Comparison of Polar® RS800CX heart rate monitor and electrocardiogram for measuring inter-beat intervals in healthy dogs

被引:26
|
作者
Essner, Ann [1 ,2 ]
Sjostrom, Rita [3 ,4 ]
Ahlgren, Erik [2 ]
Gustas, Pia [5 ]
Edge-Hughes, Laurie [6 ]
Zetterberg, Lena [1 ]
Hellstrom, Karin [1 ]
机构
[1] Uppsala Univ, Dept Neurosci, Uppsala, Sweden
[2] Evidensia Djurkliniken Gefle, SE-80321 Gavle, Sweden
[3] Jamtland Cty Council, Stromsund Hlth Ctr, Stromsund, Sweden
[4] Mid Sweden Univ, Dept Hlth Sci, Ostersund, Sweden
[5] Swedish Univ Agr Sci, Uppsala, Sweden
[6] Canine Fitness Ctr, Calgary, AB, Canada
关键词
Behavior; Dogs; Heart rate variability; Physiotherapy; Reliability; Validity; RATE-VARIABILITY; AUTONOMIC REGULATION; RESPONSES; VALIDITY; PAIN; RELIABILITY; VALIDATION; WELFARE; STRESS; HORSES;
D O I
10.1016/j.physbeh.2014.10.034
中图分类号
B84 [心理学];
学科分类号
04 ; 0402 ;
摘要
The aim of the present study was to assess the criterion validity, relative reliability and level of agreement of Polar (R) RS800CX heart rate monitor measuring inter-beat intervals (IBIs), compared to simultaneously recorded electrocardiogram (ECG) in dogs. Methods: Five continuous minutes of simultaneously recorded IBIs from Polar (R) RS800CX and Cardiostore ECG in 11 adult healthy dogs maintaining standing position were analyzed. Polar (R) data was statistically compared to ECG data to assess for systematic differences between the methods. Three different methods for handling missing IBI data were used. Criterion validities were calculated by intraclass correlation coefficients (ICCs) and corresponding 95% confidence intervals (CIs). Relative reliabilities and levels of agreement were calculated by ICCs and the Bland and Altman analysis for repeated measurements per subject. Results: Correlation coefficients between IBI data from ECG and Polar (R) RS800CX varied between 0.73 and 0.84 depending on how missing values were handled. Polar (R) was over- and underestimating IBI data compared to ECG. The mean difference in log transformed (base10) IBI data was 0.8%, and 93.2% of the values were within the limits of agreement. Internally excluding three subjects presenting IBI series containing more than 5% erroneous IBIs resulted in ICCs between 0.97 and 0.99. Bland and Altman analysis (n = 8) showed mean difference was 1.8 ms, and 985% of the IBI values were plotted inside limits of agreement. Conclusion: This study showed that Polar (R) systematically biased recorded IBI series and that it was fundamental to detect measurement errors. For Polar (R) RS800CX heart rate monitor to be used interchangeably to ECG, by showing excellent criterion validity and reliable IBI measures in group and individual samples, only less than 5% of artifacts could be accepted. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:247 / 253
页数:7
相关论文
共 24 条
  • [1] Validity and reliability of Polar (R) RS800CX heart rate monitor, measuring heart rate in dogs during standing position and at trot on a treadmill
    Essner, Ann
    Sjostrom, Rita
    Ahlgren, Erik
    Lindmark, Birgitta
    PHYSIOLOGY & BEHAVIOR, 2013, 114 : 1 - 5
  • [2] Heart rate variability assessment with fingertip photoplethysmography and polar RS800cx as compared with electrocardiography in obese adolescents
    Vasconcellos, Fabrcio V. A.
    Seabra, Andre
    Cunha, Felipe A.
    Montenegro, Rafael A.
    Bouskela, Eliete
    Farinatti, Paulo
    BLOOD PRESSURE MONITORING, 2015, 20 (06) : 351 - 360
  • [3] Two-week test-retest reliability of the Polar® RS800CX torecord heart rate variability
    Williams, DeWayne P.
    Jarczok, Marc N.
    Ellis, Robert J.
    Hillecke, Thomas K.
    Thayer, Julian F.
    Koenig, Julian
    CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2017, 37 (06) : 776 - 781
  • [4] Comparison of Polar® RS800G3 heart rate monitor with Polar® S810i and electrocardiogram to obtain the series of RR intervals and analysis of heart rate variability at rest
    da Costa de Rezende Barbosa, Marianne Penachini
    da Silva, Natalia Turri
    de Azevedo, Fabio Micolis
    Pastre, Carlos Marcelo
    Marques Vanderlei, Luiz Carlos
    CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2016, 36 (02) : 112 - 117
  • [5] Monitoring of heart rate and inter-beat intervals with wrist plethysmography in patients with atrial fibrillation
    Harju, Jarkko
    Tarniceriu, Adrian
    Parak, Jakub
    Vehkaoja, Antti
    Yli-Hankala, Arvi
    Korhonen, Ilkka
    PHYSIOLOGICAL MEASUREMENT, 2018, 39 (06)
  • [6] Multiscale Distribution Entropy Analysis of Heart Rate Variability Using Differential Inter-Beat Intervals
    Lee, Dae-Young
    Choi, Young-Seok
    IEEE ACCESS, 2020, 8 : 48761 - 48773
  • [7] Accuracy of a low-cost heart rate monitor for research involving inter-beat interval
    Fishel, SR
    Muth, ER
    Herron, R
    Hoover, AW
    Rand, JR
    PSYCHOPHYSIOLOGY, 2004, 41 : S72 - S72
  • [8] On the validity of using the Polar RS800 heart rate monitor for heart rate variability research
    Daniel S. Quintana
    James A. J. Heathers
    Andrew H. Kemp
    European Journal of Applied Physiology, 2012, 112 : 4179 - 4180
  • [9] On the validity of using the Polar RS800 heart rate monitor for heart rate variability research
    Quintana, Daniel S.
    Heathers, James A. J.
    Kemp, Andrew H.
    EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2012, 112 (12) : 4179 - 4180
  • [10] VALIDATION OF HEART RATE MONITOR POLAR RS800 FOR HEART RATE VARIABILITY ANALYSIS DURING EXERCISE
    Hernando, David
    Garatachea, Nuria
    Almeida, Rute
    Casajus, Jose A.
    Bailon, Raquel
    JOURNAL OF STRENGTH AND CONDITIONING RESEARCH, 2018, 32 (03) : 716 - 725