Examining Hemodialysis Reliable Outflow catheter performance and cost in hemodialysis access

被引:11
|
作者
Torrent, Daniel J. [1 ]
Maness, Mandy R. [1 ]
Kachare, Swapnil D. [1 ]
Zink, Jill N. [2 ]
Haisch, Carl E. [1 ]
Harland, Robert C. [1 ]
Morgan, Claire [1 ]
Guyton, Rodney L. [3 ]
Colomb, Avery G. [3 ]
Barham, David W. [3 ]
Katz, Eric C. [3 ]
Stoner, Michael C. [4 ]
机构
[1] E Carolina Univ, Dept Surg, Greenville, NC USA
[2] E Carolina Univ, Dept Cardiovasc Surg, Greenville, NC USA
[3] E Carolina Univ, Brody Sch Med, Greenville, NC USA
[4] Univ Rochester, Dept Surg, Rochester, NY USA
关键词
Vascular access; Hemodialysis access; Hemodialysis; Hemodialysis Realiable Outflow; HeRO; Arteriovenous graft; VASCULAR ACCESS; HERO DEVICE; EXPERIENCE; FISTULAS;
D O I
10.1016/j.jss.2014.03.016
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: The Hemodialysis Reliable Outflow (HeRO) vascular access device is a hybrid polytetrafluoroethylene graft-stent construct designed to address central venous occlusive disease. Although initial experience has demonstrated excellent mid-term patency rates, subsequent studies have led to external validity questions. The purpose of this study was to examine a single center experience with this vascular access device in challenging access cases with associated costs. Methods: A retrospective study representing the authors' cumulative HeRO vascular access device experience was undertaken. The primary endpoint was graft failure or death, with secondary endpoints including secondary intervention rates and cost. Results: Forty-one patients with 15,579 HeRO days and a mean of 12.7 +/- 1.5 mo with the vascular access device were available for analysis. Secondary patency was 81.6% at 6 mo and 53.7% at 12 mo. The reintervention rate was 2.84 procedures per HeRO vascular access device year. Associated HeRO costs related to subsequent procedures were estimated at $34,713.63 per patient/y. Conclusions: These data on the patency and primary outcome data diverge significantly from initial multicenter studies and represent a real-world application of this technology. It is costly to maintain patency. Use of HeRO vascular access devices should be judicious with outcome expectations reduced. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:1 / 5
页数:5
相关论文
共 50 条
  • [1] PROSPECTIVE ANALYSIS OF HEMODIALYSIS RELIABLE OUTFLOW (HeRO) VASCULAR ACCESS GRAFT VS. CUFFED CATHETER ACCESS IN HEMODIALYSIS (HD) PATIENTS
    Provenzano, Robert
    LaFleur, Matthew
    McFadden, Lori
    LaFleur, Paula
    AMERICAN JOURNAL OF KIDNEY DISEASES, 2013, 61 (04) : A78 - A78
  • [2] Cost analysis of the Hemodialysis Reliable Outflow (HeRO) Graft compared to the tunneled dialysis catheter
    Al Shakarchi, Julien
    Inston, Nicholas
    Jones, Robert G.
    Maclaine, Grant
    Hollinworth, David
    JOURNAL OF VASCULAR SURGERY, 2016, 63 (04) : 1026 - 1033
  • [3] Percutaneous Interventions on the Hemodialysis Reliable Outflow Vascular Access Device
    Gebhard, Thomas A.
    Bryant, Joshua A.
    Grezaffi, J. Adam
    Pabon-Ramos, Waleska M.
    Gage, Shawn M.
    Miller, Michael J.
    Husum, Kurt W.
    Suhocki, Paul V.
    Sopko, David R.
    Lawson, Jeffrey H.
    Smith, Tony P.
    Kim, Charles Y.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (04) : 543 - 549
  • [4] The use of HeRO (Hemodialysis Reliable Outflow) graft in hemodialysis
    Legeza Peter
    Garbaisz David
    Szeberin Zoltan
    Sotonyi Peter
    ORVOSI HETILAP, 2019, 160 (31) : 1231 - 1234
  • [5] Bilateral Pulmonary Emboli Secondary to Indwelling Hemodialysis Reliable Outflow Catheter
    Coan, Kathryn E.
    O'Donnell, Mark E.
    Fankhauser, Grant T.
    Bodnar, Zachary
    Chandrasekaran, Krishnaswamy
    Stone, William M.
    VASCULAR AND ENDOVASCULAR SURGERY, 2013, 47 (04) : 317 - 319
  • [6] Catheter access for hemodialysis
    Johnson, MS
    SEMINARS IN DIALYSIS, 1998, 11 (06) : 326 - 330
  • [7] Use of Hemodialysis Reliable Outflow (HeRO) With Immediate Access Arteriovenous Grafts
    Nag, Uttara P.
    Turner, Megan C.
    Fang, Charles
    Wagner, Jason K.
    Devulapalli, Kavi
    Cox, Mitchell
    Avgerinos, Efthymios
    Dillavou, Ellen
    JOURNAL OF VASCULAR SURGERY, 2017, 65 (06) : 97S - 98S
  • [8] A comparison of patency and interventions in thigh versus Hemodialysis Reliable Outflow grafts for chronic hemodialysis vascular access
    Huber, Thomas
    Brownie, Evan R.
    JOURNAL OF VASCULAR SURGERY, 2016, 64 (05) : 1398 - 1399
  • [9] A Review on the Hemodialysis Reliable Outflow (HeRO) Graft for Haemodialysis Vascular Access
    Al Shakarchi, J.
    Houston, J. G.
    Jones, R. G.
    Inston, N.
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2015, 50 (01) : 108 - 113
  • [10] Vascular access via translumbar Hemodialysis Reliable Outflow dialysis catheter in a case of severe central venous occlusion
    Loskutov, Anatoly
    Dave, Atman
    Gooden, Christie
    Saucier, Nathan A.
    Cho, Kenneth H.
    Wible, Brandt C.
    Borsa, John J.
    JOURNAL OF VASCULAR ACCESS, 2018, 19 (04): : 410 - 411