Validation of the RayStation Monte Carlo dose calculation algorithm using realistic animal tissue phantoms

被引:26
|
作者
Schreuder, Andries N. [1 ]
Bridges, Daniel S. [1 ]
Rigsby, Lauren [1 ]
Blakey, Marc [1 ]
Janson, Martin [2 ]
Hedrick, Samantha G. [1 ]
Wilkinson, John B. [1 ]
机构
[1] Provis Ctr Proton Therapy Knoxville, Knoxville, TN 37909 USA
[2] PhDRaySearch Labs, Stockholm, Sweden
来源
关键词
analytical dose algorithms; Monte Carlo; pencil beam scanning; proton therapy; spot scanning; PENCIL BEAM ALGORITHM; CALCULATION ACCURACY; THERAPY;
D O I
10.1002/acm2.12733
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The aim of this study is to validate the RayStation Monte Carlo (MC) dose algorithm using animal tissue neck phantoms and a water breast phantom. Methods Three anthropomorphic phantoms were used in a clinical setting to test the RayStation MC dose algorithm. We used two real animal necks that were cut to a workable shape while frozen and then thawed before being CT scanned. Secondly, we made a patient breast phantom using a breast prosthesis filled with water and placed on a flat surface. Dose distributions in the animal and breast phantoms were measured using the MatriXX PT device. Results The measured doses to the neck and breast phantoms compared exceptionally well with doses calculated by the analytical pencil beam (APB) and MC algorithms. The comparisons between APB and MC dose calculations and MatriXX PT measurements yielded an average depth difference for best gamma agreement of <1 mm for the neck phantoms. For the breast phantom better average gamma pass rates between measured and calculated dose distributions were observed for the MC than for the APB algorithms. Conclusions The MC dose calculations are more accurate than the APB calculations for the static phantoms conditions we evaluated, especially in areas where significant inhomogeneous interfaces are traversed by the beam.
引用
收藏
页码:160 / 171
页数:12
相关论文
共 50 条
  • [41] Evaluation of Collapsed Cone Dose Calculation Algorithm for HDR Brachytherapy Using Monte Carlo
    Volken, W.
    Terribilini, D.
    Fix, M. K.
    Loessl, K.
    van Veelen, B.
    Manser, P.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [42] Monte Carlo modeling of multilayer phantoms with multiple fluorophores: simulation algorithm and experimental validation
    Pery, Emilie
    Blondel, Walter C. P. M.
    Thomas, Cedric
    Guillemin, Francois
    JOURNAL OF BIOMEDICAL OPTICS, 2009, 14 (02)
  • [43] Validation of a Monte Carlo dose calculation tool for radiotherapy treatment planning
    Li, JS
    Pawlicki, T
    Deng, J
    Jiang, SB
    Mok, E
    Ma, CM
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (10): : 2969 - 2985
  • [44] Validation of DPM Monte Carlo code for internal beta dose calculation
    Owczarczyk, J.
    Owczarczyk, K.
    Recko, W.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S322 - S322
  • [45] Clinical implementation and validation of a commercial Monte Carlo dose calculation system
    Chuang, C
    Curran, B
    Verhey, L
    MEDICAL PHYSICS, 2003, 30 (06) : 1453 - 1453
  • [46] Validation of fast Monte Carlo dose calculation in small animal radiotherapy with EBT3 radiochromic films
    Noblet, C.
    Chiavassa, S.
    Smekens, F.
    Sarrut, D.
    Passal, V.
    Suhard, J.
    Lisbona, A.
    Paris, F.
    Delpon, G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (09): : 3521 - +
  • [47] Evaluation of Inhomogeneity Calculations for Electron Beams in Raystation Monte Carlo Algorithm
    Sansourekidou, P.
    Allen, C.
    Pavord, D.
    MEDICAL PHYSICS, 2014, 41 (06) : 260 - 260
  • [48] A virtual-accelerator-based verification of a Monte Carlo dose calculation algorithm for electron beam treatment planning in homogeneous phantoms
    Wieslander, E
    Knöös, T
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (06): : 1533 - 1544
  • [49] Monte Carlo dose calculations for phantoms with hip prostheses
    Bazalova, M.
    Coolens, C.
    Cury, F.
    Childs, P.
    Beaulieu, L.
    Verhaegen, F.
    INTERNATIONAL WORKSHOP ON MONTE CARLO TECHNIQUES IN RADIOTHERAPY DELIVERY AND VERIFICATION - THIRD MCGILL INTERNATIONAL WORKSHOP, 2008, 102
  • [50] A Monte Carlo dose calculation system for ophthalmic brachytherapy based on a realistic eye model
    Miras del Rio, Hector
    Ortiz Lora, Antonio
    Bertolet Reina, Alejandro
    Terron Leon, Jose Antonio