Comparison of ocular straylight after implantation of multifocal intraocular lenses

被引:25
|
作者
Labuz, Grzegorz [1 ,4 ]
Reus, Nicolaas J. [1 ,2 ]
van den Berg, Thomas J. T. P. [3 ]
机构
[1] Rotterdam Ophthalm Inst, Rotterdam, Netherlands
[2] Amphia Hosp, Dept Ophthalmol, Breda, Netherlands
[3] Royal Netherlands Acad Arts & Sci, Netherlands Inst Neurosci, Amsterdam, Netherlands
[4] Univ Murcia, Fac Opt & Optometria, Murcia, Spain
来源
关键词
SURFACE LIGHT-SCATTERING; CATARACT-SURGERY; RETINAL STRAYLIGHT; RANDOMIZED-TRIAL; VISUAL FUNCTION; DONOR LENSES; OPACIFICATION; HISTORY; QUALITY; VISION;
D O I
10.1016/j.jcrs.2016.02.022
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
A comprehensive review of the effect of multifocal intraocular lens (IOL) designs on postoperative ocular straylight was performed. Studies reporting straylight values obtained with the natural pupil using the C-Quant device after uneventful multifocal IOL implantation were included. The IOLs were categorized based on their material characteristics; that is, hydrophobicity and presence of colored chromophores. Age adjustment was achieved using the straylight age-dependency norm for pseudophakic eyes. This norm also served as a reference for comparing mean straylight levels of the various IOLs. The literature review identified 10 studies reporting 9 multifocal IOL designs. The hydrophilic IOLs showed less straylight than the hydrophobic IOLs by 0.08 log(s) (P = .001). Blue violet light filtering IOLs showed less straylight than standard IOLs by 0.04 log(s), which was not statistically significant (P = .32). Hydrophobicity was a factor that significantly affected straylight in multifocal IOLs. (C) 2016 ASCRS and ESCRS
引用
收藏
页码:618 / 625
页数:8
相关论文
共 50 条
  • [41] Combined penetrating keratoplasty and implantation of iris prosthesis intraocular lenses after ocular trauma
    Mashor, Raneen Shehadeh
    Bahar, Irit
    Kaiserman, Igor
    Berg, Amy Lauren
    Slomovic, Allan
    Rootman, David S.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (03): : 582 - 587
  • [42] Visual Performance After Contralateral Implantation of Multifocal Intraocular Lenses With +3.0 and +4.0 Diopter Additions
    Nakamura, Kunihiko
    Bissen-Miyajima, Hiroko
    Yoshino, Mami
    Oki, Shinichi
    ASIA-PACIFIC JOURNAL OF OPHTHALMOLOGY, 2015, 4 (06): : 329 - 333
  • [43] Vision-related Quality of Life after Bilateral Implantation of Monofocal and Multifocal Intraocular Lenses
    Bamdad, Shahram
    Razavizadegan, Seyyed Ahmad
    Farvardin, Mohsen
    Mohaghegh, Sahar
    JOURNAL OF OPHTHALMIC & VISION RESEARCH, 2022, 17 (01) : 19 - 26
  • [44] A comparison of the visual performance of refractive rotationally asymmetric multifocal intraocular lenses and trifocal intraocular lenses
    Qiu, Jing-Jing
    Zhang, Qian
    Fu, Shu-Hua
    Huang, Jun-Qi
    Wang, Zhi-Hong
    You, Zhi-Peng
    Liu, Fei
    Sun, Tao
    MATERIALS EXPRESS, 2023, 13 (06) : 1110 - 1115
  • [45] Dissatisfaction after multifocal intraocular lens implantation
    Woodward, Maria A.
    Randleman, J. Bradley
    Stulting, R. Doyle
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (06): : 992 - 997
  • [46] Bilateral ciliary sulcus implantation of secondary piggyback multifocal intraocular lenses
    Clare, Gerry
    Bloom, Philip
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (02): : 320 - 322
  • [47] Dissatisfaction after multifocal intraocular lens implantation
    Tarib, Imane
    Asroui, Lara
    Ferguson, Tanner
    Randleman, James
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [48] Bilateral implantation of toric multifocal additive intraocular lenses in pseudophakic eyes
    Rabsilber, Tanja M.
    Kretz, Florian T. A.
    Holzer, Mike P.
    Fitting, Anna
    Sanchez, Maria J.
    Auffarth, Gerd U.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (08): : 1495 - 1498
  • [49] Comparison of additional pseudophakic multifocal lenses and multifocal intraocular lens in the capsular bag
    Schrecker, J.
    Feith, A.
    Langenbucher, A.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2014, 98 (07) : 915 - 919
  • [50] Method for in vitro assessment of straylight from intraocular lenses
    Labuz, Grzegorz
    Vargas-Martin, Fernando
    van den Berg, Thomas J. T. P.
    Lopez-Gil, Norberto
    BIOMEDICAL OPTICS EXPRESS, 2015, 6 (11): : 4457 - 4464