An in vitro verification of strength estimation for moving an 125I source during implantation in brachytherapy

被引:3
|
作者
Tanaka, Kenichi [1 ]
Kajimoto, Tsuyoshi [1 ]
Hayashi, Takahiro [1 ]
Asanuma, Osamu [2 ]
Hori, Masakazu [3 ]
Kamo, Ken-ichi [3 ]
Sumida, Iori [4 ]
Takahashi, Yutaka [4 ]
Tateoka, Kunihiko [3 ]
Bengua, Gerard [5 ]
Sakata, Koh-ichi [3 ]
Endo, Satoru [1 ]
机构
[1] Hiroshima Univ, Grad Sch Engn, Quantum Energy Applicat, 1-4-1 Kagamiyama, Higashihiroshima, Hiroshima 7398527, Japan
[2] Sapporo Med Univ Hosp, Div Radiol & Nucl Med, Chuo Ward, South 1,West 17, Sapporo, Hokkaido 0608556, Japan
[3] Sapporo Med Univ, Grad Sch Med, Dept Radiat Oncol, Chuo Ward, South 1,West 16, Sapporo, Hokkaido 0608543, Japan
[4] Osaka Univ, Grad Sch Med, Dept Radiol, 2-2 Yamada Oka, Suita, Osaka 5650871, Japan
[5] Auckland City Hosp, Pk Rd, Auckland 1142, New Zealand
基金
日本学术振兴会; 日本科学技术振兴机构;
关键词
brachytherapy; I-125; source strength; moving source; verification; INTERSTITIAL BRACHYTHERAPY; DOSIMETRY;
D O I
10.1093/jrr/rry021
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
This study aims to demonstrate the feasibility of a method for estimating the strength of a moving brachytherapy source during implantation in a patient. Experiments were performed under the same conditions as in the actual treatment, except for one point that the source was not implanted into a patient. The brachytherapy source selected for this study was I-125 with an air kerma strength of 0.332 U (mu Gym(2)h(-1)), and the detector used was a plastic scintillator with dimensions of 10 cm x 5 cm x 5 cm. A calibration factor to convert the counting rate of the detector to the source strength was measured and then the accuracy of the proposed method was investigated for a manually driven source. The accuracy was found to be under 10% when the shielding effect of additional needles for implantation at other positions was corrected, and about 30% when the shielding was not corrected. Even without shielding correction, the proposed method can detect dead/dropped source, implantation of a source with the wrong strength, and a mistake in the number of the sources implanted. Furthermore, when the correction was applied, the achieved accuracy came close to within 7% required to find the Oncoseed 6711 (I-125 seed with unintended strength among the commercially supplied values of 0.392, 0.462 and 0.533 U).
引用
收藏
页码:484 / 489
页数:6
相关论文
共 50 条
  • [1] Strength estimation of a moving 125Iodine source during implantation in brachytherapy: application to linked sources
    Tanaka, Kenichi
    Endo, Satoru
    Tateoka, Kunihiko
    Asanuma, Osamu
    Hori, Masakazu
    Takagi, Masaru
    Bengua, Gerard
    Kamo, Ken-ichi
    Sato, Kaori
    Takeda, Hiromitsu
    Hareyama, Masato
    Sakata, Koh-ichi
    Takada, Jun
    JOURNAL OF RADIATION RESEARCH, 2014, 55 (06) : 1146 - 1152
  • [2] Dosimetric characteristics of a new 125I brachytherapy source
    Gearheart, DM
    Drogin, A
    Sowards, K
    Meigooni, AS
    Ibbott, GS
    MEDICAL PHYSICS, 2000, 27 (10) : 2278 - 2285
  • [3] DOSIMETRIC CHARACTERISTIC OF A NEW 125I BRACHYTHERAPY SOURCE
    Sadeghi, Mahdi
    Khanmohammadi, Zahra
    RADIATION PROTECTION DOSIMETRY, 2011, 147 (03) : 451 - 456
  • [4] DETERMINATION OF THE DOSIMETRIC CHARACTERISTICS OF IrSeED 125I BRACHYTHERAPY SOURCE
    Lohrabian, V.
    Ghozati, B.
    Aghamiri, M.
    Pourbeigi, H.
    Sheibani, S.
    INTERNAL MEDICINE JOURNAL, 2011, 41 : 28 - 28
  • [5] Determination of dosimetric characteristics of irseed 125I brachytherapy source
    Lohrabian, Vahid
    Sheibani, Shahab
    Aghamiri, Mahmoud Reza
    Ghozati, Behroz
    Pourbeigi, Hosein
    Baghani, Hamid Reza
    Iranian Journal of Medical Physics, 2013, 10 (03) : 109 - 117
  • [6] Model 3500 125I brachytherapy source dosimetric characterization
    Wallace, RE
    APPLIED RADIATION AND ISOTOPES, 2002, 56 (04) : 581 - 587
  • [7] 125I brachytherapy seeds implantation for an inoperable large retroperitoneal leiomyosarcoma
    Di, Xuemin
    Liang, Yansong
    Yu, Huimin
    Zhao, Jinxin
    Wang, Zeyang
    Zhao, Jing
    Sui, Aixia
    Zhang, Hongtao
    Wang, Juan
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2020, 16 (02) : 397 - 399
  • [8] Prostate brachytherapy (LDR) with 125I permanent implantation;: Experience at CROB
    Cocco, G.
    Iadanza, L.
    Zucca, S.
    Castaldo, G.
    Lapadula, L.
    Montagna, A.
    Mazziotta, M.
    Imbriani, E.
    Silvestre, G.
    Fusco, V.
    ANNALS OF ONCOLOGY, 2006, 17 : VII153 - VII153
  • [9] Measurement of the strength of iodine-125 seed moving at unknown speed during implantation in brachytherapy
    Tanaka, Kenichi
    Endo, Satoru
    Tateoka, Kunihiko
    Asanuma, Osamu
    Kamo, Ken-ichi
    Sato, Kaori
    Takeda, Hiromitsu
    Takagi, Masaru
    Hareyama, Masato
    Takada, Jun
    JOURNAL OF RADIATION RESEARCH, 2014, 55 (01) : 162 - 167
  • [10] Verification of I-125 brachytherapy source strength for use in radioactive seed localization procedures
    Metyko, John
    Erwin, William
    Landsberger, Sheldon
    APPLIED RADIATION AND ISOTOPES, 2016, 112 : 62 - 68