 1.70 +/- 0.21 mm, entry point to be 1.53 +/- 0.19 mm, and needle angle to be 0.66 degrees +/- 0.43 degrees. Qualitative imaging quality evaluation validated that the image degradation caused by the robotic assistant in the lumbar spine anatomy is negligible. Conclusions The study demonstrates that the proposed body-mounted robotic system is able to perform MRI-guided low back injection in a phantom study with sufficient accuracy and with minimal visible image degradation that should not affect the procedure.
引用
收藏
页码:321 / 331
页数:11
相关论文
共 50 条
  • [31] Evaluation of a robotic technique for transrectal MRI-guided prostate biopsies
    Martijn G. Schouten
    Joyce G. R. Bomers
    Derya Yakar
    Henkjan Huisman
    Eva Rothgang
    Dennis Bosboom
    Tom W. J. Scheenen
    Sarthak Misra
    Jurgen J. Fütterer
    European Radiology, 2012, 22 : 476 - 483
  • [32] BILATERAL MRI-GUIDED STEREOTAXIC CINGULOTOMY FOR INTRACTABLE PAIN
    PILLAY, PK
    HASSENBUSCH, SJ
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1992, 59 (1-4) : 33 - 38
  • [33] Robotic System for MRI-Guided Shoulder Arthrography: Accuracy Evaluation
    Patel, Niravkumar A.
    Azimi, Ehsan
    Monfaredi, Reza
    Sharma, Karun
    Cleary, Kevin
    Iordachita, Iulian
    2018 INTERNATIONAL SYMPOSIUM ON MEDICAL ROBOTICS (ISMR), 2018,
  • [34] MRI-guided stereotactic thalamotomy for the treatment of neurogenic pain
    Balás, I
    Horváth, ZC
    Kövér, F
    Dóczi, TP
    11TH EUROPEAN CONGRESS OF NEUROSURGERY: EUROPEAN ASSOCIATION OF NEUROSURGICAL SOCIETIES (EANS), 1999, : 725 - 728
  • [35] Palliation of intractable cancer pain by MRI-guided cingulotomy
    Wong, ET
    Gunes, S
    Gaughan, E
    Patt, RB
    Ginsberg, LE
    Hassenbusch, SJ
    Payne, R
    CLINICAL JOURNAL OF PAIN, 1997, 13 (03): : 260 - 263
  • [36] A Prototype Body-Mounted MRI-Compatible Robot for Needle Guidance in Shoulder Arthrography
    Monfaredi, R.
    Seifabadi, R.
    Iordachita, I.
    Sze, R.
    Safdar, N. M.
    Sharma, K.
    Fricke, S.
    Krieger, A.
    Cleary, K.
    2014 5TH IEEE RAS & EMBS INTERNATIONAL CONFERENCE ON BIOMEDICAL ROBOTICS AND BIOMECHATRONICS (BIOROB), 2014, : 40 - 45
  • [37] MRI-guided focused ultrasound (MRgFUS) to treat facet joint osteoarthritis low back pain—case series of an innovative new technique
    Evan M. Weeks
    Michael W. Platt
    Wladyslaw Gedroyc
    European Radiology, 2012, 22 : 2822 - 2835
  • [38] Shoulder-Mounted Robot for MRI-Guided Arthrography: Accuracy and Mounting Study
    Monfaredi, R.
    Wilson, E.
    Sze, R.
    Sharma, K.
    Azizi, B.
    Iordachita, I.
    Cleary, K.
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 3643 - 3646
  • [39] MRI Distortion Correction and Robot-to-MRI Scanner Registration for an MRI-Guided Robotic System
    Tuna, E. Erdem
    Poirot, Nate Lombard
    Franson, Dominique
    Bayona, Juana Barrera
    Huang, Sherry
    Seiberlich, Nicole
    Griswold, Mark A.
    Cavusoglu, M. Cenk
    IEEE ACCESS, 2022, 10 : 99205 - 99220
  • [40] Shoulder-mounted Robot for MRI-Guided Arthrography: Clinically Optimized System
    Kim, Gyeong Hu
    Patel, Niravkumar
    Yan, Jiawen
    Wu, Di
    Li, Gang
    Cleary, Kevin
    Iordachita, Iulian
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 1977 - 1980