Microfluidic Immunoassay System for Rapid Detection and Semi-Quantitative Determination of a Potential Serum Biomarker Mesothelin

被引:18
|
作者
Duan, Xiaoxiao [1 ]
Zhao, Linlin [2 ]
Dong, Heng [1 ]
Zhao, Wang [1 ]
Liu, Sixiu [1 ]
Sui, Guodong [1 ]
机构
[1] Fudan Univ, Shanghai Key Lab Atmospher Particle Pollut Preven, Dept Environm Sci & Engn, 220 Handan Rd, Shanghai 200433, Peoples R China
[2] Second Mil Med Univ, Changhai Hosp, Dept Gastroenterol, 168 Changhai Rd, Shanghai 200433, Peoples R China
基金
中国国家自然科学基金;
关键词
MIAS; pancreatic cancer; MSLN; rapid detection; clinical serum samples; PANCREATIC-CANCER; DIAGNOSTIC-VALUE; MARKER; ADENOCARCINOMAS; CHIP;
D O I
10.1021/acssensors.9b01430
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Mesothelin (MSLN) is considered as a potential serological tumor biomarker for early diagnosis of pancreatic cancer. Nevertheless, low sensibility, high reagent consumption, and time costs of traditional detection methods limit their utility in clinical disease diagnoses. Here, we combined the immunoassay technique with microfluidic chips to develop a microfluidic immunoassay system (MIAS) that can be used for rapid semi-quantitative detection of serum MSLN levels. The MIAS was composed of 12 uniform structures, including 12 inlets, 12 reaction chambers, and one outlet allowing measurement of four samples with three repeats, simultaneously. A unique microarray cylinder was located at the end of each reaction chamber where immunoassay was performed to trap microspheres. A feasible interception efficiency (similar to 80%) was attained, with microspheres filling the reaction column. It has been demonstrated that fluorescence intensity is proportional to the MSLN concentration on the MIAS (R-2 = 0.95). Subsequently, 16 clinical serum samples collected from Changhai Hospital, Shanghai were selected from eight patients with pancreatic cancers, four with pancreatitis, and four with other digestive system diseases (2 gastric cancer, 1 bile duct stricture, and 1 bile duct stones). MSLN levels for these samples were detected via MIAS. The results showed a significant correlation between MIAS and traditional enzymelinked immunosorbent assay (ELISA), with the correlation coefficient, 0.93. The detection limit of MSLN fluorescence intensity and concentration was similar to 6a.u. and similar to 20 pg/mL, respectively. The entire duration of analysis by MIAS decreased to similar to 40 min compared to 2 h by ELISA. Statistical analysis of MIAS data revealed that MSLN was overexpressed in pancreatic cancer than in the others (P value =, 0.0014). Moreover, the diagnostic accuracies of MSLN detected by MIAS and CA19-9 detected by ELISA in hospitals were 87.5 and 81.3%, respectively. MSLN is helpful for the early diagnosis of pancreatic cancer and other diseases, and it had a significant ability to discriminate between pancreatic and nonpancreatic cancers (P value = 0.0159). The results from this study show that MIAS has the potential to become a new serological tumor marker detection platform for rapid detection and semi-quantitative determination of MSLN and would have broad applications in early clinical diagnosis.
引用
收藏
页码:2952 / 2957
页数:11
相关论文
共 50 条
  • [41] NEW METHOD FOR IMMUNOLOGICAL DIAGNOSIS OF PREGNANCY, ALLOWING A RAPID SEMI-QUANTITATIVE DETERMINATION OF HCG IN URINE
    不详
    PATHOLOGIE BIOLOGIE, 1972, 20 (7-8): : 430 - &
  • [42] Rapid, repeatable, highly sensitive and semi-quantitative colorimetric detection of elemental sulfur with a colored clathrate
    Yu, Peng
    Zhai, Yuanming
    Bao, Rui-Ying
    Liu, Zheng-Ying
    Yang, Ming-Bo
    Yang, Wei
    SENSORS AND ACTUATORS B-CHEMICAL, 2019, 299
  • [43] Rapid detection of dibutyl phthalate in liquor by a semi-quantitative multicolor immunosensor with naked eyes as readout
    Huang, Yanling
    Lin, Yue
    Luo, Fang
    Wang, Peilong
    Wang, Jian
    Qiu, Bin
    Guo, Longhua
    Lin, Zhenyu
    ANALYTICAL METHODS, 2019, 11 (04) : 524 - 529
  • [44] A RAPID METHOD FOR THE DETECTION AND SEMI-QUANTITATIVE ASSESSMENT OF MILLED GRANULATED BLASTFURNACE SLAG IN HARDENED CONCRETE
    MCIVER, JR
    DAVIS, DE
    CEMENT AND CONCRETE RESEARCH, 1985, 15 (03) : 545 - 548
  • [45] Potential application of a semi-quantitative method for mercury determination in soils, sediments and gold mining residues
    Yallouz, A. V.
    Cesar, R. G.
    Egler, S. G.
    ENVIRONMENTAL POLLUTION, 2008, 151 (03) : 429 - 433
  • [46] On-site microfluidic paper- based titration device for rapid semi-quantitative vitamin C content in beverages
    Kaewchuay, Netnapit
    Jantra, Jongjit
    Khettalat, Chanyaphon
    Ketnok, Supansa
    Peungpra, Naruemon
    Teepoo, Siriwan
    MICROCHEMICAL JOURNAL, 2021, 164
  • [47] A MINIATURIZED SYSTEM FOR RAPID AND QUANTITATIVE DETERMINATION OF A COCAINE METABOLITE BY A HOMOGENEOUS ENZYME IMMUNOASSAY
    Weng, Xuan
    Jiang, Hai
    Li, Dongqing
    INSTRUMENTATION SCIENCE & TECHNOLOGY, 2013, 41 (05) : 512 - 523
  • [48] Rapid semi-quantitative fluorimetric determination of citrinin in fungal cultures isolated from cheese and cheese factories
    Vazquez, BI
    Fente, C
    Franco, CM
    Quinto, E
    Cepeda, A
    Prognon, P
    LETTERS IN APPLIED MICROBIOLOGY, 1997, 24 (05) : 397 - 400
  • [50] A FIELD-TEST FOR THE DETECTION AND SEMI-QUANTITATIVE DETERMINATION OF SULFUR-DIOXIDE IN AIR AND WATER
    CHAUBE, A
    GUPTA, VK
    JOURNAL OF THE INDIAN CHEMICAL SOCIETY, 1982, 59 (09) : 1106 - 1107