A depth-dose measuring device using a multichannel scintillating fiber array for electron beam therapy

被引:21
|
作者
Aoyama, T
Koyama, S
Tsuzaka, M
Maekoshi, H
机构
[1] Nagoya University, College of Medical Technology, Daikominami, Higashi-ku
关键词
electron beam; depth-dose measurement; beam energy; quality assurance; scintillating fiber;
D O I
10.1118/1.598143
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The development of a new depth-dose measuring device for electron beam therapy is described. The device employs plastic scintillating fiber detectors inserted in a polymethylmethacrylate (PMMA) phantom in line along an incident electron beam. Output photons from a fiber, the number of which is proportional to the absorbed dose at each depth of the phantom, were converted to an electric signal with a photodiode. Each signal from the photodiode was transmitted to a personal computer through a multichannel analog-digital converter, and was processed to draw a depth-dose curve on the computer display. A depth-dose curve could be obtained in a measuring time of 5 s for each incident electron beam with an energy range between 4 and 21 MeV. The mean electron energies estimated using the curves and the depth-scaling factor for PMMA were consistent with those obtained from conventional depth-dose measurements using an ion chamber and a water phantom. The newly developed system, being simple and not time consuming, could be used routinely for quality assurance purposes in electron beam therapy. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:1235 / 1239
页数:5
相关论文
共 50 条
  • [1] ELECTRON-BEAM CENTRAL AXIS DEPTH-DOSE MEASUREMENTS
    KIRBY, TH
    GASTORF, RJ
    HANSON, WF
    SHALEK, RJ
    HAZLE, JD
    MEDICAL PHYSICS, 1984, 11 (03) : 399 - 400
  • [2] The depth-dose curve depends on the inhomogeneity of the material irradiated by electron beam
    Nguyen Anh Tuan
    Chau Van Tao
    Rangacharyulu, Chary
    RADIATION PHYSICS AND CHEMISTRY, 2019, 163 : 22 - 25
  • [3] A scintillating fiber beam-energy monitor for electron beam therapy
    Aoyama, T
    Maekoshi, H
    Tsuzaka, M
    Koyama, S
    MEDICAL PHYSICS, 1995, 22 (12) : 2101 - 2102
  • [4] CALCULATION AND MEASUREMENT OF DEPTH-DOSE CURVES PRODUCED BY ELECTRON-BEAM GENERATORS
    SILVERMA.J
    ROSENSTE.M
    EISEN, H
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1972, 15 (01): : 169 - &
  • [5] Evaluation of a Scintillating Plastic Optical Fiber Device for Measuring kV-Cone Beam Computed Tomography Dose
    Popotte, Christian
    Letellier, Romain
    Paul, Didier
    Waltener, Alexandre
    Guillochon, Nicolas
    Munier, Melodie
    Retif, Paul
    SENSORS, 2023, 23 (18)
  • [6] OPTIMIZATION OF FIELD FLATNESS AND DEPTH-DOSE FOR THERAPY ELECTRON-BEAMS
    MARBACH, JR
    ALMOND, PR
    PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03): : 435 - 443
  • [7] On measuring depth-dose distribution of range-modulated proton therapy fields
    Lu, Hsiao-Ming
    MEDICAL PHYSICS, 2006, 33 (07) : 2362 - 2368
  • [8] Design and construction of a new optical depth-dose reader for electron beam energy measurement
    Askarbioki, M.
    Zarandi, M. B.
    Khakshournia, S.
    Shirmardi, S. P.
    Bouzarjomehri, F.
    Mortazavi, M.
    JOURNAL OF INSTRUMENTATION, 2019, 14 (09):
  • [9] DEPTH-DOSE FLATTENING OF ELECTRON-BEAMS USING A WIRE MESH BOLUS
    ALASTI, H
    GALBRAITH, DM
    MEDICAL PHYSICS, 1995, 22 (10) : 1675 - 1683
  • [10] FUNDAMENTAL ASPECTS OF ELECTRON-BEAM LITHOGRAPHY .1. DEPTH-DOSE RESPONSE OF POLYMERIC ELECTRON-BEAM RESISTS
    HEIDENREICH, RD
    THOMPSON, LF
    FEIT, ED
    MELLIARS.CM
    JOURNAL OF APPLIED PHYSICS, 1973, 44 (09) : 4039 - 4047