Biometry with a new swept-source optical coherence tomography biometer: Repeatability and agreement with an optical low-coherence reflectometry device

被引:77
|
作者
Kurian, Mathew [1 ]
Negalur, Nikhil [1 ]
Das, Sudeep [1 ]
Puttaiah, Narendra K. [1 ]
Haria, Dhawal [1 ]
Tejal, S. J. [1 ]
Thakkar, Maithil M. [1 ]
机构
[1] Narayana Nethralaya, Bangalore, Karnataka, India
来源
关键词
INTRAOCULAR-LENS POWER; INTERFEROMETRY; ACCURACY;
D O I
10.1016/j.jcrs.2016.01.038
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess the repeatability of a new swept-source optical coherence tomography (SS-OCT) biometer (IOLMaster 700) and evaluate the agreement with an optical low-coherence reflectometry (OLCR) biometer (Lenstar 900) in cataractous eyes. SETTING: Tertiary eyecare facility. DESIGN: Prospective evaluation of a diagnostic test. METHODS: Eyes of patients had 3 consecutive scans acquired by the same operator. The repeatability of measurements was assessed using the within-subject standard deviation and coefficient of variation (CoV). Agreement between the mean measurements of each machine was evaluated. RESULTS: The study comprised 100 eyes of 100 patients. The median lens thickness and axial length (AL) measurements were significantly greater (P < .001) with SS-OCT (4.29 mm and 23.5 mm, respectively, versus 4.25 mm and 23.4 mm, respectively); anterior chamber depth (ACD) was greater with OLCR (2.72 mm versus 2.70 mm, respectively) (P < .001). The within-subject standard deviation and CoV of the AL (0.01 and 0.05, respectively) and the ACD (0.04 and 1.22, respectively) measurements showed lower variability with SS-OCT (0.05 and 0.21, respectively, for AL; 0.06 and 1.99, respectively, for ACD) than with OLCR. Good agreement was seen for anterior segment measurements between the 2 devices. The 95% limits of agreement of most parameters fell within the test retest variability of the parameters. The SS-OCT biometer also showed better penetration in dense posterior subcapsular cataracts, measuring AL successfully in 96% of cases. CONCLUSIONS: The new SS-OCT biometer showed good repeatability and agreement with the OLCR biometer. The SS-OCT biometer measured the AL with fewer dropouts compared with the OLCR even in dense cataracts. (C) 2016 ASCRS and ESCRS
引用
收藏
页码:577 / 581
页数:5
相关论文
共 50 条
  • [41] Repeatability and reproducibility of optical biometry implemented in a new optical coherence tomographer and comparison with a optical low-coherence reflectometer
    Kanclerz, Piotr
    Hoffer, Kenneth J.
    Rozema, Jos J.
    Przewlocka, Katarzyna
    Savini, Giacomo
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (11): : 1619 - 1624
  • [42] Swept-Source Optical Coherence Tomography-Based Biometry: A Comprehensive Overview
    Borgia, Alfredo
    Raimondi, Raffaele
    Sorrentino, Tania
    Santoru, Francesco
    Buzzi, Matilde
    Borgia, Vittorio
    Scorcia, Vincenzo
    Giannaccare, Giuseppe
    PHOTONICS, 2022, 9 (12)
  • [43] Efficiency and measurements agreement between swept-source OCT and low-coherence interferometry biometry systems
    Calvo-Sanz, Jorge A.
    Portero-Benito, Alejandro
    Arias-Puente, Alfonso
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2018, 256 (03) : 559 - 566
  • [44] Efficiency and measurements agreement between swept-source OCT and low-coherence interferometry biometry systems
    Jorge A. Calvo-Sanz
    Alejandro Portero-Benito
    Alfonso Arias-Puente
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2018, 256 : 559 - 566
  • [45] Comparison of a New Optical Biometer That Combines Scheimpflug Imaging With Partial Coherence Interferometry With That of an Optical Biometer Based on Swept-Source Optical Coherence Tomography and Placido-Disk Topography
    Chen, Shihao
    Zhang, Qiaoyue
    Savini, Giacomo
    Zhang, Shuangzhe
    Huang, Xiaomin
    Yu, Jinjin
    Wang, Yirang
    Ning, Rui
    Huang, Jinhai
    Tu, Ruixue
    FRONTIERS IN MEDICINE, 2022, 8
  • [46] Agreement between a new fully automatic ocular biometer based on optical low-coherence reflectometry and an optical biometer based on Scheimpflug imaging combined with partial coherence interferometry
    Ning, Rui
    Xu, Huilin
    Li, Zheng
    Yu, Jinjin
    Xu, Shuoyu
    Lei, Chak Seng
    Wang, Yiran
    Savini, Giacomo
    Schiano-Lomoriello, Domenico
    Zhou, Xingtao
    Huang, Jinhai
    BMC OPHTHALMOLOGY, 2024, 24 (01)
  • [47] Repeatability of Keratometry Measurements with Swept-Source Optical Coherence Tomography in Normal Eyes
    Poulsen, Anne
    Cui, Amy Lu David
    Seeger, Catherine
    Lehman, Erik
    Scott, Ingrid U.
    Pantanelli, Seth
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [48] Reliability of a New Swept-Source Optical Coherence Tomography Biometer in Healthy Children, Adults, and Cataract Patients
    Huang, Jinhai
    Zhao, Yuyu
    Savini, Giacomo
    Yu, Guanshun
    Yu, Jinjin
    Chen, Zhongxing
    Tu, Ruixue
    Zhao, Yune
    JOURNAL OF OPHTHALMOLOGY, 2020, 2020
  • [49] Comparison of 2 swept-source optical coherence tomography-based biometry devices
    Fisus, Andreea D.
    Hirnschall, Nino D.
    Findl, Oliver
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2021, 47 (01): : 87 - 92
  • [50] Agreement in Biometric Parameters Between Swept-Source-Optical Coherence Tomography and Optical Low-Coherence Interferometry: Insights into Clinical Precision
    Munteanu, Mihnea
    Al Barri, Leila
    Stanca, Simona
    Mocanu, Valeria
    Rosca, Cosmin
    Balica, Nicolae-Constantin
    Stanca, Horia T.
    JOURNAL OF CLINICAL MEDICINE, 2025, 14 (05)