Relationship of blood pressure and pump flow in an implantable centrifugal blood pump during hypertension

被引:13
|
作者
Akimoto, T
Yamazaki, K
Litwak, P
Litwak, KN
Tagusari, O
Mori, T
Antaki, JF
Kameneva, MV
Watach, MJ
Umezu, M
Tomioka, J
Kormos, RL
Koyanagi, H
Griffith, BP
机构
[1] Univ Pittsburgh, Dept Surg, McGowan Ctr, Pittsburgh, PA 15219 USA
[2] Tokyo Womens Med Coll, Dept Cardiovasc Surg, Tokyo 162, Japan
[3] SunMed Technol Res Corp, Nagano, Japan
[4] Waseda Univ, Dept Mech Engn, Tokyo, Japan
关键词
D O I
10.1097/00002480-200009000-00018
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to evaluate the real time relationship between pump flow and pump differential pressure (D-P) during experimentally induced hypertension (HT). Two calves (80 and 68 kg) were implanted with the EVA-HEART centrifugal blood pump (SunMedical Technology Research Corp., Nagano, Japan) under general anesthesia. Blood pressure (BP) in diastole was increased to 100 mm Hg by norepinephrine to simulate HT. Pump flow, D-P, EGG, and BP were measured at pump speeds of 1,800, 2,100, and 2,300 rpm. All data were separated into systole and diastole, and pump flow during HT was compared with normotensive (NT) conditions at respective pump speeds. Diastolic BP was increased to 99.3 +/- 4.1 mm Hg from 66.5 +/- 4.4mm Hg (p < 0.01). D-P in systole was under 40 mm Hg (range of change was 10 to 40 mm Hg) even during HT. During NT, the average systolic pump flow volume was 60% of the total pump flow. However, during HT, the average systolic pump flow was 100% of total pump flow volume, although the pump flow volume in systole during HT decreased (33.1 +/- 5.7 vs 25.9 +/- 4.0 ml/systole, p < 0.01). In diastole, the average flow volume through the pump was 19.6 +/- 6.9 ml/diastole during NT and -2.2 +/- 11.1 ml/diastole during HT (p < 0.01). The change in pump flow volume due to HT, in diastole, was greater than the change in pump flow in systole at each pump speed (p < 0.001). This study suggests that the decrease of mean pump flow during HT is mainly due to the decrease of the diastolic pump flow and, to a much lesser degree, systolic pump flow.
引用
收藏
页码:596 / 599
页数:4
相关论文
共 50 条
  • [21] A FLOW VISUALIZATION STUDY OF THE NCVC CENTRIFUGAL BLOOD PUMP
    ARAKI, K
    TAENAKA, Y
    MASUZAWA, T
    TATSUMI, E
    WAKISAKA, Y
    WATARI, M
    NAKATANI, T
    AKAGI, H
    BABA, Y
    ANAI, H
    PARK, YH
    EYA, K
    TODA, K
    TAKANO, H
    ARTIFICIAL ORGANS, 1994, 18 (09) : 669 - 672
  • [22] PIV measurements of flow in a centrifugal blood pump: Steady flow
    Day, SW
    McDaniel, JC
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (02): : 244 - 253
  • [23] Flow Characteristics and Coherent Structures in a Centrifugal Blood Pump
    Niclas Berg
    Laszlo Fuchs
    Lisa Prahl Wittberg
    Flow, Turbulence and Combustion, 2019, 102 : 469 - 483
  • [24] Visualization of relative flow patterns in centrifugal blood pump
    Weng Kong Chan
    S. C. M Yu
    L. P. Chua
    Y. W. Wong
    KSME International Journal, 2001, 15 : 1869 - 1875
  • [25] Visualization of flow structure in a centrifugal blood pump impeller
    Tsukiya T.
    Taenaka Y.
    Tatsumi E.
    Takano H.
    Journal of Artificial Organs, 2002, 5 (1) : 13 - 17
  • [26] Centrifugal blood pump scaling
    Lorenz, M.
    Smith, W.
    ASAIO Journal, 2000, 46 (02)
  • [28] Development of a totally implantable biventricular bypass centrifugal blood pump system
    Nosé, Y
    Nakata, K
    Yoshikawa, M
    Letsou, GV
    Fujisawa, A
    Wolner, E
    Schima, H
    ANNALS OF THORACIC SURGERY, 1999, 68 (02): : 775 - 779
  • [29] Investigation of MEMS as accelerometer sensor in an Implantable Centrifugal Blood Pump prototype
    S. Silva Neto
    J. R. C. Sousa Sobrinho
    C. da Costa
    T. F. Leão
    S. A. M. M. Senra
    E. G. P. Bock
    G. A. Santos
    S. T. Souza
    D. M. Silva
    C. Frajuca
    M. A. Souza
    Journal of the Brazilian Society of Mechanical Sciences and Engineering, 2020, 42
  • [30] Japanese clinical trial results of an implantable centrifugal blood pump "EVAHEART"
    Yamazaki, K.
    Saito, S.
    Nishinaka, T.
    Kurosawa, H.
    Nakatani, T.
    Kobayashi, J.
    Kitamura, S.
    Matsumiya, G.
    Yoshiki, S.
    Nishimura, T.
    Niinami, H.
    Kyo, S.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2008, 27 (02): : S246 - S246