A protocol for patients with cardiovascular implantable devices undergoing magnetic resonance imaging (MRI): should defibrillation threshold testing be performed post-(MRI)

被引:43
|
作者
Burke, Peter Thomas [1 ]
Ghanbari, Hamid [1 ]
Alexander, Patrick B. [1 ]
Shaw, Michael K. [1 ]
Daccarett, Marcos [1 ]
Machado, Christian [1 ]
机构
[1] Providence Hosp, Providence Heart Inst, Southfield, MI 48075 USA
关键词
Magnetic resonance imaging; MRI; Cardiac implantable electronic device; CIED; Pacemaker; ICD; Defibrillator threshold testing; Defibrillation safety margin testing; DSMT; CARDIOVERTER-DEFIBRILLATORS; CARDIAC-PACEMAKERS; 1.5; TESLA; VENTRICULAR-FIBRILLATION; PERMANENT PACEMAKERS; SCIENTIFIC STATEMENT; IN-VITRO; SAFETY; CONVERSION; 1.5-TESLA;
D O I
10.1007/s10840-009-9463-y
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Magnetic resonance imaging (MRI) in patients with Cardiovascular Implantable Electronic Devices (CIED) has not been approved by the Food and Drug Administration. Recent data suggests MRI as a relative rather than absolute contraindication in CIED patients. Recently, the American Heart Association has recommended defibrillation threshold testing (DFTT) in implantable cardioverter defibrillator (ICD) patients undergoing MRI. We evaluated the feasibility and safety of a protocol for MRI in CIED patients, incorporating the new recommendations on DFTT. Consecutive patients with CIED undergoing MRI were included. The protocol consisted of continuous monitoring during imaging, device interrogation pre- and post-MRI, reprogramming of the pacemaker to an asynchronous mode in pacemaker-dependent (PMD) patients and a non-tracking/sensing mode for non-PMD patients. All tachyarrhythmia therapies were disabled. Devices were interrogated for lead impedance, battery life, pacing, and sensing thresholds. All patients with ICD underwent DFTT/defibrillator safety margin testing (DSMT) post-MRI. A total of 92 MRI's at 1.5 Tesla were performed in 38 patients. A total of 13 PMD patients, ten ICD patients, four cardiac resynchronization therapy with defibrillator (CRT-D) patients, and 11 non-PMD patients were scanned from four major manufacturers. No device circuitry damage, programming alterations, inappropriate shocks, failure to pace, or changes in sensing, pacing, or defibrillator thresholds were found on single or multiple MRI sessions. Our protocol for MRI in CIED patients appears safe, feasible, and reproducible. This is irrespective of the type of CIED, pacemaker dependancy or multiple 24-h scanning sessions. Our protocol addresses early detection of potential complications and establishes a response system for potential device-related complications. Our observation suggests that routine DFTT/DSMT post-MRI may not be necessary.
引用
收藏
页码:59 / 66
页数:8
相关论文
共 50 条
  • [21] Laryngeal mask airway Unique position in paediatric patients undergoing magnetic resonance imaging (MRI): prospective observational study
    Klucka, Jozef
    Senkyrik, Jan
    Skotakova, Jarmila
    Stoudek, Roman
    Toukalkova, Michaela
    Krikava, Ivo
    Marecek, Lukas
    Pavlik, Tomas
    Stouracova, Alena
    Stourac, Petr
    BMC ANESTHESIOLOGY, 2018, 18
  • [22] Racial and socioeconomic disparities in staging magnetic resonance imaging (MRI) utilization for prostate cancer (PCa) patients undergoing radiotherapy
    Ajayi, Ayobami
    Chapman, Christina
    Both, Stefan
    Deville, Curtiland
    CANCER EPIDEMIOLOGY BIOMARKERS & PREVENTION, 2016, 25 (03)
  • [23] Laryngeal mask airway Unique™ position in paediatric patients undergoing magnetic resonance imaging (MRI): prospective observational study
    Jozef Klučka
    Jan Šenkyřík
    Jarmila Skotáková
    Roman Štoudek
    Michaela Ťoukalková
    Ivo Křikava
    Lukáš Mareček
    Tomáš Pavlík
    Alena Štouračová
    Petr Štourač
    BMC Anesthesiology, 18
  • [24] Safe scanning, but frequent artifacts mimicking bradycardia and tachycardia during magnetic resonance imaging (MRI) in patients with an Implantable loop recorder (ILR)
    Gimbel, JR
    Zarghami, J
    Machado, C
    Wilkoff, BL
    ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2005, 10 (04) : 404 - 408
  • [25] Magnetic Resonance and Computed Tomography Imaging in Patients with Cardiovascular Implantable Electronic Devices: Analysis of Expert Consensus Data and Implications for the Perioperative Clinician
    Stawiarski, Kristin
    Sorajja, Dan
    Ramakrishna, Harish
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2018, 32 (06) : 2817 - 2822
  • [26] Temperature Management for Paediatric Patients undergoing outpatient (OP) Magnetic Resonance Imaging (MRI) scans under General Anaesthetics (GA)
    Luo, Guanmei
    Bew, S.
    ANESTHESIA AND ANALGESIA, 2021, 132 (5S_SUPPL): : 738 - 738
  • [27] Determinants of vascular stiffness in patients with advanced cardiovascular disease: Comparison of four different methods including magnetic resonance imaging (MRI)
    Delles, C.
    Zimmerli, L. U.
    Berg, G. A.
    Steedman, T.
    Dargie, H. J.
    Dominiczak, A. F.
    JOURNAL OF HYPERTENSION, 2006, 24 : S74 - S74
  • [28] Improving safety in anesthetized patients undergoing magnetic resonance imaging-Concept of time-out in the MRI suite and measures to improve adherence to time-out protocol
    Joselyn, Anita Shirley
    Teddy, Dan Anna
    Jose, Riya
    Shanthini, Sharon
    Joseph, Lallu
    Rai, Ekta
    Williams, Aparna
    PEDIATRIC ANESTHESIA, 2023, 33 (05) : 370 - 376
  • [29] Deactivation vs. asynchronous pacing - prospective evaluation of a protocol for rhythm management in patients with magnetic resonance conditional pacemakers undergoing adenosine stress cardiovascular magnetic resonance imaging
    Oliver Klein-Wiele
    Marietta Garmer
    Gianluca Barbone
    Rhyan Urbien
    Martin Busch
    Kaffer Kara
    Harald Schäfer
    Michael Schulte-Hermes
    Birgit Hailer
    Dietrich Grönemeyer
    BMC Cardiovascular Disorders, 17
  • [30] Deactivation vs. asynchronous pacing - prospective evaluation of a protocol for rhythm management in patients with magnetic resonance conditional pacemakers undergoing adenosine stress cardiovascular magnetic resonance imaging
    Klein-Wiele, Oliver
    Garmer, Marietta
    Barbone, Gianluca
    Urbien, Rhyan
    Busch, Martin
    Kara, Kaffer
    Schaefer, Harald
    Schulte-Hermes, Michael
    Hailer, Birgit
    Groenemeyer, Dietrich
    BMC CARDIOVASCULAR DISORDERS, 2017, 17