Environmental Impact Analysis and Error Correction Method of Smart Meter On-site Verification

被引:0
|
作者
Wang, Huanning [1 ]
Zhao, Zhihua [1 ]
Huang, Yan [1 ]
Yao, Hejun [1 ]
Liu, Yuan [1 ]
Gai, Xugang [1 ]
Tan, Hengyu [1 ]
机构
[1] Beijing Inst Metrol, Beijing, Peoples R China
关键词
smart meter; environmental impact; uncertainty analysis; on-site verification;
D O I
10.1109/icsgsc50906.2020.9248560
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The verification of smart meters is generally performed in a laboratory environment, which has strict requirements on environmental conditions and power quality of the power source. However, in the case of resolving power tariff disputes between power operators and consumers, or performing online metering performance evaluation of smart meters, it is necessary to verify smart meters on site. During on-site verification, environmental factors such as temperature and humidity, electromagnetic interference, and power quality factors such as frequency, harmonics, and power factor will affect the energy error of smart meters. This paper analyzes and determines the main influencing factors of on-site verification, and the proposed on-site verification plan is given.
引用
收藏
页码:124 / 128
页数:5
相关论文
共 50 条
  • [1] A Smart Meter Measurement Time Error Identification and Correction Method based on TESM and Correlation Analysis
    Zhang, Xiaopeng
    Kong, Xiangyu
    Wang, Ning
    2021 IEEE POWER & ENERGY SOCIETY GENERAL MEETING (PESGM), 2021,
  • [2] Research on error verification of harmonic smart electricity meter
    Dong, Xianguang
    Li, Zhe
    Dai, Yanjie
    Wang, Qing
    Zhang, Zhi
    Wang, Pingxin
    Chen, Zhiru
    Wang, Tingting
    Xu, Ziqian
    2018 3RD INTERNATIONAL CONFERENCE ON MECHANICAL, CONTROL AND COMPUTER ENGINEERING (ICMCCE), 2018, : 286 - 290
  • [3] Error Analysis of the Illuminance Meter Verification System
    Zou Feng
    Wang Jian-wu
    ADVANCES IN PRECISION INSTRUMENTATION AND MEASUREMENT, 2012, 103 : 378 - 382
  • [4] Mobile on-site analysis in the environmental research
    Flachowsky, J
    FIELD SCREENING EUROPE, 1997, : 259 - 262
  • [5] Error Correction Method for Smart Energy Meter Field Calibration System under Non-standard Conditions
    Tan Hengyu
    Yao Hejun
    Huang Yan
    Wang Huanning
    Zhao Zhihua
    Liu Yuan
    2020 THE 4TH INTERNATIONAL CONFERENCE ON SMART GRID AND SMART CITIES (ICSGSC 2020), 2020, : 149 - 153
  • [6] Distribution Network Topology Error Correction Using Smart Meter Data Analytics
    Luan, Wenpeng
    Peng, Joshua
    Maras, Mirjana
    Lo, Joyce
    2013 IEEE POWER AND ENERGY SOCIETY GENERAL MEETING (PES), 2013,
  • [7] A strategy for evaluating the environmental impact of on-site sanitation systems
    vanRyneveld, MB
    Fourie, AB
    WATER SA, 1997, 23 (04) : 279 - 291
  • [8] Portable instruments for on-site analysis of environmental samples
    Duan, Chunfeng
    Li, Jiamin
    Zhang, Yuanhao
    Ding, Kun
    Geng, Xuhui
    Guan, Yafeng
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2022, 154
  • [9] Development of Automated On-site Analysis Method Using a Smart Device as Detection System for Autocatalytic Reactions
    Manaka, Atsushi
    Kaburagi, Kaori
    Furuyama, Shoichi
    Tafu, Masamoto
    Irie, Mituteru
    BUNSEKI KAGAKU, 2022, 71 (4-5) : 217 - 220
  • [10] ON-SITE ENVIRONMENTAL ANALYSIS REDUCING THE LAB TESTING BURDEN
    Depalma, Angelo
    Lab Manager, 2016, 11 (09):