Ultrasensitive and high-throughput analysis of chlorophyll a in marine phytoplankton extracts using a fluorescence microplate reader

被引:4
|
作者
Mandalakis, Manolis [1 ]
Stravinskaite, Austeja [2 ]
Lagaria, Anna [3 ]
Psarra, Stella [3 ]
Polymenakou, Paraskevi [1 ]
机构
[1] Hellen Ctr Marine Res, Inst Marine Biol Biotechnol & Aquaculture, Iraklion 71500, Crete, Greece
[2] Vilnius Univ, Dept Chem, 24 Naugarduko, LT-03225 Vilnius, Lithuania
[3] Hellen Ctr Marine Res, Inst Oceanog, Iraklion 71500, Crete, Greece
基金
欧盟第七框架计划;
关键词
Phytoplankton biomass; Chlorophyll extracts; 96-well microplate; Rapid fluorometric assay; Seawater quality; OCEAN; HPLC;
D O I
10.1007/s00216-017-0392-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Chlorophyll a (Chl a) is the predominant pigment in every single photosynthesizing organism including phytoplankton and one of the most commonly measured water quality parameters. Various methods are available for Chl a analysis, but the majority of them are of limited throughput and require considerable effort and time from the operator. The present study describes a high-throughput, microplate-based fluorometric assay for rapid quantification of Chl a in phytoplankton extracts. Microplate sealing combined with ice cooling was proved an effective means for diminishing solvent evaporation during sample loading and minimized the analytical errors involved in Chl a measurements with a fluorescence microplate reader. A set of operating parameters (settling time, detector gain, sample volume) were also optimized to further improve the intensity and reproducibility of Chl a fluorescence signal. A quadratic regression model provided the best fit (r (2) = 0.9998) across the entire calibration range (0.05-240 pg mu L-1). The method offered excellent intra- and interday precision (% RSD 2.2 to 11.2%) and accuracy (% relative error -3.8 to 13.8%), while it presented particularly low limits of detection (0.044 pg mu L-1) and quantification (0.132 pg mu L-1). The present assay was successfully applied on marine phytoplankton extracts, and the overall results were consistent (average % relative error -14.8%) with Chl a concentrations (including divinyl Chl a) measured by high-performance liquid chromatography (HPLC). More importantly, the microplate-based method allowed the analysis of 96 samples/standards within a few minutes, instead of hours or days, when using a traditional cuvette-based fluorometer or an HPLC system.
引用
收藏
页码:4539 / 4549
页数:11
相关论文
共 50 条
  • [1] Ultrasensitive and high-throughput analysis of chlorophyll a in marine phytoplankton extracts using a fluorescence microplate reader
    Manolis Mandalakis
    Austėja Stravinskaitė
    Anna Lagaria
    Stella Psarra
    Paraskevi Polymenakou
    Analytical and Bioanalytical Chemistry, 2017, 409 : 4539 - 4549
  • [2] High-Throughput Chemotyping of Cannabis and Hemp Extracts Using an Ultraviolet Microplate Reader and Multivariate Classifiers
    Chen Z.
    de Boves Harrington P.
    Baugh S.F.
    Journal of Analysis and Testing, 2018, 2 (03) : 210 - 222
  • [3] High throughput Time Resolved Fluorescence in a Microplate Reader
    Petersen, Karl J.
    Muretta, Joseph M.
    Higgins, Sutton E.
    Peterson, Kurt C.
    Gillispie, Gregory D.
    Thomas, David D.
    BIOPHYSICAL JOURNAL, 2014, 106 (02) : 205A - 206A
  • [4] Fluorescence Lifetime Microplate Reader for Structural Biology: High Performance and High-Throughput in the Same Tool
    Gillispie, Gregory D.
    Peterson, Kurt C.
    Li, Ji
    Thomas, David D.
    BIOPHYSICAL JOURNAL, 2013, 104 (02) : 345A - 346A
  • [5] A High-Throughput Fluorometric Method for Quantitation of Besifloxacin Hydrochloride Using Microplate Reader
    Abd El-Hay, Soad S.
    El-Sayed, Heba M.
    CURRENT PHARMACEUTICAL ANALYSIS, 2018, 14 (03) : 198 - 203
  • [6] High-throughput measurements of the optical redox ratio using a commercial microplate reader
    Cannon, Taylor M.
    Shah, Amy T.
    Walsh, Alex J.
    Skala, Melissa C.
    JOURNAL OF BIOMEDICAL OPTICS, 2015, 20 (01)
  • [7] A Simple High-Throughput Method for Determination of Antiepileptic Analogues of γ-Aminobutyric Acid in Pharmaceutical Dosage Forms Using Microplate Fluorescence Reader
    Martinc, Bostjan
    Vovk, Tomaz
    CHEMICAL & PHARMACEUTICAL BULLETIN, 2013, 61 (10) : 1009 - 1014
  • [8] Enhanced Reproducibility and Precision of High-Throughput Quantification of Bacterial Growth Data Using a Microplate Reader
    Abkar, Leili
    Wilfart, Florentin M.
    Piercey, Marta
    Gagnon, Graham A.
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2022, (185):
  • [9] An Improved Automated High-Throughput Efficient Microplate Reader for Rapid Colorimetric Biosensing
    Yang, Jinhu
    Wu, Yue
    Wang, Hao
    Yang, Wenjian
    Xu, Zhongyuan
    Liu, Dong
    Chen, Hui-Jiuan
    Zhang, Diming
    BIOSENSORS-BASEL, 2022, 12 (05):
  • [10] Development of a micelle-enhanced high-throughput fluorometric method for determination of niclosamide using a microplate reader
    Abd El-Hay, Soad S.
    Belal, Fathalla F.