Analysis of current density and specific absorption rate in biological tissue surrounding an air-core type of transcutaneous transformer for an artificial heart

被引:0
|
作者
Shiba, Kenji [1 ]
Nukaya, Masayuki [1 ]
Tsuji, Toshio [1 ]
Koshiji, Kohji [2 ]
机构
[1] Hiroshima Univ, Grad Sch Engn, Hiroshima, Japan
[2] Tokyo Univ Sci, Chiba, Japan
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper reports on the specific absorption rate (SAR) and the current density analysis of biological tissue surrounding an air-core type of transcutaneous transformer for an artificial heart. The electromagnetic field in the biological tissue surrounding the transformer was analyzed by the transmission-line modeling method, and the SAR and current density as a function of frequency (200k-1MHz) for a transcutaneous transmission of 20W were calculated. The model's biological tissue has three layers including the skin, fat and muscle. As a result, the SAR in the vicinity of the transformer is sufficiently small and the normalized SAR value, which is divided by the ICNIRP's basic restriction, is 7 x 10(-3) or less. On the contrary, the current density is slightly in excess of the ICNIRP's basic restrictions as the frequency falls and the output voltage rises. Normalized current density is from 0.2 to 1.2. In addition, the layer in which the current's density is maximized depends on the frequency, the muscle in the low frequency (< 700kHz) and the skin in the high frequency (> 7001kHz). The result shows that precision analysis taking into account the biological properties is very important for developing the transcutaneous transformer for TAH.
引用
收藏
页码:1278 / +
页数:2
相关论文
共 17 条
  • [11] Analysis on transfer characteristics of air-core current transformer and its adaptability to relay protection
    He, Ruiwen
    Cai, Zexiang
    Wang, Yi
    Li, Jiaman
    Xie, Qiongxiang
    Dianwang Jishu/Power System Technology, 2013, 37 (05): : 1471 - 1476
  • [12] Error Analysis of Air-Core Coil Current Transformer Based on Stacking Model Fusion
    Li, Zhenhua
    Chen, Xingxin
    Wu, Lin
    Ahmed, Abu-Siada
    Wang, Tao
    Zhang, Yujie
    Li, Hongbin
    Li, Zhenxing
    Xu, Yanchun
    Tong, Yue
    ENERGIES, 2021, 14 (07)
  • [13] Thermal analysis of HTS air-core transformer used in voltage compensation type active SFCL
    Song, M.
    Tang, Y.
    Li, J.
    Zhou, Y.
    Chen, L.
    Ren, L.
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2010, 470 (20): : 1657 - 1661
  • [14] Electromagnetic Characteristics Analysis of Air-Core Transformer Used in Voltage Compensation Type Active SFCL
    Song, Meng
    Tang, Yuejin
    Zhou, Yusheng
    Ren, Li
    Chen, Lei
    Cheng, Shijie
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2010, 20 (03) : 1194 - 1198
  • [15] Analysis of Static Magnetic Flux Density and Electromagnetic Force Distribution of a Dry-type Air-core Reactor Under Different Operating Current
    Chu, Fanwu
    Fu, Ying
    Wang, Qi
    Wu, Xiong
    Liu, Zongxi
    Lei, Xiaoyan
    2019 22ND INTERNATIONAL CONFERENCE ON ELECTRICAL MACHINES AND SYSTEMS (ICEMS 2019), 2019, : 1190 - 1193
  • [16] Performance analysis on electronic current transformer based on PCB planar-type of air core coil
    Wu, Tao
    Zhou, You-Qing
    Gong, Wei
    Dianwang Jishu/Power System Technology, 2010, 34 (06): : 210 - 214
  • [17] Analysis of specific absorption rate and current density in an energy transmission system for a wireless capsule endoscope
    Shiba, Kenji
    Nagato, Tomohiro
    Tsuji, Toshio
    Koshiji, Kohji
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 6052 - +