Time-resolved beam symmetry measurement for VMAT commissioning and quality assurance

被引:1
|
作者
Barnes, Michael P. [1 ,2 ]
Greer, Peter B. [1 ,3 ]
机构
[1] Calvary Mater Hosp, Dept Radiat Oncol, Newcastle, NSW, Australia
[2] Univ Newcastle, Sch Med Radiat Sci, Newcastle, NSW 2300, Australia
[3] Univ Newcastle, Sch Math & Phys Sci, Newcastle, NSW 2300, Australia
来源
关键词
symmetry; VMAT; commissioning; quality assurance; VOLUMETRIC MODULATED ARC; DOSIMETRY; THERAPY;
D O I
10.1120/jacmp.v17i2.6026
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In volumetric-modulated arc therapy (VMAT) treatment delivery perfect beam symmetry is assumed by the planning system. This study aims to test this assumption and present a method of measuring time-resolved beam symmetry measurement during a VMAT delivery that includes extreme variations of dose rate and gantry speed. The Sun Nuclear IC Profiler in gantry mount was used to measure time-resolved in-plane and cross-plane profiles during plan delivery from which symmetry could be determined. Time-resolved symmetry measurements were performed throughout static field exposures at cardinal gantry angles, conformal arcs with constant dose rate and gantry speed, and during a VMAT test plan with gantry speed and dose rate modulation. Measurements were performed for both clockwise and counterclockwise gantry rotation and across four Varian 21iX linacs. The symmetry was found to be generally constant throughout the static field exposures to within 0.3% with an exception on one linac of up to 0.7%. Agreement in symmetry between cardinal angles was always within 1.0% and typically within 0.6%. During conformal arcs the results for clockwise and counterclockwise rotation were in agreement to within 0.3%. Both clockwise and counterclockwise tended to vary in similar manner by up to 0.5% during arc consistent with the cardinal gantry angle static field results. During the VMAT test plan the symmetry generally was in agreement with the conformal arc results. Greater variation in symmetry was observed in the low-dose-rate regions by up to 1.75%. All results were within clinically acceptable levels using the tolerances of NCS Report 24 (2015).
引用
收藏
页码:220 / 230
页数:11
相关论文
共 50 条
  • [41] MEASUREMENT MODELS FOR TIME-RESOLVED SPECTROSCOPY - A COMMENT
    HOFMANN, HF
    MAHLER, G
    QUANTUM AND SEMICLASSICAL OPTICS, 1995, 7 (04): : 489 - 497
  • [42] Time-Resolved Spectral Measurement for Fluorescence Analysis
    Saito, Mitsunori
    Koketsu, Takahiro
    Itai, Yusuke
    2019 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2019, : 450 - 454
  • [43] Time-resolved ammonia measurement in vehicle exhaust
    Mohn, J
    Forss, AM
    Brühlmann, S
    Zeyer, K
    Lüscher, R
    Emmenegger, L
    Novak, P
    Heeb, N
    INTERNATIONAL JOURNAL OF ENVIRONMENT AND POLLUTION, 2004, 22 (03) : 342 - 356
  • [44] Time-resolved measurement of stimulated polariton relaxation
    Bloch, J
    Sermage, B
    Jacquot, C
    Senellart, P
    Thierry-Mieg, V
    PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE, 2002, 190 (03): : 827 - 831
  • [45] Two-dimensional time-resolved modelling of copper HyBrID laser beam quality
    Huot, N
    Audouard, E
    Laporte, P
    OPTICS COMMUNICATIONS, 2003, 223 (4-6) : 389 - 397
  • [46] Commissioning of an Integrated Platform for Time-Resolved Treatment Delivery in Scanned Ion Beam Therapy by Means of Optical Motion Monitoring
    Fattori, G.
    Saito, N.
    Seregni, M.
    Kaderka, R.
    Pella, A.
    Constantinescu, A.
    Riboldi, M.
    Steidl, P.
    Cerveri, P.
    Bert, C.
    Durante, M.
    Baroni, G.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2014, 13 (06) : 517 - 528
  • [47] Feasibility of Time-Resolved DVH Based Treatment Delivery Verification of VMAT with EPID
    Greer, P.
    David, R.
    Wolf, J.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [48] A real-time quality assurance monitor for validation of VMAT delivery
    Smale, J. R.
    Norrlinger, B.
    Jaffray, D.
    Heaton, R.
    Huang, C.
    Ivanovski, J.
    Galbraith, D.
    Islam, M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S672 - S672
  • [49] TIME-RESOLVED SPECTROSCOPY OF A BROMINE ION-BEAM
    LEMOND, JE
    HEAD, CE
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1989, 40-1 : 209 - 210
  • [50] Time-resolved beam profilers can take the heat
    Prasad, RR
    Schein, J
    LASER FOCUS WORLD, 2000, 36 (05): : 265 - +