A Kerfless Dual-layer Transducer Combined with Beamforming by Spatial Matched Filtering for High Frame Rate Ultrasound Imaging

被引:4
|
作者
Chen, Yuling [1 ]
Yen, Jesse T. [1 ]
机构
[1] Univ So Calif, Dept Biomed Engn, Los Angeles, CA 90089 USA
来源
2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS) | 2014年
关键词
kerfless transducer; dual-layer transducer design; spatial matched filtering; transmit-receive focusing; HIGH-FREQUENCY; LINEAR ARRAYS;
D O I
10.1109/ULTSYM.2014.0245
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Dicing is typically required to create elements for array transducers made with piezoelectric ceramics. Kerfs, which are the cutting spaces generated during dicing, are introduced to prevent electrical and mechanical coupling, or cross-talk, between adjacent elements. Unfortunately, this dicing procedure can be very difficult and challenging, especially for high-frequency array transducers (> 30 MHz) because of the fine element spacing and small kerf requirements. In this work, we introduced the design, fabrication, and preliminary imaging using a transducer that does not require dicing of elements. The structure of this transducer is a dual-layer design, which has separate layers for transmit and receive respectively. A piece of un-diced PZT-5H ceramic was used as the transmitter that launched plane waves directly ahead. A sheet of P[VDF-TrFE] copolymer served as the receiver. In addition to simplifying the fabrication procedure, this kerfless dual-layer transducer also enables high frame rate imaging of the system because of the plane wave transmissions. However, the image quality of the transducer was intrinsically degraded due to the loss of transmit focusing capability. In order to compensate for the degradation, beamforming by spatial matched filtering (SMF), which is capable of dynamically focusing both the transmit and receive beams, was implemented to replace the conventional delay-andsum (DAS) beamforming onto the kerfless transducer.
引用
收藏
页码:1001 / 1004
页数:4
相关论文
共 18 条
  • [1] Improved contrast for high frame rate imaging using coherent compounding combined with spatial matched filtering
    Lou, Yang
    Yen, Jesse T.
    ULTRASONICS, 2017, 78 : 152 - 161
  • [2] Minimum variance beamforming for high frame-rate ultrasound imaging
    Holfort, Then Kraglund
    Gran, Fredrik
    Jensen, Jorgen Arendt
    2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 1541 - 1544
  • [3] High Frame-Rate Ultrasound Imaging Using Deep Learning Beamforming
    Ghani, Muhammad Usman
    Meral, F. Can
    Vignon, Francois
    Robert, Jean-luc
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 295 - 298
  • [4] Dynamic Transmit-Receive Beamforming by Spatial Matched Filtering for Ultrasound Imaging with Plane Wave Transmission
    Chen, Yuling
    Lou, Yang
    Yen, Jesse
    ULTRASONIC IMAGING, 2017, 39 (04) : 207 - 223
  • [5] High Frame Rate 3-D Ultrasound Imaging Using Separable Beamforming
    Ming Yang
    Richard Sampson
    Siyuan Wei
    Thomas F. Wenisch
    Chaitali Chakrabarti
    Journal of Signal Processing Systems, 2015, 78 : 73 - 84
  • [6] High Frame Rate 3-D Ultrasound Imaging Using Separable Beamforming
    Yang, Ming
    Sampson, Richard
    Wei, Siyuan
    Wenisch, Thomas F.
    Chakrabarti, Chaitali
    JOURNAL OF SIGNAL PROCESSING SYSTEMS FOR SIGNAL IMAGE AND VIDEO TECHNOLOGY, 2015, 78 (01): : 73 - 84
  • [7] A Dual-layer Micromachined PMN-PT 1-3 Composite Transducer for Broadband Ultrasound Imaging
    Li, Sibo
    Huang, Wenbin
    Jiang, Xiaoning
    Jian, Xiaohua
    Cui, Yaoyao
    2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 773 - 776
  • [8] A 5-MHz Cylindrical Dual-Layer Transducer Array for 3-D Transrectal Ultrasound Imaging
    Chen, Yuling
    Nguyen, Man
    Yen, Jesse T.
    ULTRASONIC IMAGING, 2012, 34 (03) : 181 - 195
  • [9] A Joint Delay-and-Sum and Fourier Beamforming Method for High Frame Rate Ultrasound Imaging
    Zhao, Wei
    Feng, Shuai
    Wang, Yadan
    Wang, Yuanguo
    Han, Zhihui
    Peng, Hu
    CMES-COMPUTER MODELING IN ENGINEERING & SCIENCES, 2020, 123 (01): : 427 - 440
  • [10] High Frame Rate Volumetric Imaging of Microbubbles Using a Sparse Array and Spatial Coherence Beamforming
    Wei, Luxi
    Wahyulaksana, Geraldi
    Meijlink, Bram
    Ramalli, Alessandro
    Noothout, Emile
    Verweij, Martin D.
    Boni, Enrico
    Kooiman, Klazina
    van der Steen, Antonius F. W.
    Tortoli, Piero
    de Jong, Nico
    Vos, Hendrik J.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (10) : 3069 - 3081