Systolic acquisition of coronary dual-source computed tomography angiography: feasibility in an unselected patient population

被引:11
|
作者
Bamberg, Fabian [1 ]
Sommer, Wieland H. [1 ]
Schenzle, Jan C. [1 ]
Becker, Christoph R. [1 ]
Nikolaou, Konstantin [1 ]
Reiser, Maximilian F. [1 ]
Johnson, Thorsten R. C. [1 ]
机构
[1] Univ Munich, Klinikum Grosshadern, Dept Clin Radiol, D-81377 Munich, Germany
关键词
Cardiac CT; Dual-source CT; Image quality; Systolic reconstruction; Dose reduction; SOURCE CT; ARTERY-DISEASE; IMAGE QUALITY; DIAGNOSTIC-ACCURACY; ATRIAL-FIBRILLATION; INITIAL-EXPERIENCE; STATEMENT;
D O I
10.1007/s00330-009-1680-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To determine the practicability and potential dose saving of an imaging algorithm incorporating a pulsing scheme applying systolic data acquisition at heart rates > 75 beats per minute (bpm). Patients clinically referred for coronary computed tomography angiography (CTA) underwent cardiac CTA using either a diastolic pulsing window (30-70%) or a narrow systolic pulsing window (150 ms at 300 ms). Independent investigators retrospectively determined image quality (1, excellent, to 5, unreadable) and derived effective radiation exposure. Among all 101 subjects (62 +/- 2 years, 59% male) the predicted decrease in the best reconstruction interval for diastolic phases was 12 ms per 1 bpm [95% confidence interval (CI): -13.5 to -11.2] and -1.9 ms for systolic phases (95% CI: -3.2 to -0.62, p = 0.004), independent of age, gender and body mass index (BMI). The systolic pulsing strategy in 47 subjects (23 subjects > 75 bpm) resulted in significantly lower radiation exposure (4.97 +/- 2.3 vs 9.38 +/- 5.5 mSv, p < 0.001 for systolic versus diastolic, respectively), whereas there was no difference with respect to image quality or heart rate (p = 0.65 and p = 0.74, respectively). Our results suggest that a systolic window for tube current modulation in subjects with higher heart rates represents a reliable tool to ensure high image quality at significantly lower dose in patients undergoing routine cardiac CTA.
引用
收藏
页码:1331 / 1336
页数:6
相关论文
共 50 条
  • [41] Dual-source computed tomography coronary artery imaging in children
    Secinaro, Aurelio
    Curione, Davide
    Mortensen, Kristian Havmand
    Santangelo, Teresa Pia
    Ciancarella, Paolo
    Napolitano, Carmela
    Del Pasqua, Alessia
    Taylor, Andrew Mayall
    Ciliberti, Paolo
    PEDIATRIC RADIOLOGY, 2019, 49 (13) : 1823 - 1839
  • [42] Coronary revascularization treatment based on dual-source computed tomography
    R. Dikkers
    T. P. Willems
    L. H. Piers
    G. J. de Jonge
    R. A. Tio
    H. J. van der Zaag-Loonen
    P. M. A. van Ooijen
    F. Zijlstra
    M. Oudkerk
    European Radiology, 2008, 18 : 1800 - 1808
  • [43] High-pitch dual-source CT coronary angiography: systolic data acquisition at high heart rates
    Robert Goetti
    Gudrun Feuchtner
    Paul Stolzmann
    Lotus Desbiolles
    Michael Alexander Fischer
    Christoph Karlo
    Stephan Baumueller
    Hans Scheffel
    Hatem Alkadhi
    Sebastian Leschka
    European Radiology, 2010, 20 : 2565 - 2571
  • [44] Image quality on dual-source computed-tomographic coronary angiography
    Rixe, Johannes
    Rolf, Andreas
    Conradi, Guido
    Elsaesser, Albrecht
    Moellmann, Helge
    Nef, Holger M.
    Bachmann, Georg
    Hamm, Christian W.
    Dill, Thorsten
    EUROPEAN RADIOLOGY, 2008, 18 (09) : 1857 - 1862
  • [45] High-pitch dual-source CT coronary angiography: systolic data acquisition at high heart rates
    Goetti, Robert
    Feuchtner, Gudrun
    Stolzmann, Paul
    Desbiolles, Lotus
    Fischer, Michael Alexander
    Karlo, Christoph
    Baumueller, Stephan
    Scheffel, Hans
    Alkadhi, Hatem
    Leschka, Sebastian
    EUROPEAN RADIOLOGY, 2010, 20 (11) : 2565 - 2571
  • [46] Image quality on dual-source computed-tomographic coronary angiography
    Johannes Rixe
    Andreas Rolf
    Guido Conradi
    Albrecht Elsaesser
    Helge Moellmann
    Holger M. Nef
    Georg Bachmann
    Christian W. Hamm
    Thorsten Dill
    European Radiology, 2008, 18 : 1857 - 1862
  • [47] Feasibility study of the best monitoring time-interval to track contrast agent bolus in dual-source coronary computed tomography angiography
    Han, Y.
    Wang, T.
    Lin, L.
    Yu, C.
    Lv, R.
    Liu, Z.
    Zhang, T.
    Han, L.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2022, 20 (01): : 145 - 150
  • [48] Accuracy of dual-source computed tomography coronary angiography: evaluation with a standardised protocol for cardiac surgeons Reply
    Plass, Andre
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2011, 39 (02) : 283 - 283
  • [49] Evaluation of coronary artery variations using dual-source coronary computed tomography angiography in neonates with transposition of the great arteries
    Odawara, Yuko
    Kawamura, Nobuko
    Yamasaki, Yuzo
    Hashimoto, Joji
    Ishikawa, Shiro
    Honda, Hiroshi
    JAPANESE JOURNAL OF RADIOLOGY, 2019, 37 (04) : 308 - 314
  • [50] Evaluation of coronary artery variations using dual-source coronary computed tomography angiography in neonates with transposition of the great arteries
    Yuko Odawara
    Nobuko Kawamura
    Yuzo Yamasaki
    Joji Hashimoto
    Shiro Ishikawa
    Hiroshi Honda
    Japanese Journal of Radiology, 2019, 37 : 308 - 314