A microwave interferometer with increased stability for diagnostics of steady-state plasma

被引:6
|
作者
Nagornyi, D. A. [1 ]
Nagornyi, A. G. [1 ]
机构
[1] Natl Acad Sci Ukraine, Inst Appl Sci, UA-40038 Sumy, Ukraine
关键词
52.70.Gw;
D O I
10.1134/S002044120606011X
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The schematic and results of studying an interferometer for a wavelength of 8 mm with homodyne frequency conversion for measuring the electron density of a steady-state plasma are described. The design of its waveguide system allows the drift of the indicated initial phase to be reduced to a level of +/- 0.1 degrees/h. The maximum phase shift measured is 360 degrees, the tolerable signal-power decay in plasma is 16 dB, and the maximum phase measurement error is similar to 5%.