Dynamic Transmit-Receive Beamforming by Spatial Matched Filtering for Ultrasound Imaging with Plane Wave Transmission

被引:6
|
作者
Chen, Yuling [1 ]
Lou, Yang [2 ]
Yen, Jesse [2 ]
机构
[1] Zonare Med Syst, Mindray North Amer Innovat Ctr, Mountain View, CA USA
[2] Univ Southern Calif, Los Angeles, CA USA
关键词
ultrafast imaging; plane wave transmissions; spatial matched filter; ultrasound imaging; transmit-receive focusing; FRAME RATE ULTRASONOGRAPHY; LAYER TRANSDUCER ARRAY; TRANSIENT ELASTOGRAPHY; DECONVOLUTION;
D O I
10.1177/0161734617692017
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
During conventional ultrasound imaging, the need for multiple transmissions for one image and the time of flight for a desired imaging depth limit the frame rate of the system. Using a single plane wave pulse during each transmission followed by parallel receive processing allows for high frame rate imaging. However, image quality is degraded because of the lack of transmit focusing. Beamforming by spatial matched filtering (SMF) is a promising method which focuses ultrasonic energy using spatial filters constructed from the transmit-receive impulse response of the system. Studies by other researchers have shown that SMF beamforming can provide dynamic transmit-receive focusing throughout the field of view. In this paper, we apply SMF beamforming to plane wave transmissions (PWTs) to achieve both dynamic transmit-receive focusing at all imaging depths and high imaging frame rate (>5000 frames per second). We demonstrated the capability of the combined method (PWT + SMF) of achieving two-way focusing mathematically through analysis based on the narrowband Rayleigh-Sommerfeld diffraction theory. Moreover, the broadband performance of PWT + SMF was quantified in terms of lateral resolution and contrast from both computer simulations and experimental data. Results were compared between SMF beamforming and conventional delay-and-sum (DAS) beamforming in both simulations and experiments. At an imaging depth of 40 mm, simulation results showed a 29% lateral resolution improvement and a 160% contrast improvement with PWT + SMF. These improvements were 17% and 48% for experimental data with noise.
引用
收藏
页码:207 / 223
页数:17
相关论文
共 41 条
  • [1] Transmit-Receive Generalized Spatial Modulation Based on Dual-layered MIMO Transmission
    Perovic, Nemanja Stefan
    Di Renzo, Marco
    Flanagan, Mark F.
    2019 IEEE 30TH ANNUAL INTERNATIONAL SYMPOSIUM ON PERSONAL, INDOOR AND MOBILE RADIO COMMUNICATIONS (PIMRC), 2019, : 682 - 688
  • [2] Integrated 16-channel Transmit and Receive Beamforming ASIC for Ultrasound Imaging
    Dusa, Chandrashekar
    Kalalii, Samiyuktha
    Rajalakshmi, P.
    Rao, Omkeshwar
    2015 28TH INTERNATIONAL CONFERENCE ON VLSI DESIGN (VLSID), 2015, : 215 - 220
  • [3] Ultrasound coherence imaging using hardware receive beamforming and broad transmit beams
    Bottenus, Nick
    Trahey, Gregg E.
    Uestuener, Kutay F.
    MEDICAL IMAGING 2015: ULTRASONIC IMAGING AND TOMOGRAPHY, 2015, 9419
  • [4] Adaptive Beamforming in Ultrasound Imaging with Plane Wave Compounding
    Zhao, Jinxin
    Zeng, Xing
    Yu, Jinhua
    Wang, Yuanyuan
    2014 INTERNATIONAL CONFERENCE ON AUDIO, LANGUAGE AND IMAGE PROCESSING (ICALIP), VOLS 1-2, 2014, : 53 - 57
  • [5] PLANE WAVE MEDICAL ULTRASOUND IMAGING USING ADAPTIVE BEAMFORMING
    Holfort, Iben Kraglund
    Gran, Fredrik
    Jensen, Jorgen Arendt
    2008 IEEE SENSOR ARRAY AND MULTICHANNEL SIGNAL PROCESSING WORKSHOP, 2008, : 288 - +
  • [6] A Kerfless Dual-layer Transducer Combined with Beamforming by Spatial Matched Filtering for High Frame Rate Ultrasound Imaging
    Chen, Yuling
    Yen, Jesse T.
    2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 1001 - 1004
  • [7] A modified sidelobe blanking beamforming method for ultrasound plane wave imaging
    Tong, Lin
    Wang, Ping
    Shen, Yue
    Shi, Xinwang
    Chen, Jinghan
    Zhao, Fenglong
    Zhou, Xiaowei
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2024, 156 (02): : 1058 - 1069
  • [8] An integrated circuit with transmit beamforming and parallel receive channels for 3D ultrasound imaging:: Testing and characterization
    Wygant, Ira O.
    Jamal, Nafis
    Lee, Hyunjoo J.
    Nikoozadeh, Amin
    Zhuang, Xuefeng
    Oralkan, Oemer
    Ergun, Arif S.
    Karaman, Mustafa
    Khuri-Yakub, Butrus T.
    2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 25 - +
  • [9] Gated Transmit and Fresnel-Based Receive Beamforming With a Phased Array for Low-Cost Ultrasound Imaging
    Yen, Jesse T.
    Nguyen, Man Minh
    Lou, Yang
    Shin, Jun S.
    Chen, Yuling
    Tarnoff, Harry L.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (06) : 2183 - 2192
  • [10] Improved Plane Wave Imaging with Retrospective Transmit Focusing and Weighting Based Post Filtering
    Hu, Chang-Lin
    Li, Meng-Lin
    2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,