The John Charnley Award Metal-on-Metal Hip Resurfacing versus Large-diameter Head Metal-on-Metal Total Hip Arthroplasty A Randomized Clinical Trial

被引:236
|
作者
Garbuz, Donald S. [1 ]
Tanzer, Michael [2 ]
Greidanus, Nelson V. [1 ]
Masri, Bassam A. [1 ]
Duncan, Clive P. [1 ]
机构
[1] Univ British Columbia, Div Lower Limb Reconstruct & Oncol, Dept Orthopaed, Vancouver, BC V5Z 4E3, Canada
[2] McGill Univ, Dept Surg, Div Orthopaed Arthroplasty, Montreal, PQ H3A 2T5, Canada
关键词
FOLLOW-UP; FEMORAL-NECK; REPLACEMENT; COBALT; OSTEOARTHRITIS; VALIDATION; PROSTHESES; CHROMIUM; RELEASE; BLOOD;
D O I
10.1007/s11999-009-1029-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Resurfacing arthroplasty has become an attractive option for young patients who want to maintain a high activity level. One recent study reported modestly increased activity levels for patients with resurfacing compared to standard total hip arthroplasty (THA). We conducted a prospective randomized clinical trial to compare clinical outcomes of resurfacing versus large-head metal-on-metal total hip arthroplasty. We randomized 107 patients deemed eligible for resurfacing arthroplasty to have either resurfacing or standard THA. Patients were assessed for quality-of-life outcomes using the PAT-5D index, WOMAC, SF-36, and UCLA activity score. The minimum followup was 0.8 years (mean, 1.1 years; range, 0.8-2.2 years). Of the 73 patients followed at least one year, both groups reported improvement in quality of life on all outcome measures. There was no difference in quality of life between the two arms in the study. Serum levels of cobalt and chromium were measured in a subset of 30 patients. In both groups cobalt and chromium was elevated compared to baseline. Patients receiving a large-head metal-on-metal total hip had elevated ion levels compared to the resurfacing arm of the study. At 1 year, the median serum cobalt increased 46-fold from baseline in patients in the large-head total hip group, while the median serum chromium increased 10-fold. At 1 year, serum cobalt was 10-fold higher and serum chromium 2.6-fold higher than in the resurfacing arm. Due to these excessively high metal ion levels, the authors recommend against further use of this particular large-head total hip arthroplasty.
引用
收藏
页码:318 / 325
页数:8
相关论文
共 50 条
  • [1] Dislocation of large-diameter head metal-on-metal total hip arthroplasty and hip resurfacing arthroplasty
    Miettinen, Simo S. A.
    Makinen, Tatu J.
    Laaksonen, Inari
    Makela, Keijo
    Huhtala, Heini
    Kettunen, Jukka S.
    Remes, Ville
    HIP INTERNATIONAL, 2019, 29 (03) : 253 - 261
  • [2] 2008 John Charnley Award: Metal Ion Levels After Metal-on-Metal Total Hip Arthroplasty: A Randomized Trial
    Engh, C. Anderson, Jr.
    MacDonald, Steven J.
    Sritulanondha, Supatra
    Thompson, Abigail
    Naudie, Douglas
    Engh, Charles A.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (01) : 101 - 111
  • [3] Hip resurfacing and metal-on-metal total hip arthroplasty
    Beverland, David E.
    Spencer, Robert F.
    HIP INTERNATIONAL, 2011, 21 (03) : 277 - 278
  • [4] Large-Diameter Metal-on-Metal Cementless Total Hip Arthroplasty in the Elderly
    Zhang, Xin
    Xu, Weidong
    Li, Jia
    Fang, Zhou
    Chen, Kangyao
    ORTHOPEDICS, 2010, 33 (12)
  • [5] HIP RESURFACING ARTHROPLASTY VERSUS LARGE-DIAMETER HEAD METAL-ON-METAL TOTAL HIP ARTHROPLASTY: COMPARISON OF THREE DESIGNS FROM THE FINNISH ARTHROPLASTY REGISTER
    Junnila, M.
    Kostensalo, I.
    Virolainen, P.
    Remes, V.
    Matilainen, M.
    Vahlberg, T.
    Pulkkinen, P.
    Eskelinen, A.
    Itala, A.
    Makela, K.
    SCANDINAVIAN JOURNAL OF SURGERY, 2014, 103 (01) : 54 - 59
  • [6] Metal Ion Release With Large-Diameter Metal-on-Metal Hip Arthroplasty
    Vendittoli, Pascal-Andre
    Amzica, Traian
    Roy, Alain G.
    Lusignan, Daniel
    Girard, Julien
    Lavigne, Martin
    JOURNAL OF ARTHROPLASTY, 2011, 26 (02): : 282 - 288
  • [7] Metal-on-metal resurfacing versus total hip replacement - the value of a randomized clinical trial
    Howie, DW
    McGee, MA
    Costi, K
    Graves, SE
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 2005, 36 (02) : 195 - +
  • [8] Revision rates for metal-on-metal hip resurfacing and metal-on-metal total hip arthroplasty - a systematic review
    Sorensen, Sofie-Amalie L. Ras
    Jorgensen, Henrik L.
    Sporing, Sune L.
    Lauritzen, Jes B.
    HIP INTERNATIONAL, 2016, 26 (06) : 515 - 521
  • [9] Taper Failure After Large-Diameter Metal-on-Metal Total Hip Arthroplasty
    Goldstein, Zachary H.
    Estrera, Kenneth
    Levine, Brett R.
    ORTHOPEDICS, 2016, 39 (05) : E984 - E987
  • [10] Metal-on-metal hip resurfacing arthroplasty
    Shimmin, Andrew
    Beaule, Paul E.
    Campbell, Pat
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (03): : 637 - 654