Developing a radon monitor with less influence of humidity based on electrostatic collection method and CR-39 detector

被引:6
|
作者
Fan, Zhongkai [1 ]
Hu, Tao [1 ]
Xie, Ruomei [1 ]
Lin, Fen [1 ]
Mo, Yixiang [1 ]
Yuan, Shuai [1 ]
Liu, Shuaibin [1 ]
Yuan, Hongzhi [1 ]
Tan, Yanliang [1 ]
机构
[1] Hengyang Normal Univ, Coll Phys & Elect Engn, Hengyang 421008, Hunan Province, Peoples R China
基金
中国国家自然科学基金;
关键词
Radon monitor; Electrostatic collection method; CR-39; Collection efficiency; EXHALATION RATE;
D O I
10.1016/j.radmeas.2022.106787
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A new radon monitor with a 50 mL hemispherical internal cell (Hynu EIRM) is designed to develop a radon monitor with less influence of humidity based on electrostatic collection and CR-39. The collection efficiency of Po-218 ions in the Hynu EIRM while the cell-wall voltage is 2000 V is estimated by a model established previously. The collection efficiency of Po-218 ions under the relative humidity of 1% is 4.81% higher than that under the relative humidity of more than 10%. Several verification experiments are performed. The Hynu EIRMs are used to measure the Rn-222 concentration under three relative humidity ranges: 5%-10%, 60%-70% and 98%-100%. Results show that the largest difference of the calibration factors of the three experiments is 9.24%, which is a little larger than the difference (4.81%) estimated by the model. The reasons for this outcome may be that the simplified model underestimates the differences of the collection efficiency under different relative humidities and the statistical uncertainty in the experiments. The difference of the calibration factors of the Hynu EIRM at any relative humidity is less than 10%. This study is the first time that the influence of humidity on the detection efficiency of radon monitors based on electrostatic collection and CR-39 is reduced by decreasing the volume of the internal cell and increasing the cell-wall voltage.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Intercomparison of radon CR-39 detector systems conducted in CLOR's calibration chamber
    Mamont-Ciesla, Kalina
    Stawarz, Olga
    Karpinska, Maria
    Kapala, Jacek
    Kozak, Krzysztof
    Grzadziel, Dominik
    Chalupnik, Stanislaw
    Chmielewska, Izabela
    Olszewski, Jerzy
    Przylibski, Tadeusz A.
    Zebrowski, Adam
    NUKLEONIKA, 2010, 55 (04) : 589 - 593
  • [32] Calibration technique for a CR-39 detector for soil and water radon exhalation rate measurements
    B. A. Almayahi
    A. A. Tajuddin
    M. S. Jaafar
    Journal of Radioanalytical and Nuclear Chemistry, 2014, 301 : 133 - 140
  • [33] Development of the ultra-high sensitivity radon monitor based on electrostatic collection method
    Xie, Ruomei
    Lin, Fen
    Yi, Haibo
    Hu, Tao
    Liu, Shuaibin
    Yuan, Shuai
    Fan, Zhongkai
    Mo, Yixiang
    Li, Huiying
    Sun, Jiale
    Liu, Zhipeng
    Wu, Jiulin
    Yuan, Hongzhi
    Xie, Dong
    Tan, Yanliang
    JOURNAL OF INSTRUMENTATION, 2023, 18 (11)
  • [34] DETERMINATION OF DETECTION EFFICIENCY FOR RADON AND RADON DAUGHTERS WITH CR-39 TRACK DETECTOR - A MONTE-CARLO STUDY
    NIKEZIC, D
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 344 (02): : 406 - 414
  • [35] CAN ELECTRIC-FIELD INFLUENCE THE SENSITIVITY OF CR-39 DETECTOR
    MATIULLAH
    TAKEDA, N
    KUDO, K
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1995, 34 (03): : 1709 - 1710
  • [36] A fast method for the determination of the efficiency coefficient of bare CR-39 detector
    Dwaikat, Nidal
    El-hasan, Mousa
    Sueyasu, Masto
    Kada, Wataru
    Sato, Fuminobu
    Kato, Yushi
    Saffarini, G.
    Iida, Toshiyuki
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2010, 268 (20): : 3351 - 3355
  • [37] Thermal effects on CR-39 detector response and permeability of some membranes used in radon measurements
    Sharaf, MA
    Abo-Elmagd, M
    RADIATION MEASUREMENTS, 2005, 39 (04) : 441 - 445
  • [38] SENSITIVITY OF RADON MEASUREMENTS WITH CR-39 TRACK ETCH DETECTOR - A MONTE-CARLO STUDY
    NIKEZIC, D
    KOSTIC, D
    KRSTIC, D
    SAVOVIC, S
    RADIATION MEASUREMENTS, 1995, 25 (1-4) : 647 - 648
  • [39] Radon Exhalation Rate from Building Materials Using CR-39 Nuclear Track Detector
    Topcu, N.
    Bicak, D.
    Cam, S.
    Erees, F. S.
    INDOOR AND BUILT ENVIRONMENT, 2013, 22 (02) : 384 - 387
  • [40] Factors affecting use of CR-39 surface monitor technology to estimate past exposure to indoor radon
    Mahaffey, JA
    Parkhurst, MA
    Hui, TE
    Brownson, RC
    Alavanja, MCR
    JOURNAL OF EXPOSURE ANALYSIS AND ENVIRONMENTAL EPIDEMIOLOGY, 1996, 6 (04): : 425 - 437