Evaluation of a new transcardiac conductance method for continuous on-line measurement of left ventricular volume

被引:13
|
作者
Steendijk, P [1 ]
Lardenoye, JW [1 ]
van der Velde, ET [1 ]
Schalij, MJ [1 ]
Baan, J [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Cardiol, Cardiac Physiol Lab, NL-2300 RC Leiden, Netherlands
关键词
cardiac function; cardiac output; conductance catheter; contractile state; end-systolic volume; end-diastolic volume; hemodynamics; monitoring; on-line measurements; sheep; thoracic impedance;
D O I
10.1097/00003246-200005000-00057
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate a new, less invasive, conductance method to measure continuous on-line left ventricular volume. End-systolic and end-diastolic volumes obtained with this transcardiac conductance method were compared with simultaneous measurements using the conventional intracardiac conductance catheter. Design: Controlled animal study. Setting: Research laboratory in a university hospital. Subjects: Six sheep. Interventions: Anesthetized sheep were instrumented and inotropic condition was varied by beta-receptor stimulation (5 mu g/kg/min of dobutamine) and beta-receptor blockade (1 mg/kg of propranolol). In each condition (control, dobutamine, repeat control, propranolol), ventricular volume was varied over a wide range by gradual preload reduction using a vena caval balloon catheter. Measurements and Main Results: We compared the two methods by performing linear regression analysis on simultaneous end-systolic and end-diastolic volumes obtained during gradual caval occlusions. We statistically analyzed the intercepts, slopes, and correlation coefficients of the regression equations relating the transcardiac and conductance catheter measurements to determine the effects of interanimal variability, inotropic condition, and cardiac phase on the relationship between the two methods. The results show an excellent linear correlation between the two methods (mean intercept, -1.82 +/- 1.24 mi; mean slope, 0.787 +/- 0.024 and r(2) = .94), Both slope and intercept of the relationship between the two methods show a significant interanimal and cardiac phase related variability but no significant dependence on inotropic condition. Conclusions: The significant interanimal variability indicates that the new method requires individual calibration in each subject. However, the small variability of the regression coefficients with changes in condition indicates that after initial calibration, end-systolic and end-diastolic volume can be followed accurately even in the presence of large changes in volume and inotropic state. This new method may facilitate quantitative continuous assessment of cardiac function in clinical practice, for example, in the intensive care unit.
引用
收藏
页码:1599 / 1606
页数:8
相关论文
共 50 条
  • [21] The importance of on-line recording of ventricular volume
    Schiereck, P
    JOURNAL OF CARDIOVASCULAR DIAGNOSIS AND PROCEDURES, 1996, 13 (02): : 171 - 176
  • [22] PROPERTIES OF LEFT-VENTRICULAR VOLUME MEASUREMENT BY CONDUCTANCE CATHETER, ASSESSED IN AN ANALYTICAL MODEL
    DEBRUIN, HG
    MUR, G
    VANDERVELDE, ET
    BAAN, J
    FEDERATION PROCEEDINGS, 1984, 43 (03) : 732 - 732
  • [23] Numerical model study of dual frequency conductance measurement of left ventricular volume in the mouse
    Pearce, JA
    Feldman, MD
    SECOND JOINT EMBS-BMES CONFERENCE 2002, VOLS 1-3, CONFERENCE PROCEEDINGS: BIOENGINEERING - INTEGRATIVE METHODOLOGIES, NEW TECHNOLOGIES, 2002, : 1260 - 1261
  • [24] Right ventricular volume measurement by conductance catheter
    Danton, MHD
    Greil, GF
    Byrne, JG
    Hsin, M
    Cohn, L
    Maier, SE
    AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2003, 285 (04): : H1774 - H1785
  • [25] On-line estimation of left ventricular stroke volume using transoesophageal echocardiography and acoustic quantification
    Greim, CA
    Roewer, N
    Laux, G
    Esch, JSA
    BRITISH JOURNAL OF ANAESTHESIA, 1996, 77 (03) : 365 - 369
  • [28] MEASUREMENT OF LEFT VENTRICULAR VOLUME BY ANGIOGRAPHY
    LETAC, B
    BARRIER, A
    TOUSSAIN.C
    LETOURNE.J
    CRIBIER, A
    OUEST MEDICAL, 1974, 27 (05): : 463 - 469
  • [29] On-line measurement method for material volume based on reference plane extraction
    Zhao, Qijie
    Meng, Qingxu
    Zhang, Xi
    Lu, Jianxia
    Tu, Dawei
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2015, 36 (12): : 2731 - 2737
  • [30] CONDUCTANCE CATHETER MEASUREMENT OF LEFT-VENTRICULAR VOLUME - EVIDENCE FOR NONLINEARITY WITHIN CARDIAC CYCLE
    SZWARC, RS
    LAURENT, D
    ALLEGRINI, PR
    BALL, HA
    AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 268 (04): : H1490 - H1498