A high-order prior for overlapped projections in the real flat-panel x-ray source imaging system

被引:1
|
作者
Duan, Jiayu [1 ]
Wang, Tianquan [2 ]
Li, Yang [1 ]
Mou, Xuanqin [1 ]
机构
[1] Xi An Jiao Tong Univ, Xian 710049, Shaanxi, Peoples R China
[2] Sun Yat Sen Univ, Canc Ctr, Guangzhou 510000, Guangdong, Peoples R China
来源
关键词
filed-emission; cold cathode; flat-panel x-ray source; high-order prior; FABRICATION; EMITTER;
D O I
10.1117/12.2594139
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
With the appearance of the field-emission cold cathode x-ray source, the fast response and small in size innovate the traditional imaging system. Although the field-emission x-ray source array generates multiple novel imaging modalities, it still faces a long stand-off distance between source and object. To realize a portable, smart, extremely low dose imaging modality, researchers proposed a concept to design a two-dimensional array field-emission source, namely the flat-panel source. In this paper, a real imaging system based on the ZnO field-emission flat-panel source is designed. Currently, the real flat-panel source faces an extremely low dose and non-addressable situation. Hence, the measurement based on the flat-panel source is overlapped and without an application potential. We first try to realize the imaging ability of the flat-panel source by designing a rebinning algorithm. With the analysis on the overlapped measurement, a high-order prior is introduced into the rebinning algorithm to improve the performance. Simulation and real data experiments verified our proposed method. Compared to the no high-order prior, the proposed algorithm can recover a more distinct measurement.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Dynamic chest radiography: flat-panel detector (FPD) based functional X-ray imaging
    Tanaka R.
    Radiological Physics and Technology, 2016, 9 (2) : 139 - 153
  • [32] The use of a flat-panel detector for seven-beam x-ray coherent scatter imaging
    Foran, Kelly A.
    Johns, Paul C.
    MEDICAL PHYSICS, 2017, 44 (08) : 4377 - 4377
  • [33] Medical imaging with mercuric iodide direct digital radiography flat-panel X-ray detectors
    Schieber, M
    Hermon, H
    Zentai, G
    Partain, L
    Street, RA
    Gilboa, H
    RADIOLOGY, 2001, 221 : 464 - 464
  • [34] Medical imaging with mercuric iodide direct digital radiography flat-panel X-ray detectors
    Gilboa, H
    Zuck, A
    Dagan, O
    Vilensky, A
    Breen, BN
    Taieb, A
    Reisman, B
    Hermon, H
    Zentai, G
    Partain, L
    Street, R
    Ready, S
    X-RAY AND GAMMA-RAY DETECTORS AND APPLICATIONS IV, 2002, 4784 : 315 - 325
  • [35] Optimal x-ray imaging geometry for flat-panel cone-beam computed tomography
    Siewerdsen, JH
    Jaffray, DA
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 106 - 107
  • [36] X-ray quantum limited portal imaging using amorphous silicon flat-panel arrays
    Munro, P
    Bouius, DC
    MEDICAL PHYSICS, 1998, 25 (05) : 689 - 702
  • [37] Direct-conversion flat-panel X-ray image detectors
    Kasap, SO
    Rowlands, JA
    IEE PROCEEDINGS-CIRCUITS DEVICES AND SYSTEMS, 2002, 149 (02): : 85 - 96
  • [38] Physical Characteristics of a New Flat-Panel Digital X-Ray Detector
    Cho, H.
    Kim, H.
    Park, H.
    Lee, C.
    Yu, A.
    Kim, D.
    Lee, Y.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [39] Diagonal 4-in ZnO Nanowire Cold Cathode Flat-Panel X-Ray Source: Preparation and Projection Imaging Properties
    Wang, Libin
    Xu, Yuan
    Cao, Xiuqing
    Huang, Jia
    Deng, Shaozhi
    Xu, Ningsheng
    Chen, Jun
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2021, 68 (03) : 338 - 345
  • [40] Performance of a static-anode/flat-panel x-ray fluoroscopy system in a diagnostic strength magnetic field: A truly hybrid x-ray/MR imaging system
    Fahrig, R
    Wen, Z
    Ganguly, A
    DeCrescenzo, G
    Rowlands, JA
    Stevens, GM
    Saunders, RF
    Pelc, NJ
    MEDICAL PHYSICS, 2005, 32 (06) : 1775 - 1784