Open-capsule intraocular lens to prevent posterior capsule opacification

被引:10
|
作者
Katsuki, Yoko [1 ,2 ]
Matsushima, Hiroyuki [1 ]
Mukai, Koichiro [1 ]
Watabiki, Satoshi [1 ]
Aose, Masamoto [1 ]
Terauchi, Wataru [1 ]
Nagamoto, Toshiyuki [3 ]
Senoo, Tadashi [1 ]
机构
[1] Dokkyo Med Univ, Dept Ophthalmol, 880 Kitakobayashi, Mibu, Tochigi 3210293, Japan
[2] HOYA Corp, Div Med, Tokyo, Japan
[3] Nagamoto Eye Clin, Tokyo, Japan
来源
关键词
OPTIC EDGE DESIGN; ENDOCAPSULAR EQUATOR RING; EPITHELIAL-CELLS; BENDING RING; AQUEOUS-HUMOR; LONG-TERM; SILICONE; SHARP; INHIBITION; EFFICACY;
D O I
10.1016/j.jcrs.2018.12.022
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To develop a single-piece open-capsule intraocular lens (IOL) that can be inserted through a small incision and that prevents posterior capsule opacification (PCO) by expanding the capsule and circulating aqueous humor into the capsular bag. Setting: Department of Ophthalmology, Dokkyo Medical University, Tochigi, Japan. Design: Experimental study. Method: Using the same hydrophobic acrylic material as the NY-60 IOL, a prototype open-capsule IOL was constructed. The IOL has a single optic and 2 haptics, with a 2.8 mm high spacer and holes through which aqueous humor circulates into the capsular bag by separating the anterior capsule from the posterior capsule and expanding the capsule. The open-capsule IOL or NY-60 (as a control group) was inserted in rabbit eyes. Posterior capsule opacification development was evaluated by measuring the thickness of the cell layer at the center of the posterior capsule on histopathologic specimens and statistically comparing the thickness between the open-capsule IOL group and control group. Results: The open-capsule IOL could be inserted through a 3.2 mm corneal incision using a D cartridge. The mean thickness of the cell layer at the center of the posterior capsule was 4.78 mu m +/- 2.61 (SD) in the open-capsule IOL group and 101.14 +/- 25.19 mu m in the control group and was significantly smaller in the open-capsule IOL group. Conclusion: The prototype single-piece IOL could be implanted through a small incision and prevented PCO by expanding the lens capsule and circulating aqueous humor into the capsular bag. (C) 2019 Published by Elsevier Inc. on behalf of ASCRS and ESCRS
引用
收藏
页码:1007 / 1012
页数:6
相关论文
共 50 条
  • [21] Effect of a silicone intraocular lens with a sharp posterior optic edge on posterior capsule opacification
    Buehl, W
    Menapace, R
    Sacu, S
    Kriechbaum, K
    Koeppl, C
    Wirtitsch, M
    Georgopoulos, M
    Findl, O
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (08): : 1661 - 1667
  • [22] Effect of an acrylic intraocular lens with a sharp posterior optic edge on posterior capsule opacification
    Buehl, W
    Findl, O
    Menapace, R
    Rainer, G
    Sacu, S
    Kiss, B
    Petternel, V
    Georgopoulos, M
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07): : 1105 - 1111
  • [23] Posterior capsule opacification prevention by an intraocular lens incorporating a micropatterned membrane on the posterior surface
    Ellis, Nathan
    Werner, Liliana
    Balendiran, Vaishnavi
    Shumway, Caleb
    Jiang, Bill
    Mamalis, Nick
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (01): : 102 - 107
  • [24] Randomized intraindividual comparison of posterior capsule opacification between a microincision intraocular lens and a conventional intraocular lens
    Cleary, Georgia
    Spalton, David J.
    Hancox, Joanne
    Boyce, James
    Marshall, John
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (02): : 265 - 272
  • [25] Active oxygen processing for acrylic intraocular lenses to prevent posterior capsule opacification
    Matsushima, Hiroyuki
    Iwamoto, Hidetoshi
    Mukai, Kouichiro
    Obara, Yoshitaka
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (06): : 1035 - 1040
  • [26] EFFECT OF INTRAOCULAR-LENS SIZE ON POSTERIOR CAPSULE OPACIFICATION AFTER PHACOEMULSIFICATION
    MAMALIS, N
    CRANDALL, AS
    LINEBARGER, E
    SHEFFIELD, WK
    LEIDENIX, MJ
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1995, 21 (01): : 99 - 102
  • [27] Effect of in-the-bag intraocular lens fixation on the prevention of posterior capsule opacification
    Ram, J
    Pandey, SK
    Apple, DJ
    Werner, L
    Brar, GS
    Singh, R
    Chaudhary, KP
    Gupta, A
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (07): : 1039 - 1046
  • [28] Posterior capsule opacification in eyes with a silicone or poly(methyl methacrylate) intraocular lens
    Kim, MJ
    Lee, HY
    Joo, CK
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1999, 25 (02): : 251 - 255
  • [29] Effect of intraocular lens design on posterior capsule opacification: an in-vitro model
    Beck, R
    zur Linden, B
    Stave, J
    Werner, C
    Nitschke, M
    Guthoff, R
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2001, 218 (02) : 111 - 115
  • [30] Posterior capsule opacification and anterior capsule opacification
    Bertelmann, Eckart
    Kojetinsky, Corina
    CURRENT OPINION IN OPHTHALMOLOGY, 2001, 12 (01) : 35 - 40