Analysis by EPR measurements and spectral deconvolution of the dosimetric properties of lithium formate monohydrate

被引:7
|
作者
Belahmar, Ayoub [1 ]
Mikou, Mohammed [1 ]
El Ghalmi, Mohammed [1 ,2 ]
机构
[1] Univ Hassan I, Lab Anal Syst & Informat Proc, Fac Sci & Technol, BP 577, Settat 26000, Morocco
[2] Al Kindy Oncol Ctr, 2-4 Rue Abou Youssef El Kindy, Casablanca, Morocco
关键词
Deconvolution; Dosimetry; Electron paramagnetic resonance; Electrons; Lithium formate monohydrate; X-ray; ALANINE/EPR DOSIMETRY; TABLE-SUGAR; ESR; RADIATION; BEAMS; RADIOTHERAPY; RADICALS; SYSTEM; ENDOR;
D O I
10.1016/j.nimb.2018.06.028
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this study we analyze the Electron Paramagnetic Resonance "EPR" response of lithium formate monohydrate "LFM" irradiated by photons and electrons using a linear accelerator. EPR measurements were performed one day after irradiation and regularly during six months of storage. Both dosimetric curves using peak to peak technique and double integration method showed linear behavior. The time dependence indicates good stability of free radicals produced on this material. The determined measurable threshold dose is 0.55 Gy. For used energies, LFM dosimeters are more sensitive to X-ray irradiation than to electrons. The deconvolution of the EPR signal of irradiated LFM using Levenberg-Marquardt (M-L) algorithm shows that this spectrum is constituting of three gaussian contributions. Each contribution has a linear evolution according to the irradiation dose, as well as an acceptable degree of stability during the storage period, approving the stability of all free radicals produced in LFM. All these characteristics confirm that LFM might be a promising dosimeter to be used in the 'clinical' dose range in radiotherapy.
引用
收藏
页码:19 / 24