TTC-Pluronic 3D radiochromic gel dosimetry of ionizing radiation

被引:39
|
作者
Kozicki, Marek [1 ,2 ]
Kwiatos, Klaudia [1 ]
Kadlubowski, Slawomir [3 ]
Dudek, Mariusz [4 ]
机构
[1] Lodz Univ Technol, Dept Man Made Fibres, Lodz, Poland
[2] GeVero Co, Lodz, Poland
[3] Lodz Univ Technol, Inst Appl Radiat Chem, Lodz, Poland
[4] Lodz Univ Technol, Inst Mat Sci & Engn, Lodz, Poland
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2017年 / 62卷 / 14期
关键词
2; 3; 5-triphenyltetrazolium chloride; Pluronic; 3D radiochromic gel; 3D dosimetry; radiotherapy dosimetry; ionising radiation; leucomalachite green; NITRO BLUE TETRAZOLIUM; HIGH-DOSE DOSIMETRY; POLYMER GELS; AQUEOUS-SOLUTIONS; COMPUTED-TOMOGRAPHY; HYDRATED ELECTRONS; HYDROXYL RADICALS; PULSE-RADIOLYSIS; ACTIVE COMPOUND; OPTICAL SCANNER;
D O I
10.1088/1361-6560/aa77eb
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work reports the first results obtained using a new 3D radiochromic gel dosimeter. The dosimeter is an aqueous physical gel matrix made of poly(ethylene oxide)-block-poly(propylene oxide)-block-poly(ethylene oxide) (Pluronic F-127, PEO-PPO-PEO) doped with a representative of tetrazolium salts, 2, 3, 5-triphenyltetrazolium chloride (TTC). There were several reasons for the choice of Pluronic as a gel forming substrate: (i) the high degree of transparency and colourlessness; (ii) the possibility of gel dosimeter preparation at both high and low temperatures due to the phase behaviour of Pluronic; (iii) the broad temperature range over which the TTC-Pluronic dosimeter is stable; and (iv) the non-toxicity of Pluronic. A reason for the choice of TTC was its ionising radiation-induced transformation to water-insoluble formazan, which was assumed to impact beneficially on the spatial stability of the dose distribution. If irradiated, the TTC-Pluronic gels become red but transparent in the irradiated part, while the non-irradiated part remains crystal clear. The best obtained composition is characterised by <4 Gy dose threshold, a dose sensitivity of 0.00231 (Gy x cm)(-1), a large linear dose range of >500 Gy and a dynamic dose response much greater than 500 Gy (7.5% TTC, 25% Pluronic F-127, 50 mmol dm(-3) tetrakis). Temporal and spatial stability studies revealed that the TTC-Pluronic gels (7.5% TTC, 25% Pluronic F-127) were stable for more than one week. The addition of compounds boosting the gels' dose performance caused deterioration of the gels' temporal stability but did not impact the stability of the 3D dose distribution. The proposed method of preparation allows for the repeatable manufacture of the gels. There were no differences observed between gels irradiated fractionally and non-fractionally. The TTC-Pluronic dose response might be affected by the radiation source dose rate-this, however, requires further examination.
引用
收藏
页码:5668 / 5690
页数:23
相关论文
共 50 条
  • [31] GLIOMA CELL MIGRATION IN 3D IS NOT IMPAIRED BY IONIZING RADIATION
    Storch, K.
    Eke, I.
    Kaestner, I.
    Mueller-Klieser, W.
    Faethe, C.
    Taucher-Scholz, G.
    Cordes, N.
    RADIOTHERAPY AND ONCOLOGY, 2012, 102 : S27 - S27
  • [32] Gd3+-doped sol-gel silica glass for remote ionizing radiation dosimetry
    El Hamzaoui, Hicham
    Bouwmans, Geraud
    Capoen, Bruno
    Cassez, Andy
    Haber, Remi
    Ouerdane, Youcef
    Girard, Sylvain
    Di Francesca, Diego
    Kerboub, Nourdine
    Morana, Adriana
    Soderstrom, Daniel
    Boukenter, Aziz
    Bouazaoui, Mohamed
    OSA CONTINUUM, 2019, 2 (03) : 715 - 721
  • [33] Tetrazolium salts-Pluronic F-127 gels for 3D radiotherapy dosimetry
    Kwiatos, Klaudia
    Maras, Piotr
    Kadlubowski, Slawomir
    Stempien, Zbigniew
    Dudek, Mariusz
    Kozicki, Marek
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (09):
  • [34] Preliminary Investigation of a Novel 3D Dosimetry System Consisting of ClearView 3D Radiochromic Dosimeters with Telecentric Optical Computed Tomography Scanning
    Hoopingarner, S.
    Penev, K.
    Oldham, M.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [35] A practical and accurate 3D dosimetry system for radiation therapy
    Guo, P.
    Adamovics, J.
    Oldham, M.
    MEDICAL PHYSICS, 2006, 33 (06) : 2193 - 2193
  • [36] Evaluation of a 3D diamond detector for medical radiation dosimetry
    Kanxheri, K.
    Servoli, L.
    Oh, A.
    Sanchez, F. Munoz
    Forcolin, G. T.
    Murphy, S. A.
    Aitkenhead, A.
    Moore, C. J.
    Morozzi, A.
    Passeri, D.
    Bellini, M.
    Corsi, C.
    Lagomarsino, S.
    Sciortino, S.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [37] 3D MAGIC gel dosimetry for verifying proton patch fields
    Shahnazi, K
    Cheng, I
    Moyers, M
    Luu, O
    Holshouser, B
    Loredo, L
    Slater, J
    Miller, D
    MEDICAL PHYSICS, 2003, 30 (06) : 1520 - 1520
  • [38] Low-density polymer gel dosimeters for 3D radiation dosimetry in the thoracic region: A preliminary study
    De Deene, Yves
    Vandecasteele, Jan
    Vercauteren, Tom
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [39] 3D Slicer Gel Dosimetry Analysis: Validation of the Calibration Process
    Alexander, K. M.
    Pinter, C.
    Andrea, J.
    Fichtinger, G.
    Schreiner, L. J.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 521 - 524
  • [40] New normoxic N-(Hydroxymethyl)acrylamide based polymer gel for 3D dosimetry in radiation therapy
    Rabaeh, Khalid A.
    Basfar, Ahmed A.
    Almousa, Akram A.
    Devic, Slobodan
    Moftah, Belal
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 33 : 121 - 126