Background and radiation resistance tests of neutral particle analyzer detectors for ITER by using a fast neutron beam

被引:5
|
作者
Afanasyev, V. I. [1 ]
Kozlovskii, S. S. [2 ]
Makar'in, D. V. [2 ]
Mel'nik, A. D. [1 ]
Mironov, M. I. [1 ]
Nesenevich, V. G. [1 ]
Petrov, M. P. [1 ]
Petrov, S. Ya. [1 ]
Chernyshev, F. V. [1 ]
机构
[1] Russian Acad Sci, AF Ioffe Phys Tech Inst, St Petersburg 194021, Russia
[2] St Petersburg State Polytech Univ, St Petersburg 195251, Russia
关键词
Plasma Physic Report; Neutron Fluence; Deuteron Beam; Sensitive Layer; Background Counting Rate;
D O I
10.1134/S1063780X10050053
中图分类号
O35 [流体力学]; O53 [等离子体物理学];
学科分类号
070204 ; 080103 ; 080704 ;
摘要
The radiation resistance and background sensitivity of scintillation (Hamamatsu H8500D photo-multiplier) and semiconductor (ORTEC BF-018-100-60 and BU-012-050-100) detectors to neutron and gamma radiation were investigated. Conclusions are drawn concerning the possibility of using such detectors in neutral particle analyzers that are being developed for ITER at the Ioffe Institute.
引用
收藏
页码:407 / 412
页数:6
相关论文
共 22 条
  • [21] Error estimation and parameter dependence of the calculation of the fast ion distribution function, temperature, and density using data from the KF1 high energy neutral particle analyzer on Joint European Torus
    Schlatter, C
    Testa, D
    Cecconello, M
    Murari, A
    Santala, M
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10): : 3547 - 3549
  • [22] Determination of the neutron and gamma ray dose rates of the irradiating beam tubes of a pool-type research reactor by using Monte Carlo simulation and experimental detectors regarding radiation protection issues
    Hedayat A.
    Emami J.
    Salartash R.
    International Journal of Advanced Nuclear Reactor Design and Technology, 2021, 3 : 27 - 43