Analysis of current density and specific absorption rate in biological tissue surrounding an air-core type of transcutaneous transformer for an artificial heart

被引:0
|
作者
Shiba, Kenji [1 ]
Nukaya, Masayuki [1 ]
Tsuji, Toshio [1 ]
Koshiji, Kohji [2 ]
机构
[1] Hiroshima Univ, Grad Sch Engn, Hiroshima, Japan
[2] Tokyo Univ Sci, Chiba, Japan
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper reports on the specific absorption rate (SAR) and the current density analysis of biological tissue surrounding an air-core type of transcutaneous transformer for an artificial heart. The electromagnetic field in the biological tissue surrounding the transformer was analyzed by the transmission-line modeling method, and the SAR and current density as a function of frequency (200k-1MHz) for a transcutaneous transmission of 20W were calculated. The model's biological tissue has three layers including the skin, fat and muscle. As a result, the SAR in the vicinity of the transformer is sufficiently small and the normalized SAR value, which is divided by the ICNIRP's basic restriction, is 7 x 10(-3) or less. On the contrary, the current density is slightly in excess of the ICNIRP's basic restrictions as the frequency falls and the output voltage rises. Normalized current density is from 0.2 to 1.2. In addition, the layer in which the current's density is maximized depends on the frequency, the muscle in the low frequency (< 700kHz) and the skin in the high frequency (> 7001kHz). The result shows that precision analysis taking into account the biological properties is very important for developing the transcutaneous transformer for TAH.
引用
收藏
页码:1278 / +
页数:2
相关论文
共 17 条
  • [1] Analysis of current density and specific absorption rate in biological tissue surrounding transcutaneous transformer for an artificial heart
    Shiba, Kenji
    Nukaya, Masayuki
    Tsuji, Toshio
    Koshiji, Kohji
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (01) : 205 - 213
  • [2] Analysis of specific absorption rate in biological tissue surrounding transcutaneous transformer for an artificial heart
    Shiba K.
    Koshiji K.
    Tatsumi E.
    Taenaka Y.
    Takano H.
    Journal of Artificial Organs, 2002, 5 (2) : 91 - 96
  • [3] Analysis of specific absorption rate and internal electric field in human biological tissues surrounding an air-core coil-type transcutaneous energy transmission transformer
    Shiba, Kenji
    Zulkifli, Nur Elina Binti
    Ishioka, Yuji
    JOURNAL OF ARTIFICIAL ORGANS, 2017, 20 (02) : 103 - 109
  • [4] Analysis of specific absorption rate and internal electric field in human biological tissues surrounding an air-core coil-type transcutaneous energy transmission transformer
    Kenji Shiba
    Nur Elina Binti Zulkifli
    Yuji Ishioka
    Journal of Artificial Organs, 2017, 20 : 103 - 109
  • [5] Energy transmission transformer for a wireless capsule endoscope: Analysis of specific absorption rate and current density in biological tissue
    Shiba, Kenji
    Nagato, Tomohiro
    Tsuji, Toshio
    Koshiji, Kohji
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (07) : 1864 - 1871
  • [6] Analysis of Specific Absorption Rate and Current Density in Biological Tissues Surrounding Energy Transmission Transformer for an Artificial Heart: Using Magnetic Resonance Imaging-based Human Body Model
    Higaki, Naoya
    Shiba, Kenji
    ARTIFICIAL ORGANS, 2010, 34 (01) : E1 - E9
  • [7] Electromagnetic effects on the biological tissue surrounding a transcutaneous transformer for an artificial anal sphincter system
    Peng Zan
    Bang-hua Yang
    Yong Shao
    Guo-zheng Yan
    Hua Liu
    Journal of Zhejiang University SCIENCE B, 2010, 11 : 931 - 936
  • [8] Electromagnetic effects on the biological tissue surrounding a transcutaneous transformer for an artificial anal sphincter system
    Peng ZANBanghua YANGYong SHAOGuozheng YANHua LIU Shanghai Key Laboratory of Power Station Automation TechnologyCollege of Mechatronics Engineering and AutomationShanghai UniversityShanghai China School of ElectronicsInformation and Electrical EngineeringShanghai Jiao Tong UniversityShanghai China
    Journal of Zhejiang University-Science B(Biomedicine & Biotechnology), 2010, (12) : 931 - 936
  • [9] Electromagnetic effects on the biological tissue surrounding a transcutaneous transformer for an artificial anal sphincter system
    Zan, Peng
    Yang, Bang-hua
    Shao, Yong
    Yan, Guo-zheng
    Liu, Hua
    JOURNAL OF ZHEJIANG UNIVERSITY-SCIENCE B, 2010, 11 (12): : 931 - 936