A design of transmission-type multi-target X-ray tube based on electric field modulation

被引:5
|
作者
Zhao, Lei [1 ]
Jia, Wenbao [1 ,3 ]
Jin, Limin [1 ]
Shan, Qing [1 ,3 ]
Cheng, Can [1 ]
Zhu, Hongkui [1 ]
Hei, Daqian [2 ]
机构
[1] Nanjing Univ Aeronaut & Astronaut, Coll Mat Sci & Technol, Nanjing 211106, Peoples R China
[2] Lanzhou Univ, Sch Nucl Sci & Technol, Tianshui South Rd 222, Lanzhou 730000, Peoples R China
[3] Jiangsu Higher Educ Inst, Collaborat Innovat Ctr Radiat Med, Suzhou 215000, Peoples R China
关键词
X-ray tube; Multi-target; Electric field modulation; Electron beam trajectory; Optimal thickness; TARGET PARAMETERS; GUN;
D O I
10.1016/j.net.2021.03.002
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Multi-target X-ray tube is a new type X-ray source, and can be applied in many fields such as sensitive X-ray fluorescence analysis and medical imaging. In this work, we report an electric field modulation multi-target X-ray tube, which contains four targets (Cr, Ni, Au, Mo) coated on a Beryllium (Be) window. A four-valve electric field deflector was developed to deflect the electron beam to bombard the corresponding targets. Particle dynamics analysis software was employed to simulate the particle tracking of electron beam. The results show that the 30 keV electron beam could get a 6.7 mm displacement on the target plane by 10(5) V/m electric field. The focus areas are about 2 mm x 5 mm and 4 mm x 2.5 mm after deflection in two directions. Thermal behavior calculated by ANSYS shows that the designed target assembly could withstand a 10 W continuous power. The optimum target thicknesses and emission spectra were obtained by Geant4 when the thickness of Be window was 300 mu m and the electron beam incident angle was 0.141 rad. The results indicate that this multi-target X-ray tube could provide different X-ray sources effectively. (C) 2021 Korean Nuclear Society, Published by Elsevier Korea LLC.
引用
收藏
页码:3026 / 3034
页数:9
相关论文
共 50 条
  • [41] A novel X-ray tube spectra reconstruction method based on transmission measurements
    Cao, Jing
    Jiang, Chun-Yu
    Zhao, Yan-Feng
    Yang, Qing-Wei
    Yin, Ze-Jie
    NUCLEAR SCIENCE AND TECHNIQUES, 2016, 27 (02)
  • [42] Low-voltage high-curent field emission from a simple pointed graphite emitter tested in a transmission-type X-ray demonstator
    Klaes, J.
    Luetzenkirchen-Hecht, D.
    Serbun, P.
    Mueller, G.
    2015 28TH INTERNATIONAL VACUUM NANOELECTRONICS CONFERENCE (IVNC), 2015, : 186 - 187
  • [43] Dual and Multi-Target Cone-Beam X-ray Luminescence Computed Tomography Based on the DeepCB-XLCT Network
    Liu, Tianshuai
    Huang, Shien
    Li, Ruijing
    Gao, Peng
    Li, Wangyang
    Lu, Hongbing
    Song, Yonghong
    Rong, Junyan
    BIOENGINEERING-BASEL, 2024, 11 (09):
  • [44] Compact X-ray tube based on carbon nanotube cold cathode emitter for handheld type X-ray application
    Lim, Jongmin
    Gupta, Amar P.
    Choi, Jinho
    Rajpoot, Shalini
    Jung, Jaeik
    Jang, Jaekyu
    Yeo, Seung Jun
    Ahn, Jeung Sun
    Ryu, Jehwang
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [45] Behavior of characteristic X-rays from a partial-transmission-type X-ray target
    Raza, Hamid Saeed
    Kim, Hyun Jin
    Ha, Jun Mok
    Cho, Sung Oh
    APPLIED RADIATION AND ISOTOPES, 2013, 80 : 67 - 72
  • [46] Determination of molybdenum target parameters for transmission X-ray tube: A Geant4 simulation study
    Mohammad Mehdi Nasseri
    Pramana, 2019, 92
  • [47] Determination of molybdenum target parameters for transmission X-ray tube: A Geant4 simulation study
    Nasseri, Mohammad Mehdi
    PRAMANA-JOURNAL OF PHYSICS, 2019, 92 (04):
  • [48] Design and Modulation Performance of Ultrafast Light-controlled Pulsed X-ray Tube with Photocathode
    Xuan Hao
    Liu Yongan
    Qiang Pengfei
    Su Tong
    Yang Xianghui
    Sheng Lizhi
    Zhao Baosheng
    ACTA PHOTONICA SINICA, 2021, 50 (07)
  • [49] Design and Characteristics of Microfocus X-ray Source with Sealed Tube and Transmissive Target on Diamond Window
    Tsunoda, Takaaki
    Tsukamoto, Takeo
    Ando, Yoichi
    Hamamoto, Yasuhiro
    Ikarashi, Yoichi
    Nagasato, Satoshi
    Ueda, Kazuyuki
    MATERIALS EVALUATION, 2021, 79 (06) : 631 - 640
  • [50] Design optimization of X-ray security scanner based on dual-energy transmission imaging with variable tube voltage
    Park, Junsung
    Ana, Geunyoung
    Seo, Hee
    JOURNAL OF INSTRUMENTATION, 2023, 18 (01):