Efficiency of different hollow-fiber hemofilters in continuous arteriovenous hemodiafiltration

被引:3
|
作者
Morabito, S
Pierucci, A
Marinelli, R
Cicciarelli, A
Guzzo, I
Cinotti, GA
Marino, B
Chiavarelli, R
机构
[1] Univ Roma La Sapienza, Policlin Umberto I, Div Nephrol, Rome, Italy
[2] Univ Roma La Sapienza, Policlin Umberto I, Dept Cardiac Surg, Rome, Italy
关键词
acute renal failure; continuous renal replacement therapy continuous arteriovenous hemodiafiltration hollow-fiber hemofilters; polyamide; polyacrylonitrile; acrylonitrile sodium methallylsulfonate; urea clearance;
D O I
10.1159/000013567
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Low dialysate to blood flow rate ratios are a unique characteristic of continuous arteriovenous hemodiafiltration (CAVHDF) that should allow complete saturation of dialysis fluid with small-molecular-weight blood solutes. The aim of the investigation was to evaluate the performance of different hemofilters in CAVHDF. In 10 critically ill patients with acute renal failure, the efficiency of four hollow-fiber hemofilters, polyamide 0.6 m(2), polyacrylonitrile (PAN) 0.3 and 0.6 m(2), acrylonitrile sodium methallylsulfonate (AN69HF) 0.6 m(2), has been evaluated. For comparison, dialysate flow rates (Q(di)) were standardized to 16.6 and 25 ml/min. Samples for urea nitrogen were obtained from the arterial blood line (C-bi) and from the dialysate exit port (C-do) within 24-hour running time. Outflowing dialysate (Q(do)) was also measured at the same time. Blood flow (Q(b)) was calculated by the bubble transit time technique. Diffusive and total urea clearances were determined. AN69HF and PAN hemofilters provided higher clearances than the polyamide hemofilter. Despite the smaller surface area, PAN 0.3 m(2) had a total urea clearance comparable to that of PAN 0.6 m(2) and AN69HF at Q(di) = 16.6 ml/min. While at Q(di) = 16.6 ml/min equilibrium between blood and dialysate (C-do/C-bi congruent to 1) occurred with the AN69HF and PAN hemofilters, at Q(di) = 25 ml/min the equilibrium was obtained only with the AN69HF hemofilter. In conclusion, almost complete urea saturation of dialysis fluid has not been obtained with all hemofilters tested here. In our experience, membrane characteristics play an important role in determining diffusive efficiency in CAVHDF. Copyright (C) 2000 S. Karger AG, Basel.
引用
收藏
页码:116 / 121
页数:6
相关论文
共 50 条
  • [41] POLYETHYLENE HOLLOW-FIBER PLASMA SEPARATOR
    ASANUMA, Y
    KOYAMA, K
    OUCHI, K
    TAKAGI, Y
    OWADA, Y
    NAKAGAWA, K
    OMOKAWA, S
    KIKUCHI, J
    GOTO, H
    SOTO, T
    IWAI, Y
    MORI, M
    INOUE, M
    NAKANISHI, H
    ARTIFICIAL ORGANS, 1984, 8 (02) : 238 - 238
  • [42] GAS SEPARATIONS IN HOLLOW-FIBER ADSORBERS
    GILLESKIE, GL
    PARKER, JL
    CUSSLER, EL
    AICHE JOURNAL, 1995, 41 (06) : 1413 - 1425
  • [43] A PROSPECTIVE COMPARATIVE-STUDY OF CONTINUOUS ARTERIOVENOUS HEMODIAFILTRATION AND CONTINUOUS VENOVENOUS HEMODIAFILTRATION IN CRITICALLY ILL PATIENTS
    BELLOMO, R
    PARKIN, G
    LOVE, J
    BOYCE, N
    AMERICAN JOURNAL OF KIDNEY DISEASES, 1993, 21 (04) : 400 - 404
  • [44] Continuous potato starch hydrolysis process in a membrane reactor with tubular and hollow-fiber membranes
    Grzeoekowiak-Przywecka, A
    Slominska, L
    DESALINATION, 2005, 184 (1-3) : 105 - 112
  • [45] REMOVAL OF TOXIC LEVELS OF N-ACETYLPROCAINAMIDE WITH CONTINUOUS ARTERIOVENOUS HEMOFILTRATION OR CONTINUOUS ARTERIOVENOUS HEMODIAFILTRATION
    DOMOTO, DT
    BROWN, WW
    BRUGGENSMITH, P
    ANNALS OF INTERNAL MEDICINE, 1987, 106 (04) : 550 - 552
  • [46] CONTINUOUS ARTERIOVENOUS HEMOFILTRATION (CAVH) AND CONTINUOUS ARTERIOVENOUS HEMODIAFILTRATION (CAVHD) IN PATIENTS WITH MULTIPLE ORGAN FAILURE
    SADJADI, SA
    SHAH, RM
    KIDNEY INTERNATIONAL, 1990, 37 (01) : 318 - 318
  • [47] CONTINUOUS ARTERIOVENOUS HEMODIAFILTRATION IN POSTOPERATIVE AND TRAUMATIC RENAL FAILURE
    TOMINAGA, GT
    INGEGNO, MD
    SCANNELL, G
    PAHL, MV
    WAXMAN, K
    AMERICAN JOURNAL OF SURGERY, 1993, 166 (06): : 612 - 616
  • [48] A MATHEMATICAL-MODEL OF CONTINUOUS ARTERIOVENOUS HEMODIAFILTRATION (CAVHD)
    AKCAHUSEYIN, E
    VINCENT, HH
    VANITTERSUM, FJ
    VANDUYL, WA
    SCHALEKAMP, MADH
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1990, 31 (3-4) : 215 - 224
  • [49] CONTINUOUS ARTERIOVENOUS HEMODIAFILTRATION (CAVHD) - DETERMINANTS OF SOLUTE TRANSPORT
    VINCENT, HH
    VANITTERSUM, FJ
    SCHALEKAMP, MADH
    KIDNEY INTERNATIONAL, 1989, 35 (06) : 1435 - 1435
  • [50] IMPROVING THE CONTINUOUS PRODUCTION OF LIPOLYZED BUTTEROIL WITH PREGASTRIC ESTERASES IMMOBILIZED IN A HOLLOW-FIBER REACTOR
    GARCIA, HS
    HILL, CG
    BIOTECHNOLOGY TECHNIQUES, 1995, 9 (07) : 467 - 470