Novel methods for the quantification of toxic, residual phase transfer catalyst in fluorine-18 labeled radiotracers

被引:4
|
作者
Blevins, David W. [1 ]
Rigney, Grant H. [1 ,2 ]
Fang, Michael Y. [1 ,2 ]
Akula, Murthy R. [1 ]
Osborne, Dustin R. [1 ]
机构
[1] Univ Tennessee, Grad Sch Med, 1929 Alcoa Hwy, Knoxville, TN 37920 USA
[2] Univ Tennessee, Dept Mech Aerosp & Biomed Engn, Knoxville, TN USA
关键词
Absorbance; Conductance; Phase-transfer-catalyst; Radiopharmaceutical; Transmittance; CROWN-ETHERS; COMPLEX-FORMATION; KRYPTOFIX; 2.2.2; ALKALI; TETRABUTYLAMMONIUM; CHROMATOGRAPHY; CRYPTANDS; METAL; IONS; FDG;
D O I
10.1016/j.nucmedbio.2019.07.008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Fluorine-18 labeled radiopharmaceuticals undergo quality control testing for residual phase-transfer-catalyst content. The almost universally used quality-control test is a silica plate spot-test comparison of the radiopharmaceutical beside a 50-ppm standard. Once developed by staining, the radiopharmaceutical spot must be of equal or less intensity to pass the test. There is currently a need for a quantitative, inexpensive, and less subjective quality control method that allows the automatic incorporation of the acquired measurement directly into electronic batch reports. Results: In the developed method, a resazurin test solution is mixed with an aliquot of the radiopharmaceutical analyte along with dichloromethane (DCM). The mixture is vortexed. The potassium resazurin-phase transfer catalyst complex solubilizes into the DCM imparting a blue color. The organic layer is then removed for analysis. Three measurement methods were utilized: visual colorimetry against pre-prepared standards, spectrophotometric measurement of transmittance, and electrical conductance. A simple prototype spectrophotometer and an electrical test cell were constructed to acquire data. Sodium Resazurin dye was found to be a suitable test chromophore for residual phase transfer catalyst analysis of aqueous solutions. Quantitative spectrophotometric measurements are possible in the 0-100-ppm range (18-crown-6) and 0-150-ppm range (Kryptofix (R) or tetrabutylammonium). Electrical resistance measurements of the phase transfer-catalyst resazurin complex in DCM are also a viable method, allowing quantitative phase transfer catalyst measurements in the 0-100-ppm range. Conclusion: The methodologies developed are more quantitative alternatives to the current spot-test method. The spectrophotometric method was determined to be the most accurate method. (C) 2019 The Authors. Published by Elsevier Inc.
引用
收藏
页码:41 / 48
页数:8
相关论文
共 43 条
  • [41] Novel fluorine-18 labeled 5-(1-pyrrolidinylsulfonyl)-7-azaisatin derivatives as potential PET tracers for in vivo imaging of activated caspases in apoptosis
    Waldmann, Christopher M.
    Hermann, Sven
    Faust, Andreas
    Riemann, Burkhard
    Schober, Otmar
    Schaefers, Michael
    Haufe, Guenter
    Kopka, Klaus
    BIOORGANIC & MEDICINAL CHEMISTRY, 2015, 23 (17) : 5734 - 5739
  • [42] Comparison of a novel fluorine-18 labeled PSMA-targeting agent, CTT1057, to gallium-68 PSMA-11 in patients with prostate cancer
    Behr, Spencer
    Aggarwal, Rahul
    Flavell, Robert
    Hope, Thomas
    Gao, Kenneth
    Ravanfar, Vahid
    Seo, Youngho
    Langton-Webster, Beatrice
    Berkman, Clifford
    VanBrocklin, Henry
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [43] Synthesis of Three Novel Fluorine-18 Labeled Analogues of L-Deprenyl for Positron Emission Tomography (PET) studies of Monoamine Oxidase B (MAO-B)
    Nag, Sangram
    Lehmann, Lutz
    Heinrich, Tobias
    Thiele, Andrea
    Kettschau, Georg
    Nakao, Ryuji
    Gulyas, Balazs
    Halldin, Christer
    JOURNAL OF MEDICINAL CHEMISTRY, 2011, 54 (20) : 7023 - 7029