Toward Efficient Low-Temperature Ozone Gas Sterilization of Medical Devices

被引:15
|
作者
Thill, Sandy A. [1 ]
Spaltenstein, Marc [1 ]
机构
[1] SteriLux SA, Chemin Viaduc 12, CH-1008 Prilly, Switzerland
关键词
Ozone; sterilization; heat-sensitive medical devices; optical ozone measurement system; V-UV radiation to generate ozone; BACILLUS-SUBTILIS; SPORICIDAL ACTIVITY; INACTIVATION; HUMIDITY; SPORES;
D O I
10.1080/01919512.2019.1704217
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
An ozone generation system using V-UV excimer lamps (172nm) and an optical measurement system based on the Beer-Lambert law were developed for an efficient low-temperature (15 degrees C-30 degrees C) sterilization of medical devices. Ozone concentrations between 100 and 1MODIFIER LETTER PRIME000 ppm (0.2-2 g/Nm(3)) were generated and the impact of several variables on the sterilization outcome were studied. The survivor curves of the sterilization process were obtained for biological indicators (BIs) inoculated with Geobacillus stearothermophilus spores. A lag phase was observed before the start of the actual sterilization which followed first-order log-linear kinetics. The results demonstrated the feasibility to sterilize medical devices with ozone at ambient temperatures, reducing the consumption of resources and the cost of sterilization.
引用
收藏
页码:386 / 398
页数:13
相关论文
共 50 条
  • [1] Research on low-temperature plasma device for rapid sterilization of medical instruments
    Li B.
    Xin Z.
    Shang Z.
    Li A.
    Jiang T.
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2022, 43 (09): : 62 - 71
  • [2] Ethylene oxide gas sterilization of medical devices
    Shintani H.
    Shintani, Hideharu, 2017, Society for Antibacterial and Antifungal Agents Japan (22) : 1 - 16
  • [3] Ethylene Oxide Gas Sterilization of Medical Devices
    Shintani, Hideharu
    BIOCONTROL SCIENCE, 2017, 22 (01) : 1 - 16
  • [4] Low-temperature sterilization and new technologies
    Goveia, Vania Regina
    Cunha Pinheiro, Silma Maria
    Graziano, Kazuko Uchikawa
    REVISTA LATINO-AMERICANA DE ENFERMAGEM, 2007, 15 (03): : 373 - 376
  • [5] LOW-TEMPERATURE STEAM AND FORMALDEHYDE STERILIZATION
    ROBERTSHAW, RG
    JOURNAL OF HOSPITAL INFECTION, 1983, 4 (03) : 305 - 314
  • [6] LOW-TEMPERATURE PYROLYSIS OF OZONE
    POPOVICH, MP
    EGOROVA, GV
    FILIPPOV, YV
    ZHURNAL FIZICHESKOI KHIMII, 1985, 59 (02): : 273 - 283
  • [7] LOW-TEMPERATURE DEVICES
    MCWHORTER, AL
    IRE TRANSACTIONS ON EDUCATION, 1960, 3 (04): : 137 - 141
  • [8] STERILIZATION - LOW-TEMPERATURE STERILIZATION ALTERNATIVES IN THE 1990S
    SCHNEIDER, PM
    TAPPI JOURNAL, 1994, 77 (01): : 115 - 119
  • [9] Application of 100% ozone gas process to rapid low-temperature oxidation
    Nonaka, H
    Ichimura, S
    Nishiguchi, T
    Morikawa, Y
    Kekura, M
    Miyamoto, M
    10TH IEEE INTERNATIONAL CONFERENCE ON ADVANCED THERMAL PROCESSING OF SEMICONDUCTORS - RTP 2002, 2002, : 119 - 123
  • [10] Characteristics of throttle low-temperature devices on ozone-safe working media
    Khimicheskoe I Neftegazovoe Mashinostroenie, 1995, (05): : 22 - 24