Electrical impedance imaging for tissue monitoring and assessment during thermal therapy

被引:0
|
作者
Paulsen, KD [1 ]
Hartov, A [1 ]
Osterman, KS [1 ]
Mazzarese, R [1 ]
Kerner, T [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
关键词
electrical impedance imaging; electrical impedance spectroscopy; thermal imaging; tissue damage assessment;
D O I
10.1117/12.304343
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Electrical properties of tissues in the 10KHz to 10MHz range are known to be temperature sensitive making the monitoring and assessment of thermal insult delivered for therapeutic purposes possible through imaging schemes which spatially resolve these changes. We have been developing electrical impedance imaging technology from both the hardware data acquisition and software image reconstruction perspectives in order to realize the capability of spectroscopically examining the electrical property response of tissues undergoing hyperthermia therapy. Results from simulations, in vitro phantom experiments and in vivo studies including in human patients are presented. Specifically, a new prototype multi-frequency data acquisition system which is functional to 1MHz in both voltage and current modes is described. In addition, recent advances in image reconstruction methods which include the enhancement techniques of total variation minimization, dual meshing and spatial filtering are discussed. It is also clear that the electrical impedance spectrum of tissue has the potential to monitor other types of treatment-induced injury. Preliminary in vivo electrical impedance measurements in a rat leg model suggest that the tissue damage from radiation therapy can be tracked with this technique. Both dose and time-dependent responses have been observed in the electrical impedance data when compared to measurements recorded in an untreated control. Correlations with histological examination have also been performed and indicate that electrical impedance spectroscopy may provide unique information regarding tissue functional status and cellular morphology. Representative results from these studies are reported.
引用
收藏
页码:193 / 204
页数:12
相关论文
共 50 条
  • [11] Methods of optimization of electrical impedance tomography for imaging tissue electroporation
    Granot, Yair
    Rubinsky, Boris
    PHYSIOLOGICAL MEASUREMENT, 2007, 28 (10) : 1135 - 1147
  • [12] Assessment of distraction callus in rabbits by monitoring of the electrical impedance of bone
    Yoshida, Takashi
    Kim, Wook-Cheol
    Oka, Yoshinobu
    Yamada, Naotake
    Kubo, Toshikazu
    ACTA ORTHOPAEDICA, 2010, 81 (05) : 628 - 633
  • [13] Noninvasive Cerebral Imaging and Monitoring Using Electrical Impedance Tomography During Total Aortic Arch Replacement
    Li, Yandong
    Zhang, Ding
    Liu, Benyuan
    Jin, Zhenxiao
    Duan, Weixun
    Dong, Xiuzhen
    Fu, Feng
    Yu, Shiqiang
    Shi, Xuetao
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2018, 32 (06) : 2469 - 2476
  • [14] Electrical impedance imaging
    Isaacson, D
    Newell, JC
    Saulnier, GJ
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 3573 - 3575
  • [15] Application of Electrical Impedance Tomography for Monitoring Tissue Water Content of the Thigh
    Chen, Shuzhe
    Zhang, Ke
    Li, Maokun
    Zhang, Yuqi
    Wang, Yibing
    Yang, Fan
    Xu, Shenheng
    Abubakar, Aria
    2022 IEEE MTT-S INTERNATIONAL MICROWAVE BIOMEDICAL CONFERENCE (IMBIOC), 2022, : 171 - 173
  • [16] Assessment of acute pulmonary edema in dogs by electrical impedance imaging
    Newell, JC
    Edic, PM
    Ren, XD
    LarsonWiseman, JL
    Danyleiko, MD
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (02) : 133 - 138
  • [17] CORRELATION OF ELECTRICAL-IMPEDANCE AND TEMPERATURE IN TISSUE DURING FREEZING
    GAGE, AA
    CRYOBIOLOGY, 1979, 16 (01) : 56 - 62
  • [18] Electrical and thermal monitoring during cardiosurgery interventions
    Nowakowski, A.
    Kaczmarek, M.
    Wtorek, J.
    Stojek, W.
    Rogowski, J.
    Siebert, J.
    13TH INTERNATIONAL CONFERENCE ON ELECTRICAL BIOIMPEDANCE AND THE 8TH CONFERENCE ON ELECTRICAL IMPEDANCE TOMOGRAPHY 2007, 2007, 17 : 150 - +
  • [19] Electrical impedance endo-tomography: Imaging tissue from inside
    Jossinet, J
    Marry, E
    Montalibet, A
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (06) : 560 - 565
  • [20] Magnetoacoustic tomography with magnetic induction for imaging electrical impedance of biological tissue
    Li, Xu
    Xu, Yuan
    He, Bin
    Journal of Applied Physics, 2006, 99 (06):