Open-Ended Coaxial Probe Technique for Dielectric Measurement of Biological Tissues: Challenges and Common Practices

被引:180
|
作者
La Gioia, Alessandra [1 ]
Porter, Emily [1 ]
Merunka, Ilja [2 ]
Shahzad, Atif [1 ]
Salahuddin, Saqib [1 ]
Jones, Marggie [1 ]
O'Halloran, Martin [1 ]
机构
[1] Natl Univ Ireland Galway, Dept Elect & Elect Engn, Galway, Ireland
[2] Czech Tech Univ, Dept Electromagnet Field, Prague 16627, Czech Republic
来源
DIAGNOSTICS | 2018年 / 8卷 / 02期
基金
爱尔兰科学基金会; 欧洲研究理事会;
关键词
dielectric measurements; biological tissues; open-ended coaxial probe; equipment-related confounders; tissue-related confounders; RADIO-FREQUENCY PERMITTIVITY; MICROWAVE THERMAL ABLATION; COMPLEX PERMITTIVITY; IN-VIVO; ELECTRICAL-PROPERTIES; BREAST-TISSUE; RADIOFREQUENCY ABLATION; ANIMAL-TISSUES; LIVER-TISSUE; LARGE-SCALE;
D O I
10.3390/diagnostics8020040
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Electromagnetic (EM) medical technologies are rapidly expanding worldwide for both diagnostics and therapeutics. As these technologies are low-cost and minimally invasive, they have been the focus of significant research efforts in recent years. Such technologies are often based on the assumption that there is a contrast in the dielectric properties of different tissue types or that the properties of particular tissues fall within a defined range. Thus, accurate knowledge of the dielectric properties of biological tissues is fundamental to EM medical technologies. Over the past decades, numerous studies were conducted to expand the dielectric repository of biological tissues. However, dielectric data is not yet available for every tissue type and at every temperature and frequency. For this reason, dielectric measurements may be performed by researchers who are not specialists in the acquisition of tissue dielectric properties. To this end, this paper reviews the tissue dielectric measurement process performed with an open-ended coaxial probe. Given the high number of factors, including equipment- and tissue-related confounders, that can increase the measurement uncertainty or introduce errors into the tissue dielectric data, this work discusses each step of the coaxial probe measurement procedure, highlighting common practices, challenges, and techniques for controlling and compensating for confounders.
引用
收藏
页数:38
相关论文
共 50 条
  • [11] Dielectric Properties for Identification of Gliomas and Normal Brain Tissues with Open-Ended Coaxial Probe
    Xu, Guofang
    Dai, Xingliang
    Yu, Xuefei
    Nan, Xiang
    Han, Jijun
    12TH ASIAN-PACIFIC CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING, VOL 1, APCMBE 2023, 2024, 103 : 63 - 69
  • [12] An application of the virtual transmission line model of an open-ended coaxial probe for dielectric properties characterization of biological tissues
    Fontana, Nunzia
    Canicatti, Eliana
    Monorchio, Agostino
    2019 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND USNC-URSI RADIO SCIENCE MEETING, 2019, : 341 - 342
  • [13] Sensing Radius of an Open-Ended Coaxial Probe for Dielectric Measurement of Tissues: Calculation with Relative Permittivity versus Conductivity
    La Gioia, Alessandra
    O'Halloran, Martin
    Porter, Emily
    2019 URSI ASIA-PACIFIC RADIO SCIENCE CONFERENCE (AP-RASC), 2019,
  • [14] IMPROVEMENT IN DIELECTRIC MEASUREMENT TECHNIQUE OF OPEN-ENDED COAXIAL LINE RESONATOR METHOD
    LIU, LP
    XU, DM
    JIANG, ZY
    ELECTRONICS LETTERS, 1986, 22 (07) : 373 - 375
  • [15] Examination of the Sensing Radius of Open-ended Coaxial Probes in Dielectric Measurements of Biological Tissues
    La Gioia, Alessandra
    Porter, Emily
    O'Halloran, Martin
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 215 - 216
  • [16] Open-Ended Coaxial Probe Technique for Dielectric Spectroscopy of Artificially Grown Sea Ice
    Komarov, Sergey A.
    Komarov, Alexander S.
    Barber, David G.
    Lemes, Marcos J. L.
    Rysgaard, Soren
    IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 2016, 54 (08): : 4941 - 4951
  • [17] Design and Calibration of a Large Open-Ended Coaxial Probe for the Measurement of the Dielectric Properties of Concrete
    Filali, Bilal
    Boone, Francois
    Rhazi, Jamal
    Ballivy, Gerard
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2008, 56 (10) : 2322 - 2328
  • [18] Open-Ended Coaxial Probe for Effective Reconstruction of Biopsy-Excised Tissues' Dielectric Properties
    Canicatti, Eliana
    Fontana, Nunzia
    Barmada, Sami
    Monorchio, Agostino
    SENSORS, 2024, 24 (07)
  • [19] Characterization of Open-Ended Coaxial Probe Sensing Depth with Respect to Aperture Size for Dielectric Property Measurement of Heterogeneous Tissues
    Aydinalp, Cemanur
    Joof, Sulayman
    Dilman, Ismail
    Akduman, Ibrahim
    Yilmaz, Tuba
    SENSORS, 2022, 22 (03)
  • [20] ON THE MEASUREMENT OF THE COMPLEX PERMITTIVITY OF MATERIALS BY AN OPEN-ENDED COAXIAL PROBE
    MISRA, D
    IEEE MICROWAVE AND GUIDED WAVE LETTERS, 1995, 5 (05): : 161 - 163