Portable gas chromatograph with integrated components

被引:63
|
作者
Dziuban, JA
Mróz, J
Szczygielska, M
Malachowski, M
Górecka-Drzazga, A
Walczak, R
Bula, W
Zalewski, D
Nieradko, L
Lysko, J
Koszur, J
Kowalski, P
机构
[1] Wroclaw Univ Technol, Fac Microsyst Elect & Photon, PL-50372 Wroclaw, Poland
[2] EMAG, Res & Dev Ctr Elect Engn, PL-40189 Katowice, Poland
[3] Lab Opt PM Duffieux, F-25030 Besancon, France
[4] Inst Electr Mat Technol, PL-02668 Warsaw, Poland
关键词
portable gas chromatograph; micromachined injector; thermal conductivity detector; silicon micromechanics;
D O I
10.1016/j.sna.2004.04.028
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In the paper, a portable gas chromatograph with silicon/glass integrated micromachined flushed injector and doubled thermal conductivity detectors (TCD) has been presented. The injector has two symmetrical dosing loops, each of 7 mul volume and one fast injection microvalve, pneumatically operated. All parts of the injector are integrated on 35 x 35 x 2.5 mm(3) glass/silicon/glass die. Teflon coated thin polyimide foil is applied as the moving actuator of the microvalve of the injector. Three types of TCDs have been tested. In the first type of TCD, four thin-film Pt heaters/thermoresistors made on thin silicon nitride CVD deposited membrane suspended over deeply etched in a silicon substrate two channels, are applied. In the next two TCDs, miniature spirals made of thin Pt solid wire are assembled inside glass/silicon chip. Two configurations of spirals position have been tested: in-line serial and in-line parallel. Computer simulations, supported by experiments have indicated better parameters of TCD with parallel configuration of spirals. The TCDs with Pt spirals are resistant against carrier gas decays. In the portable gas chromatograph two capillary columns are used: active (molecular sieve) and passive. Columns and co-working TCD are thermally stabilized in an oven equipped with a thick-film heater made on polyimide foil. The instrument is equipped with built-in microcomputer with two 16-bit microprocessors and is operated (controlled) by an external portable PC. All important parameters of analysis (possible single or continuous mode of work) are adjusted electronically, as well as data acquisition and preparing of the complete analysis effect in the form of chromatograms and tables indicating qualitative and quantitative composition of gas mixture. The instrument is able to work in on-line continuous measurements, tests made for artificial atmosphere of deep mining industry have shown successful work of the instrument. (C) 2004 Published by Elsevier B.V.
引用
收藏
页码:318 / 330
页数:13
相关论文
共 50 条
  • [21] Micro-flame ionization detector with a novel structure for portable gas chromatograph
    Wang, Jianwei
    Wang, Hua
    Duan, Chunfeng
    Guan, Yafeng
    TALANTA, 2010, 82 (03) : 1022 - 1026
  • [22] Integrated portable gas dispersion device
    Opoku, Jeffrey
    Lepage, Mark R.
    Langlois, Raymond
    Waters, Kristian E.
    CANADIAN METALLURGICAL QUARTERLY, 2024,
  • [24] ELECTROPHYSIOLOGICAL MEASUREMENTS OF THE ODOUR OF SINGLE COMPONENTS OF A MIXTURE SEPARATED IN A GAS CHROMATOGRAPH
    OTTOSON, D
    SYDOW, EV
    LIFE SCIENCES, 1964, 3 (10) : 1111 - 1115
  • [25] GAS CHROMATOGRAPH
    DONNER, W
    JOURNAL OF GAS CHROMATOGRAPHY, 1965, 3 (02): : A14 - &
  • [26] GAS CHROMATOGRAPH
    NORLEY, WN
    NEW ENGLAND JOURNAL OF MEDICINE, 1964, 271 (18): : 963 - &
  • [27] Gas chromatograph
    不详
    MICRO, 1995, 13 (02): : 54 - 54
  • [28] PORTABLE GAS CHROMATOGRAPH GIVES QUICK FIELD TEST FOR CHLORINATED HYDROCARBONS IN CRUDE.
    Bryant, M.R.
    Krottinger, D.L.
    Oil and Gas Journal, 1983, 81 (25): : 105 - 108
  • [29] A modular platform, portable gas chromatograph based on miniaturized chip injection and detection technologies
    de Zeeuw, J
    van Hove, R
    Curvers, J
    Verhoeve, L
    AMERICAN LABORATORY, 2002, 34 (25) : 25 - 27
  • [30] A method for on-site analysis of urinary benzene by means of a portable gas-chromatograph
    Suna, S
    Jitsunari, F
    Asakawa, F
    Hirao, T
    Mannami, T
    Suzue, T
    JOURNAL OF OCCUPATIONAL HEALTH, 2005, 47 (01) : 74 - 77